- This study focused on quantitative detection of cocaine on Euro banknotes in Germany. A sensitive direct competitive immunoassay was developed and optimized with a limit of detection (LOD) of 5.6 ng/L. Exhaustive cocaine extraction by solvent was tested using different methanol concentrations and buffered solutions. Crossreactivity studies were performed to determine the degree of interference of cocaine metabolites with the immunoassay. Sixty-five Euro banknotes obtained from different districts in Berlin were evaluated. A 100% contamination frequency with cocaine was detected. A comparison between the amount of cocaine extracted by cotton swabbing of one square centimeter of the banknote showed a good correlation for lower contamination levels. This assay showed high sensitivity of detecting pg of cocaine per 1 cm2 of one banknote by swabbing 1 cm2: 0, 14, and 21 pg/cm2. Moreover, three notes of different denominations revealed high cocaine concentration; 1.1 mg/note, and twice 55This study focused on quantitative detection of cocaine on Euro banknotes in Germany. A sensitive direct competitive immunoassay was developed and optimized with a limit of detection (LOD) of 5.6 ng/L. Exhaustive cocaine extraction by solvent was tested using different methanol concentrations and buffered solutions. Crossreactivity studies were performed to determine the degree of interference of cocaine metabolites with the immunoassay. Sixty-five Euro banknotes obtained from different districts in Berlin were evaluated. A 100% contamination frequency with cocaine was detected. A comparison between the amount of cocaine extracted by cotton swabbing of one square centimeter of the banknote showed a good correlation for lower contamination levels. This assay showed high sensitivity of detecting pg of cocaine per 1 cm2 of one banknote by swabbing 1 cm2: 0, 14, and 21 pg/cm2. Moreover, three notes of different denominations revealed high cocaine concentration; 1.1 mg/note, and twice 55 μg/note.…