We designed and synthesized triphenylamine based and coumarin fused rhodamine hybrid dyes and characterized using 1H, 13C NMR and HR-LCMS analysis. Both the newly synthesized hybrid dyes were found to show red shifted absorption as well as emissions and large Stokes shift (40e68 nm) as compared to the small Stokes shift (25e30 nm) of reported dyes Rhodamine B and 101. Photophysical properties of these dyes were studied in different solvents and according to the solvents acidity or basicity they preferred to remain in their spirocyclic or open form in different ratio. We studied the spirocyclic as well as open form derivatives of these dyes for their viscosity sensitivity in three different mixture of solvents i.e. polar-protic [EtOH-PEG 400], polar-aprotic [toluene-PEG 400] and non-polaraprotic [toluene-paraffin]. They are found to show very high viscosity sensitivity in polar-protic mixture of solvents [EtOH-PEG 400] and hence concluded that both polarity as well as viscosity factor worked together for the higher emission enhancement rather than only viscosity factor. As these dyes showed very high viscosity sensitivity in their spirocyclic as well as open form, they can be utilized as viscosity sensors in visible as well as deep red region. We also correlated our experimental finding theoretically by using Density Functional theory computations.
DNA origami nanostructures provide a platform where dye molecules can be arranged with nanoscale accuracy allowing to assemble multiple fluorophores without dye−dye aggregation. Aiming to develop a bright and sensitive ratiometric sensor system, we systematically studied the optical properties of nanoarrays of dyes built on DNA Origami platforms using a DNA template that provides a high versatility of label choice at minimum cost. The dyes are arranged at distances, at which they efficiently interact by Förster resonance energy transfer (FRET). To optimize array brightness, the FRET efficiencies between the donor fluorescein (FAM) and the acceptor cyanine 3 were determined for different sizes of the array and for different arrangements of the dye molecules within the array. By utilizing nanoarrays providing optimum FRET efficiency and brightness, we subsequently designed a ratiometric pH nanosensor using coumarin 343 as a pH-inert FRET donor and FAM as a pHresponsive acceptor. Our results indicate that the sensitivity of a ratiometric sensor can be improved simply by arranging the dyes into a well-defined array. The dyes used here can be easily replaced by other analyte-responsive dyes, demonstrating the huge potential of DNA nanotechnology for light harvesting, signal enhancement, and sensing schemes in life sciences.
Optical spectroscopy – Techniques, instrumentation, and typical molecular and nanoscale reporters
(2018)
Different types of optical spectroscopies are introduced with special emphasis on method-inherent limitations and reliable instrument calibration and performance validation. In addition, different classes of molecular and nanocrystalline emitters are presented and the underlying photophysical processes are briefly described.
Mandatory for the comparison of different emitter classes and the rational design of the next generation of molecular and nanoscale reporters are reliable and quantitative photoluminescence measurements. This is of special relevance for all fluorescence applications in the life and material sciences. In the following, procedures for the determination of this spectroscopic key parameter are presented including pitfalls and achievable uncertainties and material-specific effects related to certain emitter classes are addressed.
Organic/inorganic mixtures were prepared from ordinary Portland cement (OPC), water (w/c 0.22), a fluorescent dye in aqueous solution (stable at alkaline pH; BAM-I), and two different comb shape polycarboxylates (PCEs), i.e., high charge (PCE-HC) and low charge (PCE-LC), respectively. Rheological and calorimetric measurements were performed prior to optical studies in order to select PCE concentrations. Absorption and fluorescence spectroscopy of the system OPC + BAM-I (CBAM-I) revealed maxima of dye BAM-I located at 645 nm and 663 nm, respectively. In presence of PCE-HC and PCE-LC, these mixtures displayed a small red shift in reflectance and a faster decrease in intensity compared to studies with CBAM-I; however, only slight differences were observed between the different PCEs. With time, all systems exhibited a decrease in intensity of BAM-I in absorption/reflectance and emission. This could be caused by dye adsorption and possibly decomposition when in contact with cement particles or hydration products.
