Filtern
Erscheinungsjahr
- 2010 (3) (entfernen)
Dokumenttyp
Sprache
- Englisch (3)
Referierte Publikation
- ja (3)
Schlagworte
- Dissolved organic matter (DOM) (1)
- Flame retardants (1)
- Humic substance (1)
- Interlaboratory comparison (ILC) (1)
- Internal standard (1)
- International intercomparison (1)
- International measurement evaluation programme (IMEP) (1)
- Polybrominated biphenyls (PBBs) (1)
- Polybrominated diphenyl ethers (PBDEs) (1)
- Polycyclic aromatic hydrocarbon (PAH) (1)
Recently, the International Measurement Evaluation Programme (IMEP) organized an interlaboratory comparison on total concentrations of eight polycyclic aromatic hydrocarbons (PAHs) in natural inland waters. It was carried out in support to the European Union Water Framework Directive (EU-WFD) that lists these eight as priority substances.
As sample matrix, we used groundwater spiked with humic acid as a model for the colloidal substances that are present in natural inland waters. Humic substances can adsorb PAHs, but we found that some laboratories did not apply analytical procedures that sufficiently accounted for this. One of these laboratories was involved in establishing the reference values. We show how this impacted on the reliability of their measurements.
Many participants accepted our invitation to report their measurement uncertainties. We assessed their results against our reference values and uncertainties, and provided z and zeta scores. Although the overall measurement capability appears satisfactory, there is room for improvement of analytical procedures as regards the use of measurement standards.
An international intercomparison involving eight national metrology institutes (NMIs) was conducted to establish their current measurement capabilities for determining five selected congeners from the brominated flame retardant classes polybrominated diphenyl ethers and polybrominated biphenyls. A candidate reference material consisting of polypropylene fortified with technical mixtures of penta-, octa- and decabromo diphenyl ether and decabromo biphenyl, which was thoroughly assessed for material homogeneity and stability, was used as study material. The analytical procedures applied by the participants differed with regard to sample pre-treatment, extraction, clean-up, employed calibrants and type of calibration procedure as well as regarding analytical methods used for separation, identification and quantification of the flame retardant congeners (gas chromatography coupled to an electron capture detector (GC-ECD), gas chromatography-mass spectrometry in the electron ionisation mode (GC-EI-MS), gas chromatography-mass spectrometry in the electron capture negative ionisation mode (GC-ECNI-MS), and liquid chromatography-inductive coupled plasma-mass spectrometry (LC-ICP-MS)). The laboratory means agreed well with relative standard deviations of the mean of means of 1.9%, 4.8%, 5.5% and 5.4% for brominated diphenyl ether (BDE) 47, 183 and 209 and for the brominated biphenyl (BB) congener 209, respectively. For BDE 206, a relative standard deviation of 28.5% was obtained. For all five congeners, within-laboratory relative standard deviations of six measurements obtained under intermediate precision conditions were between 1% and 10%, and reported expanded measurements uncertainties typically ranged from 4% to 10% (8% to 14% for BDE 206). Furthermore, the results are in good agreement with those obtained in the characterization exercise for determining certified values for the flame retardant congeners in the same material. The results demonstrate the state-of-the-art measurement capabilities of NMIs for quantifying representative BDE congeners and BB 209 in a polymer. The outcome of this intercomparison (pilot study) in conjunction with possible improvements for employing exclusively calibrants with thoroughly assessed purity suggests that a key comparison aiming at underpinning calibration and measurement capability (CMC) claims of NMIs can be conducted.