6.6 Physik und chemische Analytik der Polymere
Filtern
Dokumenttyp
- Zeitschriftenartikel (12)
- Vortrag (7)
- Posterpräsentation (5)
- Sonstiges (1)
Schlagworte
- Microplastic (8)
- Microplastics (8)
- TED-GC/MS (8)
- Mikroplastik (7)
- TED-GC-MS (7)
- Analysis (2)
- Internal Standard (2)
- Thermoanalytics (2)
- Thermoanalytik (2)
- Tire wear (2)
Organisationseinheit der BAM
- 6.6 Physik und chemische Analytik der Polymere (25) (entfernen)
Eingeladener Vortrag
- nein (7)
Introduction to microplastics, definition, relevance, analytical approaches & challenges. TED-GC/MS working principle, features & performance. Optimisation in TED-GC/MS methodology, new validation data. Replacement of Internal Standard, outlook towards upcoming publication and application within EU´s revised drinking water directive.
A fast method for microplastic detection is thermal extraction desorption‐gas chromatography/mass spectrometry (TED‐GC/MS), which uses polymer‐specific thermal decomposition products as marker compounds to determine polymer mass contents in environmental samples. So far, matrix impacts of different environmental matrices on TED‐GC/MS performance had not yet been assessed systematically. Therefore, three solid freshwater matrices representing different aquatic bodies with varying organic matter contents were spiked with a total of eight polymers. Additionally, for the first time, the two biodegradable polymers polybutylene adipate terephthalate (PBAT) and polylactide (PLA) were analysed using TED‐GC/MS. The methodological focus of this work was on detectability, quality of signal formation as well as realisation of quantification procedures and determination of the limit of detection (LOD) values. Overall, TED‐GC/MS allowed the unambiguous detection of the environmentally most relevant polymers analysed, even at low mass contents: 0.02 wt% for polystyrene (PS), 0.04 wt% for the tyre component styrene butadiene rubber (SBR) and 0.2 wt% for polypropylene (PP), polyethylene (PE) and PBAT. Further, all obtained LOD values were increased in all matrices compared to the neat polymer without matrix. The LOD of the standard polymers were increased similarly (PS: 0.21–0.34 μg, SBR: 0.27–0.38 μg, PP: 0.32–0.36 μg, PMMA: 0.64–1.30 μg, PET: 0.90–1.37 μg, PE: 3.80–6.99 μg) and their decompositions by radical scission processes were not significantly influenced by the matrices. In contrast, matrixspecific LOD increases of both biodegradable polymers PBAT (LOD: 1.41–7.18 μg) and PLA (0.84–20.46 μg) were observed, probably due to their hetero‐functional character and interactions with the matrices. In conclusion, the TED‐GC/MS performance is not solely determined by the type of the polymers but also by the composition of the matrix.
In this study, carried out within the Joint Danube Survey 4, a comprehensive microplastic screening in the water column within a large European river basin from its source to estuary, including major tributaries, was realized. The objective was to develop principles of a systematic and practicable microplastic monitoring strategy using sedimentation boxes for collection of suspended particulate matter followed by its subsequent analysis using thermal extraction desorption-gas chromatography/mass spectrometry. In total, 18 sampling sites in the Danube River Basin were investigated. The obtained suspended particulate matter samples were subdivided into the fractions of >100 μm and <100 μm and subsequently analyzed for microplastic mass contents. The results showed that microplastics were detected in all samples, with polyethylene being the predominant polymer with maximum contents of 22.24 μg/mg, 3.23 μg/mg for polystyrene, 1.03 μg/mg for styrene-butadiene-rubber, and 0.45 μg/mg for polypropylene. Further, polymers such as different sorts of polyester, polyacrylates, polylactide, and natural rubber were not detected or below the detection limit. Additional investigations on possible interference of polyethylene signals by algae-derived fatty acids were assessed. In the context of targeted monitoring, repeated measurements provide more certainty in the interpretation of the results for the individual sites. Nevertheless, it can be stated that the chosen approach using an integrative sampling and determination of total plastic content proved to be successful.
Die Thermo Extraktion Desorption Gas Chromatograpie Massen Spektrometrie (TED-GC/MS) ist ein schnelles thermoanalytisches Verfahren zur Bestimmung von Gehalten an Mikroplastik (MP). Die Kombination aus teils schwerflüchtigen Analyten und Kältebrücken führt zu Memory-Effekten, denen u.a. durch häufige Blindwertmessungen begegnet werden muss. Reduzierte Probedurchsätze sind die Folge. Um Messergebnisse abzusichern und den Gerätestatus bei gleichzeitig hoher Messleistung überwachen zu können, wurde die sequentielle Standardmethode überarbeitet.
