Filtern
Dokumenttyp
Sprache
- Englisch (7)
Referierte Publikation
- ja (7) (entfernen)
Schlagworte
- IDMS (3)
- Metrology (2)
- Analyte-matrix separation (1)
- Archaeometry (1)
- Automotive exhaust emissions (1)
- CCQM (1)
- CRM (1)
- Certification (1)
- Environmental pollution monitoring (1)
- Finow Canal (1)
Organisationseinheit der BAM
The quantitative analysis of toxic metals in plastics is very important for different sectors of industry and daily life. Many routine procedures have been established based on X-ray fluorescence or inductively coupled plasma atomic emission spectrometry. However, all of them require suitable reference materials to calibrate or validate. These reference materials ideally are being certified by reference procedures. The development of such reference procedures for sulfur and the four toxic elements cadmium, chromium, mercury and lead in plastics is described here. The procedures are based on double isotope dilution mass spectrometry including analyte–matrix separation. The applied mass spectrometric techniques are thermal ionization mass spectrometry as well as inductively coupled plasma mass spectrometry. Memory effects of mercury and dissolution of chromium(III) oxide have been considered especially. The expanded uncertainties have been improved from the percent range down to the per mill range during the development of the procedure from the early analysis of BCR-680/681 to the recent analysis of CCQM-P106. With the fully developed procedures expanded uncertainties (k = 2) between 0.1 and 0.4% for cadmium, chromium, lead and sulfur and around 1% for mercury can be achieved. The so developed procedures have been successfully applied to the certification of reference materials as well as to intercomparisons organized by CCQM.
A two-step separation procedure for the quantification of Pd and Pt in automotive exhaust emissions using isotope dilution mass spectrometry was established using a combination of cation and anion exchange chemistry. AG 50W-X12 was used as cation exchange resin and DGA as weakly basic anion exchange resin. This procedure enabled the effective separation of Pd and Pt from the matrix and from interfering elements. Additionally Pd and Pt were collected in separate chromatographic fractions, which increased the precision of the isotope ratio determination by separate measurements using single collector sector field ICPMS. The analytical procedure was validated by analysing the synthetically prepared samples and the certified reference materials BCR-723 (road dust) and IAEA-450 (algae). For the SI-traceable results complete uncertainty budgets were calculated yielding relatively expanded uncertainties (k = 2) of ≈1% for analyte masses in the ng range. Procedure blanks of 55 pg Pd and 3 pg Pt were obtained. The detection limits were calculated as 12 pg for Pd and 7 pg for Pt. Additionally, Pd and Pt blank levels of different filter materials are presented as well as the first results for automotive exhaust particles collected on cellulose filters.
The CCQM-K88 key comparison was organized by the Inorganic Analysis Working Group of CCQM to test the abilities of the national metrology institutes to measure the mass fraction of lead in lead-free solder containing silver and copper. National Metrology Institute of Japan (NMIJ), National Institute of Metrology of China (NIM) and Korea Research Institute of Standards and Science (KRISS) acted as the coordinating laboratories. The participants used different measurement methods, though most of them used inductively coupled plasma optical emission spectrometry (ICP-OES) or isotope-dilution inductively coupled plasma mass spectrometry (ID-ICP-MS). Accounting for relative expanded uncertainty, comparability of measurement results was successfully demonstrated by the participating NMIs for the measurement of the mass fraction of lead in lead-free solder at the level of 200 mg/kg.
It is expected that metals at mass fractions greater than approximately 100 mg/kg in lead-free solder containing silver and copper can be determined by each participant using the same technique(s) employed for this key comparison to achieve similar uncertainties mentioned in the present report.
A modified Pb–matrix separation procedure using NH4HCO3 solution as eluent has been developed and validated for determination of Pb isotope amount ratios by thermal ionization mass spectrometry. The procedure is based on chromatographic separation using the Pb·Spec resin and an in-house-prepared NH4HCO3 solution serving as eluent. The advantages of this eluent are low Pb blanks (<40 pgmL-1) and the property that NH4HCO3 can be easily removed by use of a heating step (>60 °C). Pb recovery is >95 % for water samples. For archaeological silver samples, however, the Pb recovery is reduced to approximately 50 %, but causes no bias in the determination of Pb isotope amount ratios. The validated procedure was used to determine lead isotope amount ratios in Trojan silver artefacts with expanded uncertainties (k=2) <0.09 %.
