The quantification of the sulphur mass fraction in pure copper and copper alloys by GDMS and LA-ICP-MS revealed a lack of traceability mainly due to a lack of suitable certified reference materials for calibrating the instruments. Within this study GDMS and LA-ICP-MS were applied as routine analytical tools to quantify sulphur in copper samples by applying reference materials as calibrators, which were characterized for their sulphur mass fraction by IDMS beforehand. Different external calibration strategies were applied including a matrix cross type calibration. Both techniques with all calibration strategies were validated by using certified reference materials (others than those used for calibration) and good agreement with the reference values was achieved except for the matrix cross type calibration, for which the agreement was slightly worse. All measurement results were accompanied by an uncertainty statement. For GDMS, the relative expanded (k = 2) measurement uncertainty ranged from 3% to 7%, while for LA-ICP-MS it ranged from 11% to 33% when applying matrix-matched calibration in the sulphur mass fraction range between 25 mg kg-1 and 1300 mg kg-1. For cross-type calibration the relative expanded (k = 2) measurement uncertainty need to be increased to at least 12% for GDMS and to at least 54% for LA-ICP-MS to yield metrological compatibility with the reference values. The so obtained measurement results are traceable to the international system of units (SI) via IDMS reference values, which is clearly illustrated by the unbroken chain of calibrations in the metrological traceability scheme.
The Uncertainty Paradox: Molar Mass of Enriched Versus Natural Silicon Used in the XRCD Method
(2020)
The X-ray crystal density method uses silicon spheres highly enriched in 28Si as a primary method for the dissemination of the SI base unit kilogram yielding smallest possible uncertainties associated with the mass m within a few parts in 10-8. This study compares different available and newly developed analytical methods and their results for the determination of the molar mass M of silicon highly enriched in 28Si (Me) and of silicon (Mx) with an almost natural isotopic distribution. While for Me relative uncertainties urel(Me) in the lower 10-9 range are obtained routinely, it was not possible to fall below a value of urel(Mx) < 4 x 10-6 in the case of natural silicon, which is approximately three orders of magnitude larger. The application of the state-of the-art isotope ratio mass spectrometry accompanied with sophisticated thoroughly investigated methods suggests an intrinsic cause for the large uncertainty associated with the molar mass of natural silicon compared to the enriched material.
Isotope amount ratios of krypton were measured on subsamples from one large batch of high purity krypton separated from the atmosphere. Synthetic mixtures of enriched krypton isotopes were used to calibrate the measurements with small uncertainties. The result is a primary isotopic gas standard (PIGS) IRMM-2030 with certified values for isotope ratios, isotopic composition, and molar mass of krypton with small combined uncertainties uc, evaluated according to the ISO/BIPM Guide (GUM). It is commercially available from IRMM-Geel or from MESSER (Duisburg, D). The certified krypton isotope amount ratios in the PIGS IRMM-2030 are as follows: n(78Kr)/n(84Kr)=0.006 232 5(55), n(80Kr)/n(84Kr)=0.040 107(17), n(82Kr)/n(84Kr)=0.203 43(12), n(83Kr)/n(84Kr)=0.201 79(11), and n(86Kr)/n(84Kr)=0.303 205(59) with expanded uncertainty U=kuc and coverage factor k=2. The molar mass of Kr in this sample is M(Kr)=83.798 02(16) g/mol. These values are in good agreement with published measurements of atmospheric krypton but have smaller combined uncertainties and are calibrated by means of synthetic isotope mixtures. The values of the PIGS are traceable to the SI. Measurements of isotope amount ratios of krypton in other samples can be linked to SI using this PIGS.