A layer of Sb atoms, implanted with an energy of 400 keV and a nominal dose of 5×1016 atoms/cm2 into a high purity silicon wafer, was certified for its areal density (atoms/cm2) using Rutherford backscattering spectrometry (RBS), instrumental neutron activation analysis (INAA) and inductively coupled plasma isotope dilution mass spectrometry (ICP-IDMS) and for its isotope ratio using INAA and ICP-IDMS. Excellent agreement between the results of the different independent methods was found. In the present work, the measurements of the homogeneity of the areal density of Sb, previously determined with RBS in spots having 1 mm diameter, are improved with synchrotron X-ray fluorescence analysis: Higher precision in even smaller sample spots allows to estimate a reduced inhomogeneity of the whole batch of samples of the order of only 0.4%. Thus the uncertainty of the certified value can further be reduced. Down to fractions of a chip with 0.3×0.4 mm2 area, the areal density is now certified as (4.81±0.06)×1016 Sb atoms/cm2, where the expanded uncertainty 0.06 (coverage factor k=2) corresponds to only 1.2%. The relative merits of the different analytical methods are discussed.
This Report describes the certification of the reference material antimony implanted in Si/SiO2 intended to be used for calibration of surface and near surface analytical methods. It describes the preparation, homogeneity measurements and the analytical work performed for the certification of both Areal density of antimony Atoms (retained dose) and the isotope amount Ratio as well as giving considerations on the stability of the material.