Bioanalytical, diagnostic, and security applications require the fast and sensitive determination of a steadily increasing number of analytes or events in parallel in a broad variety of detection formats.[1,2] Ideal candidates for spectral encoding and multiplexing schemes are luminescent nanocrystals like semiconductor quantum dots (QDs), particularly Cd-containing II/VI QDs with their narrow and symmetric emission bands. With the availability of relatively simple and inexpensive instrumentation for time-resolved fluorescence measurements, similar strategies utilizing the compound-specific parameter fluorescence lifetime or fluorescence decay kinetics become increasingly attractive.[3-5] The potential of different types of QDs like II/VI, III/V and Cd-free ternary QDs such as AgInS (AIS) QDs for lifetime-based encoding and multiplexing has been, however, barely utilized, although the lifetimes of these nanocrystals cover a time windows which is barely accessible with other fluorophores. Here we present a brief insight into the photophysics of AIS QDs and show the potential of dye- and QD-encoded beads for lifetime-based encoding and detection schemes in conjunction with flow cytometry and fluorescence lifetime imaging microscopy
Singlet oxygen can severely damage biological tissue, which is exploited in photodynamic therapy (PDT). In PDT, the effective range is limited by the distribution of the photosensitizer (PS) and the illuminated area. However, no distinction is made between healthy and pathological tissue, which can cause undesired damage. This encouraged us to exploit the more acidic pH of cancerous tissue and design pH-controllable singlet oxygen-generating boron-dipyrromethene (BODIPY) dyes. A pH sensitivity of the dyes is achieved by the introduction of an electronically decoupled, photoinduced electron transfer (PET)-capable subunit in meso-position of the BODIPY core. To favor triplet-state formation as required for singlet Oxygen generation, iodine substituents were introduced at the chromophore core. The resulting pH-controlled singlet oxygen-generating dyes with pKa values in the physiological range were subsequently assessed regarding their potential as pH-controlled PS for PDT.
Using HeLa cells, we could successfully demonstrate markedly different pH-dependent cytotoxicities upon illumination.
The photoluminescence quantum yield characterizes the performance of emitters for applications in optical devices, as reporters or probes in material and analytical sciences, and for sensing applications. Quantum yield measurements are challenging for luminescent molecules and nanocrystals immobilized in thin films for many sensor applications, particularly if spatially resolved quantitative luminescence information is desired. We show here that a Drexhage-type experiment, where a silver-coated millimeter-sized sphere is used to modify the local density of states, can provide an elegant approach to counter this challenge. As a representative example of the potential of this method, we measure the pH-dependent photoluminescence quantum yield of fluorescein isothiocyanate bound to a thin mesoporous silica film. The results were compared with those of the studies on the pH dependence of the same dye in solution. We found that our approach can link single fluorophore studies to ensemble measurements and pave the way for the spatially resolved fluorescence measurements of ultralow concentrations of emitters utilized as optically active elements and reporters in thin sensor films or incorporated into membranes.
The comparison of different emitter classes and the rational design of the next generation of molecular and nanoscale reporters require accurate and quantitative photo-luminescence measurements. This is of special importance for all photoluminescence applications in the life and material sciences and nanobiophotonics. In the following, procedures for the determination of the spectroscopic key parameter photoluminescence quantum yield, i.e., the number of emitted per absorbed photons, in the UV/vis/NIR/SWIR are presented including pitfalls and achievable uncertainties and material-specific effects related to certain emitter classes are addressed.
In summary, by combining the NIR-emissive [Cr(ddpd)2][BPh4]3 complex CrBPh4 with its extremely large energy gap between the longest wavelength absorption and emission maxima with a pH-responsive fluorescein derivative (FITC) and an inert reference dye like Nile Red (NR) and 5,10,15,20tetrakis(pentafluorophenyl) porphyrin (TFPP), we developed nanosensors for simultaneously sensing temperature, O2 partial pressure, and pH. These novel TOP nanosensors (temperature, oxygen, pH) cover the biologically and physiologically relevant concentration ranges of these parameters/analytes with single wavelength excitation in PBS buffer and in a cell culture medium containing bovine serum albumin (BSA). The response of both nanosensors to all parameters is fully reversible and only minimally affected by the presence of BSA, the most common serum albumin. Moreover, comparative studies with nanosensors containing only a single type of stimuli-responsive molecule and with the respective molecular systems revealed that the different sensor components do not interfere with each other. Future research will include the testing of these nanosensors in cellular uptake studies and, after surface modification with targeted bioligands, eventually in in vivo experiments as previously done by some of us with other polystyrene nanoparticle reporters and nanosensors.46,47 Moreover, this concept of multianalyte sensing will be expanded to nanosensors derived from differently sized premanufactured biocompatible polymer particles and different stimuli-responsive dyes like fluorescent indicators for biologically and bioanalytically relevant metal ions.