Filtern
Dokumenttyp
- Zeitschriftenartikel (5)
- Vortrag (2)
Sprache
- Englisch (7)
Schlagworte
- GDMS (3)
- Atomic Force Microscopy (2)
- AFM (1)
- Aluminium (1)
- CaCl emission bands (1)
- CdSe/ZnS quantum Dots (1)
- Cell (1)
- Cellular uptake (1)
- Chlorine determination (1)
- Depht profiling (1)
Organisationseinheit der BAM
Eingeladener Vortrag
- nein (2)
The formation of diffusion layers on Zn layers deposited on Al substrates is mainly used to prevent corrosion effects. Evaluation of the influence exerted by different coating methodologies and heat treatments on the formation of these diffusion layers is of great interest for the aluminium industry. Particularly, multi-elemental in-depth distributions of major, minor and trace elements in Zn-coatings is highly demanded before and after heat treatments. A fast characterization of these materials require a direct solid analytical technique able to provide high sensitivity and high depth resolution. For this purpose, an improved analytical method based on the use of pulsed glow discharge sector field mass spectrometry (pulsed-GD-SFMS) is investigated. Glow discharge operating parameters (e.g. pulse duration, pulse frequency), glow discharge source design (e.g. flow tube lengths), and SFMS mass spectra acquisition conditions (e.g. integration time) are evaluated to achieve low sputtering rates, high mass spectra acquisition rates and improved depth resolution. At the optimize conditions Zn coatings deposited by arc-spray and electrodeposition are analysed before and after heat treatments to evaluate the diffusion of different key elements. Moreover, results are validated using femtosecond laser ablation (fs-LA)-ICP-MS, which provides additional information about the heterogeneous distribution of some elements in the Zn coatings.
The analytical potential of a nanosecond laser ablation inductively coupled plasma mass spectrometer (ns-LA-ICP-SFMS) system, equipped with an ultra-fast wash-out ablation chamber, is critically investigated for fast and highly spatially resolved (∼μm) qualitative elemental distribution within single cells. Initially, a low surface roughness (< 10 nm) thin In–SnO2 layer (total coating thickness ∼200 nm) deposited on glass is employed to investigate the size, morphology and overlapping of laser-induced craters obtained at different laser repetition rates, making use of Atomic Force Microscopy (AFM). Conical craters with a surface diameter of about 2 µm and depths of about 100 nm were measured after a single laser shot. Furthermore, the influence of the sampling distance (i.e. distance between the sample surface and the inner sniffer of the ablation chamber) on the LA-ICP-MS ion signal wash-out time is evaluated. A significant decrease of the transient 120Sn+ ion signal is noticed after slight variations (±200 μm) around the optimum sampling position. Ultra-fast wash-outs (< 10 ms) are achieved reducing the aerosol mixing from consecutive laser shots even when operating the laser at high repetition rates (25 – 100 Hz). Fast and highly spatially resolved images of elemental distribution within mouse embryonic fibroblast cells (NIH/3T3 fibroblast cells) and human cervical carcinoma cells (HeLa cells), incubated with gold nanoparticles (Au NPs) and Cd-based quantum dots (QDs), respectively, are determined at the optimized operating conditions. Elemental distribution of Au and Cd in single cells is achieved using a high scanning speed (50 µm/s) and high repetition rate (100 Hz). The results obtained for the distribution of fluorescent Cd-based QDs within the HeLa cells are in good agreement with those obtained by confocal microscopy. The size, morphology and overlapping of laser-induced craters in the fixed cells are also investigated using AFM, observing conical craters with a surface diameter of about 2.5 µm and depths of about 800 nm after a single laser shot.
The analytical potential of a nanosecond laser ablation coupled plasma mass spectrometer (ns-LA-ICP-SFMS)system is investigated for fast and highly spatially resolved (~µm) elemental distribution within single cells. The size, morphology and overlapping of laser-induced craters has been investigated with Atomic Force Microscopy (AFM).