An investigation of microplastic (MP) occurrence in a municipal wastewater treatment plant (WWTP) effluent with tertiary treatment was carried out. Representative sample volumes of 1 m3 were taken by applying a fractionated filtration method (500, 100, and 50 µm mesh sizes). The detection of MP mass fractions by thermal extraction desorption–gas chromatography/mass spectrometry (TED-GC/MS) was achieved without the previously required additional sample pretreatment for the first time. Different types of quantification methods for the evaluation of TED-GC/MS data were tested, and their accuracy and feasibility have been proven for real samples. Polyethylene, polystyrene, and polypropylene were identified in effluent samples. The polymer mass content varied significantly between 5 and 50 mg m−3. A correlation between the MP load and the quantity of suspended matter in the WWTP effluents, particle size distribution, particle type, and operation day (i.e., weekday, season, and capacity) was not found. It can be concluded that a meaningful assessment of WWTPs requires a comprehensive sampling campaign with varying operation conditions.
Die große Herausforderung in der Detektion von Mikroplastik (MP)‑Partikeln (1-1000 µm) besteht darin, sie neben Millionen von natürlichen Partikeln in Umweltproben nachzuweisen. Zum Aufspüren von Senken, Eintragspfaden und für Monitoring-Kampagnen ist ein schnelles, automatisiertes Verfahren unerlässlich. Ein solches Verfahren wurde an der Bundesanstalt für Materialforschung und -prüfung (BAM) mit der ThermoExtraktion/Desorption-Gaschromatographie-Massenspektrometrie (TED-GC/MS) in den letzten Jahren entwickelt.
Deuterated polystyrene (D-PS) improves microplastic detection with TED-GC / MS, among other things through higher reproducibility. It is inexpensive (0.1 ct / analysis) and can be dosed quickly and reproducibly thanks to its good solubility (e.g. in toluene). However, certain matrices lead to an exchange of the deuterium with hydrogen. The poster shows the measuring principle, in which matrices H/D-exchange takes place, the results of kinetic studies and ways of preventing or circumventing the H/D-exchange.
Fast quantitative detection of microplastics using TED-GC/MS, an innovative thermoanalytical method
(2021)
In the talk, the BAM is introduced and the motivation for our engagement in the field of microplastic analysis is explained. The challenges in the analysis of microplastics are shown and how we deal with them. The working principle of TED-GC/MS and its performance parameters are presented, as well as the procedure for identifying and quantifying polymers in environmental samples. Interesting results are shown using selected examples. Finally, BAM's microplastic reference materials are highlighted.
Deuteriertes Polystyrol (D-PS) verbessert die Mikroplastikdetektion mit der TED-GC/MS, u.a. durch höhere Reproduzierbarkeit. Es ist kostengünstig (0,1 ct/Analyse) und durch die gute Löslichkeit (z.B. in Toluol) schnell und reproduzierbar zu dosieren. Bestimmte Matrices führen jedoch zu einem Austausch des Deuteriums durchWasserstoff. Das Poster zeigt das Messprinzip, bei welchen Matrices H/D-Austausch erfolgt, Ergebnisse von Kinetikuntersuchungen sowie Möglichkeiten, den Austausch zu verhindern oder zu umgehen.
There is growing interest in quantifying microplastics in environmental samples. This application note presents a thermal extraction desorption-gas chromatography/mass spectrometry (TED-GC/MS) method that is well suited to automation and increased sample throughput. The method is also able to detect all particle sizes in the sample as long as the limit of detection (LOD) is reached and allows analysis of larger samples of 15 to 25 mg or more. Samples were decomposed by thermogravimetric analysis (TGA), and the gaseous decomposition products were trapped on a solid-phase sorbent, followed by thermal desorption‑gas chromatography/mass spectrometry (TD-GC/MS) using an Agilent 5977B GC/MSD coupled to an Agilent 7890B GC. Target microplastic particle (MP) polymers were identified in environmental samples including surface water, finished compost, house dust, and drinking water. Quantification of MP polymers in environmental samples provided LODs of 0.06 to 2.2 μg, allowing the detection of MPs in trace amounts with sample weights of up to 1 g. Method repeatability was adequate for reliable quantification with RSDs of approximately 6 to 12%.