The elemental analysis of seawater is often critical to the understanding of marine chemistry, marine geochemistry, and the deep-sea ecosystems. Laser-induced breakdown spectroscopy (LIBS) with the advantage of rapid multi-elements detection, has a great potential for in-situ elemental analysis of seawater. In practice, it is crucial to create a compact, low cost and power saving instrument for the long-term deep-sea observation. A recently appeared diode-pumped solid-state (DPSS) laser seems to be a promising candidate as it is both compact and robust. Additionally, its high repetition rate up to hundreds of kHz can provide a considerable throughput for LIBS analysis. However, the DPSS lasers operate at moderate pulse energies, usually less than one mJ, which cannot sustain stable breakdowns in bulk water. To ensure stable laser-induced plasmas underwater with such a μJ-DPSS laser, we introduced an ultrasound source to assist the breakdown process. The phase interface and mass flow generated by the near-field ultrasound can greatly reduce the breakdown threshold and enhance element-specific emissions. Meanwhile, the high repetition-rate pulses can also improve the breakdown probability and generate unique emission lines originated from the water molecule. We further demonstrate that the high repetition-rate DPSS laser combined with the Echelle spectrometer can provide effective quantitative analysis for metal elements in bulk water.