Polymeric nanoparticles (NP) are of increasing importance for a wide range of applications in the material and life sciences, including their use as drug carriers, fluorescent sensors and multimodal reporters in a large variety of bioassays and bioimaging studies. Application-relevant properties of NP include their size (and size distribution), shape, optical properties, and ease of subsequent functionalization, e.g. with linkers, sensor molecules, and bioligands. In this respect, knowledge of the chemical nature, the total number of surface groups and the number of groups accessible for subsequent coupling reactions is mandatory.1 Commercially available polystyrene NP often contain different additives like stabilizers, radical starters and crosslinkers, which can influence the quantification of surface functionalities. Moreover, they often have unknown surface group densities that may vary from batch to batch, which complicates or even hampers their reliable use in many (bio)applications. To circumvent these issues, we synthesized amino- and carboxy-functionalized, monodisperse 100 nm polystyrene NP with three different, well-defined surface group densities. Using a recently developed approach for the quantification of functional groups on nano- and microparticles with cleavable reporters,2 we quantified the assessible functional groups on the self-synthesised PS particles via optical spectroscopy (photometry) and inductively coupled optical emission spectrometry (32S ICP-OES). In addition, we developed a fluorimetric approach for the quantification of surface functional groups on nanoparticles based on the labelling with luminescent lanthanide complexes (LLC). In contrast to common organic dyes, LLC are not prone to photo¬luminescence quenching arising from reabsorption or dye aggregation, and thus, enable a reliable fluorometric quantification of the assessible functional groups on NP surfaces. Moreover, lanthanide tags can be detected with high specificity and sensitivity with analytical techniques such as XPS and ICP-MS, which allow for the multimodal validation of the fluorometric quantification approach.
The overall interest in nanotoxicity, triggered by the increasing use of nanomaterials in the material and life sciences, and the synthesis of an ever increasing number of new functional nanoparticles calls for standardized test procedures and for efficient approaches to screen the potential genotoxicity of these materials. Aiming at the development of fast and easy to use, automated microscopic methods for the determination of the genotoxicity of different types of nanoparticles, we assess the potential of the fluorometric γH2AX assay for this purpose. This assay, which can be run on an automated microscopic detection system, relies on the detection of DNA double strand breaks (DSB) as a sign for genotoxicity. Here, we provide first results obtained with broadly used nanomaterials like CdSe/CdS and InP/ZnS quantum dots as well as iron oxide, gold, and polymer particles of different surface chemistry with previously tested colloidal stability and different cell lines like Hep-2 and 8E11 cells, which reveal a dependence of the genotoxicity on the chemical composition as well as the surface chemistry of these nanomaterials. These studies will be also used to establish nanomaterials as positive and negative genotoxicity controls or standards for assay performance validation for users of this fluorometric genotoxicity assay. In the future, after proper validation, this microscopic platform technology will be expanded to other typical toxicity assays.
Risk assessment of nanomaterials requires not only standardized toxicity studies but also validated methods for nanomaterial surface characterization with known uncertainties. In this context, a first bilateral interlaboratory comparison on Surface group quantification of nanomaterials is presented that assesses different reporter-free and labeling methods for the quantification of the total and accessible number of amine functionalities on commercially available silica nanoparticles that are widely used in the life sciences. The overall goal of this comparison is the identification of optimum methods as well as achievable measurement uncertainties and the comparability of the results across laboratories. We also examined the robustness and ease of implementation of the applied analytical methods and discussed method-inherent limitations. In summary, this comparison presents a first step toward the eventually required standardization of methods for surface group quantification.
Inductively coupled plasma mass spectrometry (ICP-MS) is a powerful method for the matrix-independent quantitative analysis of target elements. Developed for the use in inorganic trace analysis, ICP-MS is nowadays a valuable tool for bioanalytical questions. Especially the use of ICP-MS for quantitative proteomics by measuring heteroatoms has gained recognition in the last decade, considering that established quantification methods like organic mass spectrometry depend on labelling of the target protein or the existence of matched protein and peptide standards. The need for reliable quantification of proteins is continuously growing, but only a limited number of well-characterized and quantified protein standards are available so far. Accurately quantified, traceable protein standards are necessary to ensure comparability of measurements between laboratories, not only in basic research but also in a clinical context. One example of this is the Alzheimer’s disease biomarker tau protein. However, existing tau standards lack comparability, emphasizing the need for a well-quantified protein standard.
