Photoluminescence techniques are amongst the most widely used tools in the life sciences, with new and exciting applications in medical diagnostics and molecular imaging continuously emerging. Advantages include their comparative ease of use, unique sensitivity, non-invasive character, and potential for multiplexing, remote sensing, and miniaturization. General drawbacks are, however, signals, that contain unwanted wavelength- and polarization contributions from instrument-dependent effects, which are also time-dependent due to aging of instrument-components, and difficulties to measure absolute fluorescence intensities. Moreover, scattering systems require special measurement geometries and the interest in new optical reporters with emission > 1000 nm strategies for reliable measurements in the second diagnostic for the comparison of material performance and the rational design of new fluorophores with improved properties.
Here, we present strategies to versatile method-adaptable liquid and solid fluorescence standards for different fluorescence parameters including traceable instrument calibration procedures and the design of integrating sphere setups for the absolute measurement of emission spectra and quantum yields in the wavelength region of 350 to 1600 nm. Examples are multi-emitter glasses, spectral fluorescence standards, and quantum yield standards for the UV/vis/NIR
Scanning Electron Microscopy (SEM) or Transmission Electron Microscopy (TEM) ? This is a question, nowadays discussed in EM labs of research and industry involved in the characterization and metrology of nanoparticles. The Scanning principle is adapted to TEM, the Transmission mode is adapted to Ultra High Resolution SEM. Can modern SEM replace TEM, reach atomic resolution even without Cs corrector or nm lateral resolution for energydispersive X-ray spectroscopy (EDX)?
Due to the development of a New Cold Field Emission (NCFE) electron source Hitachi SEMs SU9000 and SU8200 can deliver routinely sub-nm image resolution and EDX mappings at very high count rates and a lateral EDX resolution of a few nm. A TiO2 sample provided by BAM was analysed at 30kV using low kV STEM – simultaneously with the Through-the-lens (TTL) SE detector, the Bright Field transmitted and Dark Field transmitted signals. By this method a pixel precise information of the particles surface using SE, its chemical nature using DF-STEM and its crystalline structure using Bright Field signal is given in one 40sec scan. The advantage of this observation mode compared to Ultra Low Voltage imaging is outlined.
The EC recommendation for the definition of nanomaterial [2011/696/EU] requires the quantitative size determination of constituent particles in samples down to 1 nm. Accordingly, a material is a nanomaterial if 50 % or more of the particles are in the size range 1-100 nm. The fact that engineered nanomaterials already exist in many industrial and consumer products challenges the development of measurement methods to reliably identify, characterize and quantify their occurrence as substance and in various matrices.
The EU FP7 NanoDefine project [www.nanodefine.eu] has addressed this challenge by developing a robust, readily implementable and cost-effective measurement strategy to obtain quantitative particle size distributions and to distinguish between nano and non-nano materials according to the EU definition. Based on a comprehensive evaluation of existing methodologies and intra- and inter-lab comparisons, validated measurement methods and instrument calibration procedures have been established to reliably measure the size of particles within 1-100 nm, and beyond, including different shapes, coatings and chemical compositions in industrial materials and consumer products. Case studies prove their applicability for various sectors, including food, pigments and cosmetics.
Main outcome is the establishment of an integrated tiered approach including rapid screening (tier 1) and confirmatory methods (tier 2), and a user manual to guide end-users, such as manufacturers, in selecting appropriate methods. Another main product is the “NanoDefiner” e-Tool allowing the standardised / semi-automated selection of appropriate methods for material classification according to the EU definition. Results also contribute to standardization efforts, such as CEN TC 352 or ISO TC 229.
The synthesis of TiO2 nanoplatelets with fluorine-containing reactants is carried out using titanium (IV) butoxide as precursor and concentrated HF as shape controller, the final product requires a working up in order to eliminate or at least to reduce the amount of residual fluorides, which is realized here by well-defined thermal treatment of the samples. Bulk and surface sensitive methods namely scanning electron microscopy with energydispersive X-ray spectroscopy (SEM-EDX), Auger electron spectroscopy (AES) and time-of-flight secondary ion mass spectrometry (ToF-SIMS) have been applied to trace the presence of any fluorides in dependence on different information depths and measurement sensitivities of these methods.
Parameters of TiO2 coatings can greatly influence their final performance in largescale applications such as photocatalytic measurements, orthopedic and/or dental prostheses, cell cultures, and dye-sensitized solar cells. From different film deposition procedures, self-assembly of TiO2 NPs in multiple layers was selected for systematic characterization. EDX, AES and ToF-SIMS analysis have been carried out in order to evaluate the functionalization of several types of TiO2 NPs differing in size, shape and surface area.
