The aim of this article is to illustrate the need for an improved quality assurance in fluorescence microscopy. From the instrument-side, this can be achieved by a better understanding, consideration, and regular control of the instrument-specific parameters and quantities affecting measured fluorescence signals. Particularly, the need for requirements on physical- and chemical-type instrument standards for the characterization and performance validation of spectral fluorescence microscopes (SFMs) is discussed and suitable systems are presented. Special emphasis is given to spectral fluorescence standards and to day-to-day intensity standards for SFMs. Fluorescence standards and well-characterized fluorescence microscopes are the first and essential steps towards the comparability and the understanding of the variability in fluorescence microscopy data in medical and life sciences. In addition, standards enable the distinction between instrument-specific variations and fluorescent label- or probe-related uncertainties as well as generally sample-related effects.
This report summarizes problems associated with the comparability of measurements of photoluminescence and procedures for the characterization of relevant instruments, focusing on physical and chemical fluorescence standards. To provide recommendations on selecting and using such standards, we derive general and scope-specific requirements and quality criteria for suitable devices and materials and briefly address metrological requirements linked to the realization of comparable measurements. Special emphasis is dedicated to liquid and solid chromophore-based fluorescence standards developed or currently tested by us.