High spatially resolved quantitative bioimaging of CdSe/ZnS Quantum Dots uptake in two kinds of cells is investigated combining laser ablation inductively coupled plasma mass spectrometry and the spatially resolved analysis of dried pL-droplets from a solution with a known concentration of Quantum Dots. Single cells and dried pL-droplets are morphologically characterized by Atomic Force Microscopy. A number concentration of CdSe/ZnS QDs between 3.5 104 and 48 104 is estimated to be uptaken by several selected single cells, after being incubated in the presence of a QDs suspension added to a standard cell culture medium. Mono-elemental bioimaging at subcellular resolution seems to show a higher number concentration of the CdSe/ZnS QDs in the cytosol around the cell nucleus.
The entire process involving the determination of Cl by molecular emission detection in Laser-Induced Breakdown Spectroscopy (LIBS) is thoroughly studied in this paper. This critical evaluation considers how spectra are normalized, how interferences from other molecular species signals are removed, and how signal integration is applied. Moreover, a data treatment protocol is proposed to achieve reliable and accurate Cl determination from the CaCl molecular spectral signal, not requiring the use of more complex numerical approaches. Calcium chloride dihydrate (CaCl2⋅2H2O) and high purity anhydrite samples (CaSO4) are used to optimize the acquisition conditions and data treatment of CaCl emission signal. Using the developed protocol, calibration curves for Cl, covering the concentration range from 0 μg/g to 60,000 μg/g of Cl, are successfully achieved. Finally, the suitability of the proposed methodology for Cl determination is successfully applied in industrial gypsum waste samples, where the results obtained by LIBS are validated using high-resolution molecular absorption spectroscopy (HR-CS-MAS) and potentiometric titration.
Innovative material with outstanding physical and chemical properties are produced upon the deposition of thin and ultrathin coatings on different substrates. For instance, hard disks able to store Tera-bytes of information are based on the deposition of multiple nano-layers, which include magnetic and non-magnetic layers. Many other analogous examples, including photovoltaic cells, coated Al substrates, oxidized thin film composite membranes, coated glasses, coated polymers, etc. could be cited. The physical performance of these emerging materials is directly related to their chemical properties, including the elemental distribution within the different layers and at the layer-interfaces, or the presence of critical non-desired trace elements. Therefore, quality control and R+D advances require the development of direct solid analytical techniques able to provide fast multi-elemental chemical analysis of these materials, with high depth resolution (to monitor the different layers) and high sensitivity (to detect major, minor and trace elements).
Atomic spectrometry techniques have long been used for direct elemental chemical characterization. Techniques, such as Secondary Ion Mass Spectrometry or Auger Electron Spectroscopy provide very valuable information about the chemical composition of the surfaces/coatings; however, they also have some major drawbacks, such as high operating costs, complex sample pre-treatment and handling, low sample throughput and/or severe matrix effects that result in difficult quantification procedures. To overcome some of these drawbacks Glow Discharge Mass Spectroscopy is proposed as a complementary methodology that provides an ideal solution for fast and accurate bulk and layer analyses. In this work, we evaluate the advantages and limitations of this technique and we discuss about recent progresses and new applications.
A plethora of innovative materials are produced by depositing thin and ultrathin coatings on different substrates. For instance, hard disks able to store terabytes of information are based on multiple magnetic and non-magnetic nano-layers. Moreover, protectives coatings might be used to enhance the mechanical properties of substrate materials; for instance Zn-based coatings are used in the Al industry.
In this context, solid analytical techniques are required to provide fast multi-elemental chemical analyses with high depth resolution (to monitor the different layers) and high sensitivity (to detect major, minor and trace elements). Direct solid analytical techniques, including Secondary Ion Mass Spectrometry, Secondary Neutral Mass Spectrometry, Auger Electron Spectroscopy or X-Ray Photoelectron Spectroscopy provide very valuable information about the atomic chemical composition of the surfaces/coatings; however, they also have some major drawbacks, such as high operating costs, complex sample pre-treatment and handling, low sample throughput and/or severe matrix effects that result in difficult quantification procedures. In order to overcome some of these drawbacks Glow Discharge Mass Spectroscopy (GD-MS) is proposed as a complementary methodology that provides an ideal solution for fast and accurate bulk and layer analyses. In this work, we evaluate the advantages and limitations of this technique and we discuss about recent progresses and new applications.