Special new techniques of atomic spectrometric methods for trace analysis of high purity metals
(2003)
In this work, different calibration strategies for the determination of trace elements in pure copper metal by nanosecond laser ablation ICP-MS were investigated. In addition to certified reference materials (CRMs), pellets of doped copper powder were used for calibration. The micro homogeneity of the CRMs as well as the solution-doped pellets was sufficient to use them as calibration samples in combination with a laser spot size of 200 µm. In contrast, pellets doped with analytes in solid form showed a significant heterogeneity. For most of the investigated analytes and copper CRMs the measured mass fractions were within ± 20% of their certified values when other copper CRMs were used as calibration samples. When solution-doped powder pellets were used as calibration samples a systematic trend towards mass fractions below the certified values was observed for nearly all elements determined in the analysed CRMs. Thermal fractionation effects during the ablation of the solution-doped pellets were suspected as the extent of the fractionation depends on the irradiance, whereas fractionation is reduced at higher irradiance.