Double entry method for the verification of data a chromatography data system receives

  • The importance of software validation increases since the need for high usability and suitability of software applications grows. In order to reduce costs and manage risk factors, more and more recommendations and rules have been established. In the field of pharmacy the vendors of so-called chromatography data systems (CDSs) had to implement the guidelines of the Code of Federal Regulations Title 21 (CFR 21) during the last few years in order to fulfill the increasing requirements. The CFR 21 part 11 deals with electronic records and signatures. This part is binding for each company in the regulated environment that wishes to create, edit and sign electronic information instead of printing them on paper. Subsection CFR 21 part 11.10(h) explains how to perform an input check for manual user entries as well as for data that will be collected from an external device. In this article we present an approach performing the double entry method on data provided by the hardware instrument in order to investigate possible influences on the raw data by the handling CDS. A software tool has been written which allows us to communicate with a high-performance liquid chromatography (HPLC) detector and acquire data from it. The communication is completely independent of a CDS which is started separately and connected to the same system. Using this configuration we made a parallel data acquisition of two instances at the same time possible. Two CDSs have been tested and for at least one of them it has been shown that a comparison of the acquired data can be done as with the double entry method for the data verification. For the second CDS we checked whether it would be applicable after a few modifications. The given approach could be either used for a live data verification of produced raw data or as a single test during a software operational qualification to verify the data acquisition functionality of the software.

Download full text files

Export metadata

Additional Services

Search Google Scholar Statistics
Metadaten
Author:David Thomas Marehn, Detlef Wilhelm, Heike Pospisil, Roberto Pizzoferrato
URN:urn:nbn:de:kobv:526-opus4-12349
DOI:https://doi.org/10.5194/jsss-8-207-2019
ISSN:2194-878X
Parent Title (English):Journal of Sensors and Sensor Systems
Document Type:Article
Language:English
Year of Publication:2019
Publishing Institution:Technische Hochschule Wildau
Release Date:2019/06/06
Volume:8
First Page:207
Last Page:214
Source:Marehn, D. T., Wilhelm, D., Pospisil, H., and Pizzoferrato, R.: Double entry method for the verification of data a chromatography data system receives, J. Sens. Sens. Syst., 8, 207-214, https://doi.org/10.5194/jsss-8-207-2019, 2019.
Faculties an central facilities:Fachbereich Ingenieur- und Naturwissenschaften
Dewey Decimal Classification:5 Naturwissenschaften und Mathematik / 57 Biowissenschaften; Biologie / 570 Biowissenschaften; Biologie
Licence (German):Creative Commons - CC BY - Namensnennung 4.0 International
Verstanden ✔
Diese Webseite verwendet technisch erforderliche Session-Cookies. Durch die weitere Nutzung der Webseite stimmen Sie diesem zu. Unsere Datenschutzerklärung finden Sie hier.