For ten years µ-XRF (micro-focus X-ray fluorescence) analysis has been performed with SEM/EDS (scanning electron microscope with an energy dispersive X-ray detector) so that non-destructive analysis of elements at trace level concentrations below 100 µg g-1 becomes possible. This can be considered as a valuable completion of the classical electron probe microanalysis by EDS, an analytical method 'suffering' from rather poor limits of detection in the range of one to two orders of magnitude higher than those of µ-XRF. Based on a representative actual application, namely analysis of RoHS relevant elements at trace concentration levels, the performance of the rather new analytical method with respect to its limits of detection is systematically evaluated. CRMs (certified reference materials) specially prepared to support the quantitative XRF analysis of RoHS relevant elements were employed. On the other side, based on calculations of µ-XRF spectra according to a recently developed physical model the optimization of the analytical performance is also successfully undertaken.
Determination of the efficiency of an energy dispersive X-ray spectrometer up to 50 keV with a SEM
(2009)
Both electron and polychromatic photon excitations (micro-focus X-ray source) at a scanning electron microscope (SEM) are used to determine the efficiency of an energy dispersive X-ray spectrometer up to 50 keV by means of a calibrated X-ray spectrometer and reference materials (RM) specially selected for this purpose.
Popular X-ray tube models available in the literature, i.e. 'Pella', 'Ebel', and 'Finkelshtein and Pavlova', are systematically evaluated with the focus on the estimation of the associated uncertainties. Also taken in consideration and compared is our recent semi-empirical own approach already employed in our lab. This has been working for the common target elements rhodium, molybdenum and tungsten and was further extended in the present work for the target elements copper, chromium and vanadium. By using a modern scanning electron microscope/energy dispersive spectroscopy (SEM/EDS) system this time, higher performances such as stability of the beam current and especially the better energy resolution of the EDS have enabled the reliable extension of our own X-ray tube spectrum approach into the low-energy range, due to increasing interest. Hence, also the more challenging X-ray lines of copper, chromium and vanadium L-series lying in the energy range below 12?keV are included into the model. Such low-energy L-lines or, e.g. M-lines of tungsten, are not treated explicitly by the other existing popular algorithms for the nowadays widely used geometries of side-window tubes, offering a unique virtue to our present, modern approach. With our own model, a measurement uncertainty of the X-ray tube spectra (considering the uncertainties associated with the SEM beam current, the detector acceptance solid angle and efficiency of the spectrometer) within 15% has been estimated. The validation of the approach is demonstrated with metrological measurements with a calibrated SEM/EDS system geometrically configured as a side-window X-ray tube.
The accurate knowledge of the properties of polycapillary X-ray semi-lenses has a significant influence on the quantitative results in a widespread field of applications involving microfocus X-ray beams. A routine procedure for the characterisation of a polycapillary X-ray semi-lens with a scanning electron microscope (SEM) having attached an energy dispersive spectrometer (EDS) is presented in this paper. A key issue of the procedure consists of fitting the semi-lens in front of the EDS for spectra acquisition. Relevant semi-lens parameters such as focal distance, full width at half maximum (FWHM) of the acceptance area, and transmission are determined in parallelising mode of the semi-lens. Special attention has been paid to the calculation of the transmission.