Within the Collaborative Research Center InPROMPT a novel process concept for the hydroformylation of long-chained olefins is studied in a mini-plant, using a rhodium complex as catalyst in the presence of syngas. Recently, the hydroformylation in micro¬emulsions, which allows for the efficient recycling of the expensive rhodium catalyst, was found to be feasible. However, the high sensitivity of this multi-phase system with regard to changes in temperature and composition demands a continuous observation of the reaction to achieve a reliable and economic plant operation. For that purpose, we tested the potential of both online NMR and Raman spectroscopy for process control. The lab-scale experiments were supported by off-line GC-analysis as a reference method.
A fiber optic coupled probe of a process Raman spectrometer was directly integrated into the reactor. 25 mixtures with varying concentrations of olefin (1-dodecene), product (n-tridecanal), water, n-dodecane, and technical surfactant (Marlipal 24/70) were prepared according to a D-optimal design. Online NMR spectroscopy was implemented by using a flow probe equipped with 1/16” PFA tubing serving as a flow cell. This was hyphenated to the reactor within a thermostated bypass to maintain process conditions in the transfer lines.
Partial least squares regression (PLSR) models were established based on the initial spectra after activation of the reaction with syngas for the prediction of unknown concentrations of 1-dodecene and n-tridecanal over the course of the reaction in the lab-scale system. The obtained Raman spectra do not only contain information on the chemical composition but are further affected by the emulsion properties of the mixtures, which depend on the phase state and the type of micelles. Based on the spectral signature of both Raman and NMR spectra, it could be deduced that especially in reaction mixtures with high 1-dodecene content the formation of isomers as a competitive reaction was dominating. Similar trends were also observed during some of the process runs in the mini-plant. The multivariate calibration allowed for the estimation of reactants and products of the hydroformylation reaction in both laboratory setup and mini-plant.
Monitoring chemical reactions is the key to chemical process control. Today, mainly
optical online methods are applied. NMR spectroscopy has a high potential for direct
loop process control. Compact NMR instruments based on permanent magnets
are robust and relatively inexpensive analysers, which feature advantages like low
cost, low maintenance, ease of use, and cryogen-free operation. Instruments for
online NMR measurements equipped with a flow-through cell, possessing a good
signal-to-noise-ratio, sufficient robustness, and meeting the requirements for
integration into industrial plants (i.e., explosion safety and fully automated data
analysis) are currently not available off the rack.
Intensified continuous processes are in focus of current research. Flexible (modular)
chemical plants can produce different products using the same equipment with short
down-times between campaigns and quick introduction of new products to the
market. In continuous flow processes online sensor data and tight closed-loop control
of the product quality are mandatory. If these are not available, there is a huge risk of
producing large amounts of out-of-spec (OOS) products. This is addressed in the
European Unionʼs Research Project CONSENS (Integrated Control and Sensing)
by development and integration of smart sensor modules for process monitoring and
control within such modular plant setups.
The presented NMR module is provided in an explosion proof housing of 57 x 57 x
85 cm module size and involves a compact 43.5 MHz NMR spectrometer together
with an acquisition unit and a programmable logic controller for automated data
preparation (phasing, baseline correction) and evaluation. Indirect Hard Modeling
(IHM) was selected for data analysis of the low-field NMR spectra. A set-up for
monitoring continuous reactions in a thermostated 1/8” tubular reactor using
automated syringe pumps was used to validate the IHM models by using high-field
NMR spectroscopy as analytical reference method.
Ugly Spectra and Lousy interfaces – Challenges for Compact NMR Spectroscopy in Process Control
(2016)
With the introduction of advanced process analytical technology, the closeness of key process variables to their limits can be directly controlled and the product can be classified or even released in real time.
Compact NMR instruments can make NMR spectroscopy accessible in industrial and harsh environments for process control.