Although there are many experimental techniques for measuring particle sizes and size distributions, electron microscopy (EM) is still considered as the gold standard in this field, especially when it comes to particle sizes in the nanorange (1 nm – 100 nm). Furthermore, high-resolution X-ray spectroscopy (EDS) can be applied to individual nanoparticles. To be able to extract accurate information from the EM micrographs and EDS elemental maps that are representative for the material under investigation, one needs to assure the representativity of the particles as sampled on the substrate and their homogeneous spatial distribution, to avoid operator bias when selecting the imaged area. Furthermore, agglomeration should be avoided as far as possible. Several sample preparation techniques exist since a long time, the most common way being suspending the particles in a liquid and depositing them on the grid. However, this procedure includes the drying of larger solvent amounts on the substrate itself, which can affect the spatial distribution of the deposited particles. One possibility to overcome this problem is the use of an electrospray system, where the suspension of particles is sprayed onto the substrate in charged droplets that are so small that they either dry off on the substrate immediately without affecting the position of particles, or even already during their flight time to the substrate. No dedicated commercial instruments are available for the preparation of TEM grids yet, only electrostatic deposition of aerosols on TEM grids has been reported so far. To test the prototype and assess its performance, several materials have been sprayed onto TEM grids and the resulting particle distributions were compared to more traditional sample preparation strategies like the “drop on grid” method. Operation parameters such as the sample flow-rate, capillary – substrate distance, electric field strength and sampling period length have been optimised. It was found that the particles deposited by electrospray generally show a much more homogeneous spatial distribution on the substrate and a substantial increase of the number of single particles, which are much better suited to an automatic image evaluation procedure than the agglomerated particles observed otherwise. The applicability of the technique to a broad range of materials is demonstrated by various examples, but also the influence of the substrate, the choice of the particular TEM grid, on the obtained spatial particle distribution is assessed.
Introduction: Laser Induced Breakdown Spectroscopy (LIBS) is an atomic emission analytical technique, wide spreading in laboratories and industries. One way to dramatically increase its analytical results is to deposit metal NPs on the sample surface, resulting in an better version called Nanoparticle Enhanced LIBS (NELIBS). In order to better know and use this technique, the evolution of the plasma has been studied with Tomography.
Colloidal metalorganic frameworks as novel biofunctional nanoparticles for immunoassay applications
(2017)
Metal-organic framework (MOF) colloids have unique features that render them ideal signalling agents for realizing advanced immunoassay-based detection systems. MOFs are porous coordination polymers of metal nodes and organic linkers. The pore size of MOFs can be engineered and tailored to allow specific host (MOF) and guest (analyte) interactions. The particle sizes of the colloidal MOF can be tailored by employing methods from colloidal chemistry in wet synthesis. The adaption of established Layer-by-Layer polyelectrolyte coating protocols [1] allows equipping colloidal MOF particles with a nanometer thin polyelectrolyte membrane. This polyelectrolyte membrane serves as an interface for antibody binding. These biofunctional MOF nanoparticles have shown a strong immuno-binding that is sufficient for solid state immunoassays. Our current research addresses the design of luminescence encoded colloidal particle libraries by adjusting the ratios of e.g. Terbium (green) and Europium (red) metal nodes in mixed lanthanide based MOF-76. These mixed lanthanide MOF-76 particles are envisioned to allow multiplexed immuno-detection of endocrine disruptors such as bisphenol A.
In addition we investigate the detection of analytes that do not allow the production of antibodies due to their inherent properties. Such “difficult analytes” have a strong hydrophobicity or are very small or highly toxic molecules. One example is the common plasticizer dioctylphthalate that is also a potent endocrine disruptor. MOF colloids can address this issue by specific host (MOF) : guest (analyte) interactions that result in analyte-specific colour change or exciplex-based fluorescence emission. Our overall aim is to develop methodologies that allow parallel sensing of two endocrine disruptors (e.g. bisphenol A & phthalates) by simultaneous immuno-detection and MOF:analyte specific interactions.
The overall objective of project Improved traceability chain of nanoparticle size measurements is to improve the traceability chain for nanoparticle size measurements. The main impact will be achieved by manifold contributions to standard documents for CEN/TC 352 “Nanotechnologies”, which directly addresses the research needs of CEN, CENELEC and ETSI mandated by EC to develop standards for methods and reference materials to accurately measure the size and size distribution of nanoparticles. This will take place in collaboration with ISO/TC229 ‘Nanotechnologies’, ISO/TC24/SC4 ‘Particle characterization’ and ISO/TC201 ‘Surface analysis’/ SC9 ‘Scanning probe microscopy’.
The presentation goes into the characterization of TiO2 engineered nanoparticles within EU/FP7 SETNanoMetro Project and the nanomaterial classification according to the EC definition tested within EU/FP7 NanoDefine Project. Further, ISO/TC 229/JWG 2 activities related on ISO standards in development and inter-laboratory comparisons on measurement of nanoparticle size and shape distribution by SEM and TEM are discussed.