Hepcidin-25 concentrations measured by various methods differ considerably, complicating interpretation. Here, a previously identified plasma-based candidate secondary reference material (csRM) was modified into a serum-based two-leveled sRM. We validated its functionality to increase the equivalence between methods for international standardization. We applied technical procedures developed by the International Consortium for Harmonization of Clinical Laboratory Results. The sRM, consisting of lyophilized serum with cryolyoprotectant, appeared commutable among nine different measurement procedures using 16 native human serum samples in a first round robin (RR1). Harmonization potential of the sRM was simulated in RR1 and evaluated in practice in RR2 among 11 measurement procedures using three native human plasma samples. Comprehensive purity analysis of a candidate primary RM (cpRM) was performed by state-of-the-art procedures. The sRM was value assigned with an isotope dilution mass spectrometry-based candidate reference method calibrated using the certified pRM. The inter-assay CV without harmonization was 42.1% and 52.8% in RR1 and RR2, respectively. In RR1, simulation of harmonization with sRM resulted in an inter-assay CV of 11.0%, whereas in RR2 calibration with the material resulted in an inter-assay CV of 19.1%. Both the sRM and pRM passed international homogeneity criteria and showed long-term stability. We assigned values to the low (0.95 ± 0.11 nmol/L) and middle concentration (3.75 ± 0.17 nmol/L) calibrators of the sRM. Standardization of hepcidin is possible with our sRM, which value is assigned by a pRM. We propose the implementation of this material as an international calibrator for hepcidin-25.
Hepcidin-25 was identified as the main iron regulator in the human body, and it by binds to the sole iron-exporter ferroportin. Studies showed that the N-terminus of hepcidin is responsible for this interaction, the same N-terminus that encompasses a small copper(II)-binding site known as the ATCUN (amino-terminal Cu(II)- and Ni(II)-binding) motif. Interestingly, this copper-binding property is largely ignored in most papers dealing with hepcidin-25. In this context, detailed investigations of the complex formed between hepcidin-25 and copper could reveal insight into its biological role. The present work focuses on metal-bound hepcidin-25 that can be considered the biologically active form. The first part is devoted to the reversed-phase chromatographic separation of copper-bound and copper-free hepcidin-25 achieved by applying basic mobile phases containing 0.1% ammonia. Further, mass spectrometry (tandem mass spectrometry (MS/MS), high-resolution mass spectrometry HRMS)) and nuclear magnetic resonance (NMR) spectroscopy were employed to characterize the copper-peptide. Lastly, a three-dimensional (3D)model of hepcidin-25with bound copper(II) is presented. The identification of metal complexes and potential isoforms and isomers, from which the latter usually are left undetected by mass spectrometry, led to the conclusion that complementary analytical methods are needed to characterize a peptide calibrant or reference material comprehensively. Quantitative nuclear magnetic resonance (qNMR), inductively-coupled plasma mass spectrometry (ICP-MS), ion-mobility spectrometry (IMS) and chiral amino acid analysis (AAA) should be considered among others.
Mass spectrometry-based methods play a crucial role in the quantification of the main iron metabolism regulator hepcidin by singling out the bioactive 25-residue peptide from the other naturally occurring N-truncated isoforms (hepcidin-20, -22, -24), which seem to be inactive in iron homeostasis. However, several difficulties arise in the MS analysis of hepcidin due to the sticky character of the peptide and the lack of suitable standards. Here, we propose the use of amino- and fluoro-silanized autosampler vials to reduce hepcidin interaction to laboratory glassware surfaces after testing several types of vials for the preparation of stock solutions and serum samples for isotope dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). Furthermore, we have investigated two sample preparation strategies and two chromatographic separation conditions with the aim of developing an LC-MS/MS method for the sensitive and reliable quantification of hepcidin-25 in serum samples. A chromatographic separation based on usual acidic mobile phases was compared with a novel approach involving the separation of hepcidin-25 with solvents at high pH containing 0.1% of ammonia. Both methods were applied to clinical samples in an intra-laboratory comparison of two LC-MS/MS methods using the same hepcidin-25 calibrators with good correlation of the results. Finally, we recommend an LC-MS/MS-based quantification method with a dynamic range of 0.5–40 μg/L for the assessment of hepcidin-25 in human serum that uses TFA-based mobile phases and silanized glass vials.
