The mechanism of this system is based on kinetic competition. This biosensor consists of a monolithic glass column with a vast excess of immobilized hapten, which traps the fluorescently labeled antibody as long as no explosive is present. If the explosive 2,4,6-trinitrotoluene (TNT) is introduced some binding sites of the antibody will be blocked, which leads to an immediate breakthrough of the labeled protein. The fluorescence is detected by highly sensitive laser-induced fluorescence with a conventional CMOS camera. The system achieved limits of detection of approx.1 pM (1 ppt) of the fluorescent label and around 100 pM (20 ppt) of TNT. The total assay time is less than 8 minutes. A cross-reactivity test with 5000 pM solutions of pentaerythritol tetranitrate (PETN), 1,3,5-trinitroperhydro-1,3,5-triazine (RDX), and octahydro-1,3,5,7-tetranitro-1,3,5,7-tetrazocine (HMX) showed no cross reactivity.