Filtern
Erscheinungsjahr
- 2016 (2) (entfernen)
Dokumenttyp
- Zeitschriftenartikel (1)
- Sonstiges (1)
Sprache
- Englisch (2)
Schlagworte
- EDX (2) (entfernen)
Ionic liquids (ILs) are proposed as simple and efficient test materials to evaluate the performance of energy dispersive X-ray spectrometers (EDS) in the low energy range below 1 keV. By only one measurement, C Kα, N Kα, O Kα, and F Kα X-ray lines can be excited. Additionally, the S Kα line at 2.3 keV and, particularly, the S L series at 149 eV complete the picture with X-ray lines offered by the selected ILs. The well-known (certifiable) elemental composition of the ILs selected in the present study can be used to check the accuracy of results produced with the available EDS quantification routines in the low energy range, simultaneously, for several low atomic number elements. A comparison with other reference materials in use for testing the performance of EDS in the low energy range is included.
Materials and chemical producers require detailed knowledge of surface chemistry for research into new products. One way to understand a surface without damaging it is to bombard it with an electron beam, causing its atoms to emit characteristic X-rays enabling identification. The measurement of these must be precise as many elements emissions are close in energy – traceable reference materials will ensure instruments using this technique are stable and accurate.