Filtern
Erscheinungsjahr
- 2018 (2) (entfernen)
Dokumenttyp
Sprache
- Englisch (2)
Referierte Publikation
- nein (2)
Schlagworte
- Trace humidity measurement (2) (entfernen)
Organisationseinheit der BAM
This work presents a first attempt to use ionic liquids as a new coating for planar coulometric sensors. These sensors are used for the measurement of trace humidity in various gases. Usually, the coating of the sensors is tetraphosphorus decaoxide and its hydrolysis products. Instead, a hygroscopic ionic liquid was used as sensor coating in this work. Generated frost point temperatures tf in the gas ranged from -80 °C to -30 °C, which is equivalent to vapour mole fractions xv from 0.5 μmol·mol-1 to 376 μmol·mol-1. In addition to the coulometric sensors, the generated humidity is determined by a precision dew point hygrometer as reference. First results show that it is possible to measure humidity with ionic liquid based coulometric sensors.
Planar coulometric sensors were investigated in humidified synthetic air at various absolute gas pressures, i. e. 2 bar, 5, bar, 10 bar, and 15 bar. Humidified gas flow at adjusted gas pressure was split into two flows, one passed a coulometric sensor and the other one passed a reference hygrometer after decompression. Both signals were recorded and then compared after calculation of resulting frost point temperature. Calculation is based on a calibration function obtained at ambient pressure. Comparison showed that an increasing pressure resulted in a higher derivation between sensor signal (calculated frost point temperature) and reference frost point temperature. At an absolute pressure of 2 bar the differences were minor in consideration of the uncertainty, however at 15 bar the differences were 6.77 K. Nevertheless, it was possible to measure the gas humidity at higher pressure with coulometric trace humidity sensors.