CCQM-K143 is a key comparison that assesses participants’ ability to prepare single element calibration solutions. Preparing calibration solutions properly is the cornerstone of establishing a traceability link to the International System of Units (SI), and therefore should be tested in order to confirm the validity of CCQM comparisons of more complex materials. CCQM-K143 consisted of participants each preparing a single copper calibration solution at 10 g/kg copper mass fraction and shipping 10 bottled aliquots of that solution to the coordinating laboratory, the National Institute of Standards and Technology (NIST). The masses and mass fraction for the prepared solutions were documented with the submitted samples.
The solutions prepared by all participants were measured at NIST by high performance inductively coupled plasma optical emission spectroscopy (HP-ICP-OES). The intensity measurements for copper were not mapped onto values of mass fraction via calibration. Instead, ratios were computed between the measurements for copper and simultaneous measurements for manganese, the internal standard, and all subsequent data reductions, including the computation of the KCRV and the degrees of equivalence, were based on these ratios. Other than for two participants whose measurement results appeared to suffer
from calculation or preparation errors, all unilateral degrees of equivalence showed that the measured values did not differ significantly from the KCRV. These results were confirmed by a second set of ICP-OES measurements performed by the Physikalisch-Technische Bundesanstalt (PTB). CCQM-K143 showed that participants are capable of preparing calibration solutions starting from high purity, assayed copper metal.
Similar steps are involved when preparing solutions for other elements, so it seems safe to infer that similar capabilities should prevail when preparing many different, single-element solutions.
A recently introduced inductively coupled plasma-time-of-flight-mass spectrometer (ICP-ToF-MS) shows enhanced sensitivity compared to previous developments and superior isotope ratio precision compared to other ToF and commonly used single-collector ICP-MS instruments. Following this fact, an improvement for isotope dilution ICP-MS using the new instrumentation has been reported. This study aimed at investigating whether this improvement also meets the requirements of species-specific isotope dilution using GC/ICP-MS, where short transient signals are recorded. The results of the analysis of monomethylmercury (MMHg) of a sediment reference material show that isotope ratio precision of ICP-MS instruments equipped with quadrupole, sector-field, and time-of-flight mass analyzers is similar within a broad range of peak signal-to-noise ratio when analyzing one isotopic system. The procedural limit of quantification (LOQ) for MMHg, expressed as mass fraction of Hg being present as MMHg, w(Hg)MMHg, was similar as well for all investigated instruments and ranged between 0.003 and 0.016 μg/kg. Due to the simultaneous detection capability, the ICP-ToF-MS might, however, be more favorable when several isotopic systems are analyzed within one measurement. In a case study, the GC/ICP-ToF-MS coupling was applied for analysis of MMHg in sediments of Finow Canal, a historic German canal heavily polluted with mercury. Mass fractions between 0.180 and 41 μg/kg (w(Hg)MMHg) for MMHg, and 0.056 and 126 mg/kg (w(Hg)total) for total mercury were found in sediment samples taken from the canal upstream and downstream of a former chemical plant.
Soil certified reference material (CRM), UME EnvCRM 03 was produced by a collaborative approach among national metrology institutes, designated institutes and university research laboratories within the scope of the EMPIR project: Matrix Reference Materials for Environmental Analysis. This paper presents the sampling and processing methodology, homogeneity, stability, characterization campaign, the assignment of property values and their associated uncertainties in compliance with ISO 17034:2016. The material processing methodology involves blending a natural soil sample with a contaminated soil sample obtained by spiking elemental solutions for 8 elements (Cd, Co, Cu, Hg, Ni, Pb, Sb and Zn) to reach the level of warning risk monitoring values specified for metals and metalloids of soils in Europe. Comparative homogeneity and stability test data were obtained by two different institutes, ensuring the reliability and back up of the data. The certified values and associated expanded uncertainties for the total mass fractions of thirteen elements (As, Cd, Co, Cr, Cu, Fe, Hg, Mn, Ni, Pb, Sb, V and Zn) are established. The developed CRM can be used for the development and validation of measurement procedures for the determination of the total mass fractions of elements in soil and also for quality control/assurance purposes. The developed CRM is the first example of a soil material originating from Türkiye.