A procedure for the quantification of the PCB marker congeners 28, 52, 101, 138, 153, 180 in plastic waste form the recycling of cables in the range of content between 0.1 mg kg-1 and 20 mg kg-1 was validated. Representative reference materials for laboratory analyses were produced using granulates from different recycling processes. The method development included aspects of grinding, sieving, homogenising as well as the comparison of extraction methods and chromatographic conditions. The validated procedure foresees grinding of the technical sample to an upper particle size limit of 0.50 mm followed by extraction with an unpolar solvent such as n-hexane and gas chromatographic determination. LOD and LOQ for individual congeners ranged between 0.01 mg kg-1 and 0.08 mg kg-1. Electron capture detection (ECD) and mass selective detection (MS) were equivalent if chromatographic conditions were appropriately selected. The application of MS was advantageous in presence of significant amounts of interfering components which have been found in some cables.
Three reference materials containing the marker congeners in the range of 0.1 mg kg-15 mg kg-1 were submitted to a final validation intercomparison with 23 selected field laboratories using the developed method. Relative reproducibility standard deviations varied between 15% and 33%.
ISO Guide 35 deals with RM stability issues and scrutinizes the evaluation of stability testing results under the assumption that either there is no trend at all (a rather rare situation), or any observed deterministic change is insignificant and thus can be neglected. However, market demands for reliable reference materials are obviously not limited to stable or at least seemingly stable materials. In many analytical applications, analytes and measurands under consideration are known, or at least suspected, to be unstable on time scales that may vary widely from measurand to measurand. The Federal Institute for Materials Research and Testing (BAM) has developed (and successfully uses) an integrated approach in its certification practice. The approach is based on an initial stability study and subsequent post-certification monitoring. Data evaluation is model-based and takes advantage of all information collected in the stability testing scheme(s). It thus allows one to deal with any kind of instability observed, to assess limiting time intervals at any stress condition in the range tested, to estimate a final expiry date for materials with detected instabilities or the maximum admissible re-testing interval for seemingly stable materials, and to assess maximum admissible stress loads during delivery of the material to the customer. The article describes (and exemplifies) typical study layout, the model selection, and the integrated data assessment.