Therefore, we developed a method for the quantification of pure proteins via sulfur isotope dilution ICP-MS (IDMS). As sulfur is present in two amino acids, cysteine and methionine, it exists in nearly all proteins and can be used for the quantification of proteins of known stoichiometry. We employed simple offline strategies for the separation of non-protein bound sulfur species. Quantification of these contaminations by IDMS allows for correction of the protein content and enables reliable protein quantification. We report the protein mass fractions of a standard reference material and commercially available proteins determined by sulfur IDMS, including the expanded uncertainties. The developed method can be applied for the reliable and traceable quantification of pure proteins for use as in-house standards. Here, we successfully used this method for the quantification of the tau protein.
Inductively coupled plasma mass spectrometry (ICP-MS) is a powerful method for the matrix-independent quantitative analysis of target elements. Developed for the use in inorganic trace analysis, ICP-MS is nowadays emerging as a valuable tool for bioanalytical questions. Especially the use of ICP-MS for quantitative proteomics by measuring heteroatoms has gained popularity in the last decade, considering that established quantification methods like organic mass spectrometry depend on the existence of matched protein and peptide standards or labelling of the target protein. The need for reliable quantification of proteins is constantly growing, but only a limited number of well characterized and quantified protein standards are available so far. Not only in basic research, but also in a clinical context, accurately quantified, traceable protein standards are needed to ensure comparability of measurements between laboratories. One disease with a major impact on our ageing society is Alzheimer’s disease (AD), which is still challenging to diagnose. As this is also due to a lack in comparability and accuracy of existing biomarker assays, the community would greatly benefit from well quantified protein biomarker standards.
In this work, we applied isotope dilution analysis (IDA) using ICP-MS to quantify proteins of known stoichiometry via their sulfur content. Sulfur is present in two amino acids, cysteine and methionine, and hence exists in nearly all proteins. Simple strategies were employed for the detection of low molecular sulfur species to correct for sulfur contaminants and allow for reliable quantification of various proteins. We report the protein mass fractions with expanded uncertainties of a standard reference material and commercially available proteins determined by sulfur IDA. The herein developed method can be applied for the reliable and traceable quantification of pure proteins and will be used for the quantification of an AD biomarker. Our target is the tau protein, as brain load and distribution of tau is highly correlated with the clinical progression of AD.
LA-ICP-MS is increasingly used for single cell analysis in two different detection modes using either the imaging mode with subcellular resolution or alternatively single spot analysis of cells with a larger laser spot size. This study compares the analytical figures of merit of both detection modes (signal to noise, precision, accuracy, throughput), as well as ease of operation and data evaluation. Adherent 3T3 fibroblast cells were stained with two metal dyes (mDOTA-Ho, Ir-DNA-intercalator) and several dozen cells were measured using both modes. We found a ten times higher throughput for single spot analysis, which has as well a straightforward data analysis, shortening the total analysis time further. The signal to noise ratio for single spot analysis was found to be slightly better compared to the signal to noise of pixels in imaging. The mean metal intensity per single cell differed by only 10% between both modes and obtained distributions were found to show no statistically significant differences. Using matrix matched calibration based on standards spotted onto nitrocellulose membrane, we achieved detection limits (10s) of 12 fg for Ir and 30 fg for Ho and quantified 57 +/-35 fg Ir and 1192 +/- 707 fg Ho per single cell.
Compared to a conventional ICP-MS measurement of a digest of about 60000 cells, 54% of Ir content and 358% Ho content was found using quantitative LA-ICP-MS. The difference might be a consequence of the two metal dyes binding to different structures of the cell and therefore might behave differently in sample preparation for conventional and LA-ICP-MS.
Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) has been revealed as a convenient technique for trace elemental imaging in tissue sections, providing elemental 2D distribution at a quantitative level. For quantification purposes, in the last years several approaches have been proposed in the literature such as the use of CRMs or matrix matched standards. The use of Isotope Dilution (ID) for quantification by LA-ICP-MS has been also described, being mainly useful for bulk analysis but not feasible for spatial measurements so far. In this work, a quantification method based on ID analysis was developed by printing isotope-enriched inks onto kidney slices from rats treated with antitumoral Pt-based drugs using a commercial ink-jet device, in order to perform an elemental quantification in different areas from bio-images. For the ID experiments ¹⁹⁴Pt enriched platinum was used. The methodology was validated by deposition of natural Pt standard droplets with a known amount of Pt onto the surface of a control tissue, where could be quantified even 50 pg of Pt, with recoveries higher than 90%. The amount of Pt present in the whole kidney slices was quantified for cisplatin, carboplatin and oxaliplatin-treated rats. The results obtained were in accordance with those previously reported. The amount of Pt distributed between the medullar and cortical areas was also quantified, observing different behavior for the three drugs.
A brief introduction to fluorescence spectroscopy will be provided, ranging from typically measured fluorescence quantities over instrument-specific contributions to measured fluorescence signals to selected applications. In this context, an overview of the photoluminescence properties of molecular and nanoscale luminescence reporters will be given including a brief insight into their photophysics and fluorescence standards designed by division Biophotonics for the calibration and instrument performance validation of fluorescence measuring devices will be presented.
Surface functionalization of nanomaterials is nowadays at the core of many applications of functional materials in the life and material sciences. Examples range from membranes and microarrays over bead-based assays, and next generation sequencing to nanometer-sized optical reporters, nanosensors, and magnetic and optical contrast agents. Typical function-nalization steps include silanization and grafting reactions with reactive monomers to introduce functional groups like amino or carboxylic acid groups or the attachment of ligands like polyethylene glycol (PEG) molecules and biomolecules. [1-3] This enables to tune e.g., dis-persibility, hydrophilicity and biocompatibility, minimize unspecific interactions, improve biofunctionalization efficiencies, and enhance blood circulation times and allows for the use of nanomaterials as reporters in assays or the design of targeted probes for bioimaging.
At the core of all functionalization strategies are reliable and validated methods for surface group and ligand quantification that can be preferably performed with routine laboratory instrumentation, require only small amounts of substances, and are suitable for many different types of nanomaterials. [3] There is meanwhile a considerable need to make these methods traceable. We present here versatile and simple concepts for the quantification of common functional groups, ligands, and biomolecules on different types of organic and inorganic nanomaterials, using conventional and newly developed cleavable and multimodal reporters, that can be detected with optical spectroscopy. [4-7] These reporters are chosen to enable method validation with the aid of method comparisons and mass balances. Also, strategies how to make these simple assays traceable to SI units using quantitative nuclear resonance spectroscopy (qNMR) and X-ray photoelectron spectroscopy (XPS) are derived.
Inorganic nanocrystals with linear and nonlinear luminescence in the ultraviolet, visible, near infrared and short-wave infrared like semiconductor quantum dots and spectrally shifting lanthanide-based nanophosphors have meanwhile found applications in the life and material sciences ranging from optical reporters for bioimaging and sensing over security barcodes to solid state lighting and photovoltaics. These nanomaterials commonly have increasingly sophisticated core/shell particle architectures with shells of different chemical composition and thickness to minimize radiationless deactivation at the particle surface that is usually the main energy loss mechanism [1]. For lanthanide-based spectral shifters, particularly for very small nanoparticles, also surface coatings are needed which protect near-surface lanthanide ions from luminescence quenching by high energy vibrators like O-H groups and prevent the disintegration of these nanoparticles under high dilution conditions. [2,3,4]. The identification of optimum particle structures requires quantitative spectroscopic studies focusing on the key performance parameter photoluminescence quantum yield [5,6], ideally flanked by single particle studies to assess spectroscopic inhomogeneities on a particle-to-particle level for typical preparation methods [7,8], Moreover, in the case of upconversion nanoparticles with a multi-photonic and hence, excitation power density (P)-dependent luminescence, quantitative luminescence studies over a broad P range are required to identify particle architectures that are best suited for applications in fluorescence assays up to fluorescence microscopy. Here, we present methods to quantify the photoluminescence of these different types of emitters in the vis/NIR/SWIR and as function of P and demonstrate the importance of such measurements for a profound mechanistic understanding of the nonradiative deactivation pathways in semiconductor and upconversion nanocrystals of different size and particle architecture in different environments.