Hepcidin-25 has attracted much attention ever since its discovery in 2001. It is widely recognized that this peptide hormone plays a major role in the regulation of iron levels in mammals and can reveal important clinical information about several iron-related disorders. However, the development of a reliable assay to quantify hepcidin proved to be problematic and serum hepcidin-25 concentrations determined by various assays differ substantially. Challenges arise in the MS analysis of hepcidin due to the “sticky” character of the peptide and the lack of suitable standards.
With the aim to tackle the current difficulties in hepcidin quantification and improve the status of this promising biomarker in the clinical field, we developed a rapid and robust analytical strategy for the quantification of hepcidin-25 in human samples based on HPLC-MS/MS (QqQ) as a reference method candidate to be implemented in routine laboratories. The novelty of the method is the use of amino- and fluoro-silanized autosampler vials to reduce hepcidin interaction to laboratory glassware surfaces. Furthermore, we have investigated two sample preparation strategies and two chromatographic separation conditions where the use of acidic mobile phases was compared with a novel approach involving solvents at high pH containing 0.1% of ammonia. Both methods were carefully validated and applied to clinical samples in an intra-laboratory comparison of two LC-MS/MS methods using the same hepcidin-25 calibrators with very good correlation of the results.
Isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) is emerging in the field of clinical chemistry and laboratory medicine as an alternative to immunoassays and is acknowledged as the MS “gold standard” for small biomolecule quantification. Hepcidin-25, a key iron-regulatory peptide hormone discovered in 2000, has revolutionized the understanding of iron disorders and its quantitative determination in biological samples should advance the management of iron-related pathology (diagnosis, prognosis and treatment). This study applied LC-MS/MS, using the triple quadrupole (QqQ) mass spectrometer, in a rapid and robust analytical strategy for the quantification of hepcidin-25 in human serum, to be implemented in routine laboratories. For this purpose, two sample preparation strategies and two complementary chromatographic separation conditions were investigated, where the use of acidic mobile phases (0.1% trifluoroacetic acid) was compared with a novel approach involving solvents at high pH (containing 0.1% ammonia). The application of these LC-MS/MS methods to human samples in an intra-laboratory comparison, using the same hepcidin-25 calibrators, yielded a very good correlation of the results. The LC-MS/MS employing trifluoroacetic acid-based mobile phases was selected as a highly sensitive (limit of quantification LOQ of 0.5 μg/L) and precise (coefficient of variation CV<15%) method and was recommended as a reference method candidate for hepcidin-25 quantification in real samples (in the dynamic range of 0.5-40 μg/L). One of the novel aspects of the methodology was the use of amino- and fluoro-silanized autosampler vials to reduce the interaction of the 25-residue peptide to laboratory glassware surfaces. Moreover, this LC-MS/MS method was used for an international round robin study, applying a secondary reference material as a calibrator. By determining the degree of equivalence between the results of the ten participating methodologies, the performance of the method developed in this study was found to be in the optimal range as defined by the International Consortium for Harmonization of Clinical Laboratory Results (ICHCLR). In this work, the formation of hepcidin-25 complexes with copper(II) was investigated. The first reversed-phase chromatographic separation of hepcidin-25/Cu2+ and hepcidin-25 (copper “free”) was achieved by applying mobile phases containing 0.1% of ammonia (pH 11). LC-MS/MS and high-resolution mass spectrometry (Fourier-transform ion cyclotron resonance (FTICR) MS) were applied for the mass spectrometric characterization of the formed hepcidin-25-Cu(II) species at pH values of 11 and 7.4 respectively. A new species corresponding to hepcidin-25 complexed with two copper ions was identified at high pH.