The vanadium redox flow battery (VRFB) is currently a potential candidate for stationary energy storage. A major challenge is the unintended vanadium transport through the separator, which results in a fade of capacity. To overcome this issue, it is necessary to understand the transport processes in the membrane on a more fundamental level. In this work, the vanadium species distribution in Nafion™ 117 after cyclization was investigated. Two membranes, one from a charged VRFB and another from a discharged VRFB, were analyzed using ultraviolet–visible spectroscopy (UV/VIS) and X-ray absorption near edge structure spectroscopy (XANES). Little difference between the two membranes was recognizable according to the UV/VIS results. In comparison, the XANES results showed that the membrane from the charged VRFB contains more V3+ than VO2+, whereas for the discharged case, more VO2+ is present in the membrane.
A core component of energy storage systems like vanadium redox flow batteries (VRFB) is the polymer electrolyte membrane (PEM). In this work, the frequently used perfluorosulfonic-acid (PFSA) membrane Nafion™ 117 and a novel poly (vinylidene difluoride) (PVDF)-based Membrane are investigated. A well-known problem in VRFBs is the vanadium permeation through the membrane. The consequence of this so-called vanadium crossover is a severe loss of capacity. For a
better understanding of vanadium transport in membranes, the uptake of vanadium ions from electrolytes containing Vdimer(IV–V) and for comparison also V(II), V(III), V(IV), and V(V) by both membranes was studied. UV/VIS spectroscopy, X-ray absorption near edge structure spectroscopy (XANES), total reflection X-ray fluorescence spectroscopy (TXRF), inductively coupled plasma optical emission spectrometry (ICP-OES), and micro X-ray fluorescence spectroscopy (microXRF) were used to determine the vanadium concentrations and the species inside the membrane. The results strongly support that Vdimer(IV–V), a dimer formed from V(IV) and V(V), enters the nanoscopic water-body of Nafion™ 117 as such. This is interesting, because as of now, only the individual ions V(IV) and V(V) were considered to be transported through the membrane. Additionally, it was found that the Vdimer(IV–V) dimer partly dissociates to the individual ions in the novel PVDF-based membrane.
The Vdimer(IV–V) dimer concentration in Nafion™ was determined and compared to those of the other species. After three days of equilibration time, the concentration of the dimer is the lowest compared to the monomeric vanadium species. The concentration of vanadium in terms of the relative uptake λ = n(V)/n(SO3
) are as follows: V(II) [λ = 0.155] > V(III) [λ = 0.137] > V(IV) [λ = 0.124] > V(V) [λ = 0.053] > Vdimer(IV–V) [λ = 0.039]. The results show that the Vdimer(IV–V) dimer Needs to be considered in addition to the other monomeric species to properly describe the transport of vanadium through Nafion™ in VRFBs.
A major source of capacity fade of the common vanadium redox flow battery (VRFB) is the vanadium ion transport through the separator. However, different transport models disagree significantly in the diffusion coefficient for the different V species and the influence of different transport mechanisms. The underlying hypothesis of this work is that reactions inside the membrane are partly responsible for these discrepancies. Accordingly, it was investigated if redox reactions inside the nanoscopic water body of Nafion 117 can occur. X-ray absorption near edge structure spectroscopy (XANES) was used to distinguish between the different V species inside hydrated Nafion 117 and novel PVDF-based membranes. It was validated that the speciation of vanadium can be performed using the pre-edge peak energy and intensity. The experiments were performed as follows: strips of the membrane were exposed from one site to a V3+ solution (green) and from the other site to a VO2+ solution (yellow). The ions could diffuse into the membrane from both sides. A change of color of the membrane strip was observed. The blue color in the middle of the strip indicated that VO2+ was formed where V3+ and VO2+ got in contact. Using XANES this reaction inside Nafion was proven.
Absorption effects in total reflection X-ray fluorescence (TXRF) analysis are important to consider, especially if external calibration is to be applied. With a color X-ray camera (CXC), that enables spatially and energy resolved XRF analysis, the absorption of the primary beam was directly visualized for mL-droplets and an array of pL-droplets printed on a Si-wafer with drop-on-demand technology. As expected, deposits that are hit by the primary beam first shade subsequent droplets, leading to a diminished XRF signal. This shading effect was quantified with enhanced precision making use of sub-pixel analysis that improves the spatial resolution of the camera. The measured absorption was compared to simulated results using three different model calculations. It was found they match very well (average deviation < 10%). Thus errors in quantification due to absorption effects can be accounted for in a more accurate manner.
Shading Effects in SR-TXRF: Calculations and Experimental Visualization using a Color X-ray Camera
(2014)
An automatic sample changer chamber for total reflection X-ray fluorescence (TXRF) and X-ray absorption near-edge structure (XANES) analysis in TXRF geometry was successfully set up at the BAMline at BESSY II. TXRF and TXRF-XANES are valuable tools for elemental determination and speciation, especially where sample amounts are limited (<1 mg) and concentrations are low (ng ml⁻¹ to μg ml⁻¹). TXRF requires a well defined geometry regarding the reflecting surface of a sample carrier and the synchrotron beam. The newly installed chamber allows for reliable sample positioning, remote sample changing and evacuation of the fluorescence beam path. The chamber was successfully used showing accurate determination of elemental amounts in the certified reference material NISTwater 1640. Low limits of detection of less than 100 fg absolute (10 pg ml⁻¹) for Ni were found. TXRF-XANES on different Re species was applied. An unknown species of Re was found to be Re in the +7 oxidation state.
