During its 25 years of existence, the Inorganic Analysis Working Group of the Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology (CCQM IAWG) has achieved much in establishing comparability of measurement results. Impressive work has been done on comparison exercises related to real-world problems in fields such as ecology, food, or health. In more recent attempts, measurements and comparisons were focused on calibration solutions which are the basis of most inorganic chemical measurements. This contribution deals with the question of how to achieve full and transparent SI traceability for the values carried by such solutions. Within this framework, the use of classical primary methods (CPMs) is compared to the use of a primary difference method (PDM). PDM is a method with a dual character, namely a metrological method with a primary character, based on the bundling of many measurement methods for individual impurities, which lead to materials with certified content of the main component. As in classical methods, where small corrections for interferences are accepted, in PDM, many small corrections are bundled. In contrast to classical methods, the PDM is universally applicable to all elements in principle. Both approaches can be used to certify the purity (expressed as mass fraction of the main element) of a high-purity material. This is where the metrological need of National Metrology Institutes (NMIs) for analytical methods meet the challenges of analytical methods. In terms of methods, glow discharge mass spectrometry (GMDS) with sufficient uncertainties for sufficiently small impurity contents is particularly noteworthy for the certification of primary transfer standards (PTS), and isotope dilution mass spectrometry (IDMS), which particularly benefits from PTS (back-spikes) with small uncertainties, is particularly noteworthy for the application. The corresponding relative uncertainty which can be achieved using the PDM is very low (< 10−4). Acting as PTS, they represent the link between the material aspect of the primary calibration solutions and the immaterial world of the International System of Units (SI). The underlying concepts are discussed, the current status of implementation is summarised, and a roadmap of the necessary future activities in inorganic analytical chemistry is sketched. It has to be noted that smaller measurement uncertainties of the purity of high-purity materials not only have a positive effect on chemical measurements, but also trigger new developments and findings in other disciplines such as thermometry or materials science.
Within the Working Group on Inorganic Analysis (IAWG) of the Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology (CCQM) international key comparisons and pilot studies related to inorganic analysis are carried to ensure consistency in this field at the highest level. Some of these comparisons deal directly with the preparation and characterization of monoelemental solutions or with topics, closely related. The importance of monoelemental solutions lies in the fact that almost every measurement in inorganic analysis relies on the comparison with either a reference material, or references in form of solutions, usually (mono)elemental solutions. All quantitative measurement approaches, e.g. isotope dilution or standard addition, need an accurate reference solution made from a well characterized reference material, prepared under full gravimetric control. These primary (monoelemental) solutions do not only serve as arbitrary references/calibration solutions, but they also link up measurement results to the International System of units (SI), this way establishing the so-called metrological traceability to a measurement unit of the SI. Without such solutions on the highest possible level of accuracy and with the smallest possible associated uncertainties (for e.g. element content and/or impurities), an analysis itself can never be as good as it could be with appropriate reference solutions. This article highlights select key comparisons and pilot studies dealing with monoelemental solution related topics within the IAWG from the foundation of CCQM – 25 years ago – up to latest achievements in the field of inorganic analysis.
A candidate material for the use as primary standard for silver determination was characterized with respect to total purity. Except the radioactive elements and He, all possible impurities were considered. Based on glow discharge mass spectrometry, inductively coupled plasma-mass spectrometry and carrier gas hot extraction measurements, the demonstrated total purity and its standard uncertainty is w(Ag) = (99.999 52 ± 0.000 11) %. The purity value and its uncertainty is dominated by the contributions from the measurements of the nonmetallic impurities, namely S, N, C and O.
Calibration of two commercially available glow discharge double focusing mass spectrometers, the VG 9000 and Element GD, is described using synthetic pin standards pressed from solution doped copper and zinc matrices. A special pressing die was developed for this purpose and optimal results were obtained with the highest possible pressures, i.e., 95 kN·cm- 2. This calibration approach permits the determination of trace element mass fractions down to µg·kg- 1 with small uncertainties and additionally provides traceability of the GD-MS results in the most direct manner to the SI (International System of Units). Results were validated by concurrent measurements of a number of compact copper and zinc certified reference materials. The impact of the sample pin cross-section (circular or square) was investigated with the use of a new pin-sample holder system for the Element GD. The pin-sample holder was designed by the manufacturer for pin-samples having circular cross-section; however, samples with square pin cross-section were also shown to provide acceptable results. Relative Sensitivity Factors for some 50 analytes in copper (VG 9000, Element GD) and zinc matrices (VG 9000) are presented. The field of applicability of GD-MS may be considerably extended via analysis of pin geometry samples based on their ease of preparation, especially with respect to the accuracy and traceability of the results and the enhanced number of analytes which can be reliably calibrated using such samples.
Determining the influence of impurities on the fixed-point emperatures
of the ITS-90 requires the completion of several tasks. In this paper, the Progress made at Physikalisch-Technische Bundesanstalt (PTB) and BAMFederal Institute for Materials Research and Testing is presented and remaining questions are discussed.
