Bundling analytical capacities to understand phase formation in recycling of functional materials
(2019)
Transitioning from combustion engine-driven transportation to e-mobility demands a paradigm shift – from a system geared to maximize energy efficiency (i.e. fuel consumption) to a system that may be constrained by the availability of high technology (critical) metals required for electrical energy storage systems and drives. In the wake of these developments efforts in securing new resources of these metals from recycling of end-of-life products are increasing steadily.
Recycling of Li-Ion batteries has recently been evaluated. The results pinpoint to a critical need for understanding slag Formation and its dependence on metal components like Mn under extreme conditions. This will allow researchers to predict optimal Operation setting and to react quickly to changing market demands (which may be Li or Co at one point but may also shift to Ni or rare earth elements (REE)). The long-term goal is to control the formation of specific phases in slags allowing for a Maximum yield of elements of interest and optimal recovery in the separation processes that follows. The combination of data on the physical micro structure and local chemistry of the multi-Phase products during and after processing will help to understand and derive thermodynamic and kinetic data on its formation.
In this paper we are giving an overview on the analytical challenges and approaches to provide robust data on local element concentration and species (especially Mn which is a common component of next generation Li-ion batteries cathodes), spanning the dimensions from the nanometer scale to the bulk material. The complementary interactions of X-rays and electrons make them ideal probes to collect Interface and “in-depth” information. Before- and -after studies as well as in situ structural changes and Phase (trans)formation, changes in elemental and elemental species (e.g. oxidation state) distribution may be tracked by X-ray diffraction (XRD), X-ray fluorescence microscopy and X-ray Absorption spectroscopy. The application of such advanced analytical tools will not only provide essential clues during early lab-based experiments towards the development of new recycling technologies, but may also be deployed for on-line and in-line monitoring of industrial processes.
Confocal Micro X-ray fluorescence analysis (μXRF) is a powerful method for the destruction-free investigation of the elemental composition of various samples in three dimensions. Quantitative results rely heavily on the transmission properties of polycapillary optics, used in confocal setups.
Different models for the functional correlation between energy and transmission of a polycapillary half-lens are compared with experimental data, established with a monochromatic confocal laboratory setup [1], as well as Monte-Carlo-Simulations [2]. The effect of the different approaches on quantitative results are compared using well known SRMs, ranging from thick samples to thin layers and structured materials.
Along with this the basic principles for quantitative evaluations of XRF spectra taken with confocal setups are presented, and compared with experimental results.
X-ray fluorescence imaging is a well-established tool in materials characterization. In this work, we present the adaption of coded aperture imaging to full-field X-ray fluorescence imaging at the synchrotron. Coded aperture imaging has its origins in astrophysics, and has several advantages: Coded apertures are relatively easy to fabricate, achromatic, allow a high photon throughput, and high angular acceptance. Coded aperture imaging is a two-step-process, consisting of the measurement process and a reconstruction step. Different programs have been written, for the raytracing/forward projection and the reconstruction. Experiments with coded aperture in combination with a Color X-ray Camera and an energy-dispersive area detector, have been conducted at the BAMline. Measured samples were successfully reconstructed, and gave a 9.1-fold increase in count rate compared to a polycapillary optic.
The vanadium redox flow battery (VRFB) is currently a potential candidate for stationary energy storage. A major challenge is the unintended vanadium transport through the separator, which results in a fade of capacity. To overcome this issue, it is necessary to understand the transport processes in the membrane on a more fundamental level. In this work, the vanadium species distribution in Nafion™ 117 after cyclization was investigated. Two membranes, one from a charged VRFB and another from a discharged VRFB, were analyzed using ultraviolet–visible spectroscopy (UV/VIS) and X-ray absorption near edge structure spectroscopy (XANES). Little difference between the two membranes was recognizable according to the UV/VIS results. In comparison, the XANES results showed that the membrane from the charged VRFB contains more V3+ than VO2+, whereas for the discharged case, more VO2+ is present in the membrane.
In this talk an overview about artificial intelligence/machine learning applications @BAMline is given. In the first part, the use of neural networks for the quantification of XRF measurements and the decoding of coded-aperture measurements are shown. Then it is shown how Gaussian processes and Bayesian statistics can be used to achieve an optimal alignment of the set-up and in general for optimization of measurements.
Gold is one of the seven metals already known in antiquity and was used from time immemorial as a medium of exchange and for the production of jewelry because of its luster and rarity. In addition, it is easy to work and largely resistant to chemical influences. Investigations of gold using synchrotron radiation excited X-ray fluorescence analysis are non-destructive and provide information about the chemical elements present in the sample under investigation. The investigations presented here at BAMline focus on questions such as the origin, manufacturing process, and association of gold findings. The different questions are explained by a number of examples ranging from the Viking treasure from Hiddensee to the Nebra Sky Disk and finds from Egypt. The find from Bernstorf is discussed in detail. A Bayesian treatment of the authenticity is shown.
News from the BAMline
(2022)
Our aim is to develop a simple and inexpensive method for full field X-ray fluorescence imaging.We combine an energydispersive array detector with a coded aperture to obtain high resolut ion images. To obtain the information from the recorded image a reconstruction step is necessary. The reconstruction methods we have developed, were tested on simulated data and then applied to experimental data. The first tests were carried out at the BAMline @BESSY II. This method enables the simultaneous detection of multiple elements,which is important e.g. in the field of catalysis.
Gold ist eines der sieben bereits in der Antike bekannten Metalle und wurde wegen seines Glanzes und seiner Seltenheit seit jeher als Tauschmittel und zur Herstellung von Schmuck verwendet. Außerdem ist es leicht zu bearbeiten und weitgehend resistent gegen chemische Einflüsse.
Die Analyse von Gold mit der durch Synchrotronstrahlung angeregten Röntgenfluoreszenzanalyse ist zerstörungsfrei und liefert Informationen über die in der untersuchten Probe vorhandenen chemischen Elemente.
Im Mittelpunkt der hier vorgestellten Untersuchungen an der BAMline stehen Fragen nach der Herkunft, dem Herstellungsprozess und der Zugehörigkeit von Goldfunden. Die verschiedenen Fragestellungen werden anhand einer Reihe von Beispielen erläutert, die vom Wikingerschatz von Hiddensee über die Himmelsscheibe von Nebra bis hin zu Funden aus Ägypten reichen. Darüber hinaus werden die heute am Synchrotron verfügbaren modernen Messmethoden vorgestellt.