Polymer electrolyte membranes (PEM) are polymers which act as separator in an electrochemical cell and allow ionic charge flow to close the circuit. Widely used membranes are perfluorinated sulfonic-acid (PFSA) ionomers (e.g. Nafion™), a class of ion-conducting polymers with remarkable ion conductivity and chemical-mechanical stability. A nanoscopic water system around the sulfonic acid capped side-chains of the otherwise hydrophobic polymer allows the ion transport. Only approx. 14 molecules of water per sulfonic group are present in the fully hydrated state. In vanadium redox flow batteries (VRFB) ideally only protons are transported. The weak ion selectivity of Nafion™ is the main cause for the transportation of vanadium, briefly vanadium crossover, in VRFBs a system we have investigated more closely, lately. The consequences of crossover are a concentration imbalance and a self-discharge of the battery, which leads to a decrease of the capacity. The development of efficient energy storage systems is crucial for the transformation towards a renewable energy based economy. The VRFB has a great potential as a commercial electrochemical energy storage system due to properties including, but not limited to, no cross-contamination, a long cycle-life and a theoretically unlimited capacity. VRFB consists of two half-cells, which are linked to electrolyte tanks and separated by a membrane. The membrane plays a major role in overall cell performance. So far, vanadium transportation models, which include diffusion, migration, electroosmotic convection and pressure gradients, are inconsistent. There is no agreement in the literature on the diffusion coefficients of vanadium species (e.g. published diffusion coefficients of V2+ are located between 3.13·10-12 m²s-1 and 9.44·10-12 m²s-1) indicate that thevanadium crossover is not well understood and there is a lack on a more fundamental level. Since the membrane transport is the rate-determining step of the crossover as well as of the proton exchange it is extremely important to understand these phenomena on a fundamental level. This will eventually allow us to design better membranes. Alternative materials need to show equivalent performance compared with Nafion™ and ideally be more selective regarding ion cross-membrane transport.There are several options to approach the chemistry i.e. the interaction of dissolved and bound ions inside the confined water body of ionomeric membranes. Vanadium ions are a versatile model as UV/VIS data can be used to distinguish between the 5 species V2+, V3+, VO2+, VO2+ and V2O33+ of the electrolyte. Infrared spectroscopy has been applied to study the interaction of sulfonic groups and the ions. Molecular dynamic modeling is another approach to study the distribution and distances of ions. We now introduce X-ray absorption near edge structure spectroscopy (XANES) to study species and species changes inside Nafion™ and a novel membrane based on poly(1,1-difluoroethylene) (PVDF). We evaluated the methods and investigated the influence of irradiation, temperature and hydration on the measurements.
Total Reflection X-ray Fluorescence (TXRF) is a small footprint, ressource efficient micro-analytical tool for trace elemental determination. However, depending on the matrix TXRF is also challenging in several ways: the preparation of a representative aliquot maybe difficult for slurries, shading effects and matrix effects may occur and the applicability of an IS maybe hampered by interferences (fitting may be impaired as well) or inhomogeneities. It is therefore crucial to understand and if applicable mitigate the influence of the before mentioned phenomena. We have used the small volume approach using pL droplets to study shading in TXRF previously.Using this approach thin specimens in a favorable geometry are prepared with a well defined morphology; this way minimizing matrix effects and shading. To be used as standard it is also necessary to determine the delivered elemental amounts. Here we will present on the performance of a commercial ink-jet printer cartridge to deliver defined volumes and elemental amounts. The microscopic specimens obtained have been successfully applied to determine relative sensitivities in TXRF and prepare references to study coded apertures in grazing incidence full field micro-XRF.The homogeneous lateral distribution of analyte and IS may be probed by micro-XRF, to obtain information on alikeness of analyte and IS. To study the alikeness of the in depth distribution (film-like or particle-like) angle scans can deliver valuable data. Here we present first results on angle scans using a prototype GIXRF set up and a commercial TXRF instrument.
We have studied the micro-homogeneity of boron carbide powders by inductively coupled Plasma optical emission spectrometry (ICP-OES) and total reflection X-ray fluorescence spectrometry (TXRF) using slurry sampling. To get information on the particle size distributions of the powders, the stabilized slurries of boron Carbide powders were nebulized, the aerosols were transported into a Batelle impactor and the droplets were collected on the impactor stages bearing TXRF sample holders. In a first series of measurements, parameters of the impaction like the duration of the impaction and the use of glutinous substance on the sample holders were optimized. The different mass size fractions for industrial boron Carbide powders were determined by weight measurements of the fractions collected on the different stages. The established particle size distributions were in the range of 0.5 to >16 µm and found similar to those determined by laser diffraction reported elsewhere. Analyses of the mass fractions by slurry sampling TXRF showed that Ca, Ti, Cr, Mn, Fe, Ni and Cu within the measurements errors were homogeneously distributed over the mass fractions between 0.5 and 4 µm and that their concentrations agreed with the bulk composition, as determined with ICP-OES subsequent to digestion. However, light underestimates were found at the 5 (Mn) up to 150 µg g-1 (Fe) level. Finally, boron carbide powders were washed out with nitric acid with different concentrations and leaching solutions and the residues were analyzed by ICP-OES and TXRF respectively.
It is shown that up to 60% of the residual trace impurities in the powder studied can be removed by leaching with 34% (v/v) of nitric acid.