There are many different methods to characterize air-coupled ultrasonic transducers for non-destructive testing. Data sheets of various manufacturers contain information about some parameters important for the performance of transducers, but this information is not standardized, so that a comparison between probes of different manufacturers is difficult. Therefore, the German Society for Non-Destructive Testing (DGZfP) is working on a guideline to characterization of air-coupled probes.
One of the topics in this guideline is the application of thermoacoustic transducers for the characterization of receivers and another topic is the application of microphones for the characterization of transmitters. In this presentation we compare various characterization methods with the particular focus on the characterization of thermoacoustic transducers using an optical microphone. Both thermoacoustic transmitters and optical microphones have a very large bandwidth compared to conventional air-coupled transducers, but their spectrum is not entirely linear, which needs to be taken into account if they are applied as reference transducers.