The projected characterization procedure at PTB, which is based on the established SIE method (sum of the individual estimates), using a new indium fixed-point cell is described as an example. This procedure includes an SI-traceable chemical Analysis of the material in the fixed-point cell with sufficiently low uncertainties, the individual experimental determination of the influence of the quantified impurities on the fixed-point temperature, and the establishment of direct links to the phase-transition temperatures of the national standard and of an assumed material of ideal purity.
A characteristic difference to the common practice is the chemical analysis of the fixed-point metal being done after determining the cells freezing temperature. This allows for the detection and consideration of contamination and purification effects due to the filling process, or due to the contact with the carbon crucible and other parts of the fixed-point cell. A chemical analysis of an indium fixed-point was carried out by BAM with relative measurement uncertainties below 30 % which have not been previously achieved. The results provide evidence for the precipitation of some impurities,
which is apparently inconsistent with the corresponding binary phase diagrams, but was explained in a recent publication. Implications for the use of the SIE method shall be described briefly at the end.
Mass spectrometric detection can play a prominent role in the assessment of different reactions occurring at varied temperatures in a milled ZnO/Al mixture. This is the first time that online mass spectrometric information was used as a tool for monitoring and understanding the chemical reduction process mechanism. We have observed four different types of reaction taking place: (1) distillation of Zn metal, (2) reduction of ZnO by activated Al, (3) melting of Al and finally (4) reduction of ZnO by inactivated Al. The experimental conditions and results observed by QMS were supported with literature data and physical measurement data from X-Ray Diffraction (XRD) which gave us an idea about the complex reaction cascade which occurred during the formation of the zinc metal.
Synthetic pressed metal powder standards doped with standard solutions were used for the calibration of both commercially available high resolution GD-MS instruments Element GD and VG 9000 for zinc matrix. Different quantification procedures (IBR, Standard RSF, matrix matched RSF from the calibration with CRMs and use of doped synthetic standards) are compared using zinc matrix as an example, whereas the calibration with doped pellets turned out to be the best quantification technique for high-purity materials. The applicability of the Standard RSF concept is scrutinised. In this context, RSF values for several Matrices (Co, Cu, Fe, In and Zn) are reported additionally.
In this work, different calibration strategies for the determination of trace elements in pure copper metal by nanosecond laser ablation ICP-MS were investigated. In addition to certified reference materials (CRMs), pellets of doped copper powder were used for calibration. The micro homogeneity of the CRMs as well as the solution-doped pellets was sufficient to use them as calibration samples in combination with a laser spot size of 200 µm. In contrast, pellets doped with analytes in solid form showed a significant heterogeneity. For most of the investigated analytes and copper CRMs the measured mass fractions were within ± 20% of their certified values when other copper CRMs were used as calibration samples. When solution-doped powder pellets were used as calibration samples a systematic trend towards mass fractions below the certified values was observed for nearly all elements determined in the analysed CRMs. Thermal fractionation effects during the ablation of the solution-doped pellets were suspected as the extent of the fractionation depends on the irradiance, whereas fractionation is reduced at higher irradiance.
Ultra high purity Zn (m6N5) was obtained through triple vacuum distillation using an m4N5 Zn as input material. High-volatile impurities were removed from the zinc matrix by vacuum evaporation, while traces having lower volatility than the matrix remained in the residual material after finishing of the entire purification process. The time required for the removal of the main high-volatile impurity (Cd) was monitored using a connected gas source mass spectrometer. During distillation the Zn vapor of the distillate was condensed on the cap of a glassy carbon crucible which was water cooled by a movable copper block. The temperatures were optimized for the removal of the volatiles and for the distillation process using vapor pressure data. High resolution glow discharge mass spectrometry was applied for the analysis and purity evaluation of the distilled and input zinc. The analysis confirmed the reduction of the metallic impurities from 43 mg kg-1 to 0.5 mg kg-1 (m6N5) after three consecutive vacuum distillations. An increase in the grain size and a decrease in the micro-hardness were observed for the purified Zn material.
Within the National Metrology Institutes (NMIs) and designated laboratories, an interlaboratory comparison, CCQM-P107, was conducted to verify the degree of international comparability concerning the results of purity analysis. The mass fractions of Ag, Bi, Cd, Cr, Ni, Tl at the lower mg/kg-level in a high purity zinc material were determined, but the real measurand in metrological sense was the sum of the six mass fractions. Homogeneity was investigated by glow discharge mass spectrometry, reference values were obtained using isotope dilution mass spectrometry. Six NMIs participated, contributing eight independent data sets. The agreement amongst the results of the participants, their median and the agreement with the reference values were usually excellent and in almost all cases below the target uncertainty of 30% relative. In this manner, the accuracy of results and the comparability between the participants was demonstrated to be established.