Objectives: The mechanical properties and seeding with endothelial cells were investigated in fresh and cryopreserved human umbilical vein.
Methods: Human umbilical veins (HUV) were frozen in Euro-Collins/1M DMSO at –1°C/min and stored in liquid nitrogen. Stress-strain relationships of fresh and thawed veins were determined in an uniaxial tension-testing rig. HUV endothelial cells (HUVEC) were seeded onto denuded HUV under static conditions and grown for 3d. Luminal surfaces were analyzed by scanning electron microscopy. Calcein-stained cells were seeded hyperconfluently to determine the cell retention capacity of fresh and cryopreserved veins.
Results: The stress-strain relationships of HUV followed a biphasic pattern typical for natural vessels. Neither the failure stress (2.71±0.36 vs. 3.25±0.97 N, n=3) nor the displacement required to achieve failure (9.73±0.9 vs. 7.43±2.07mm, n=3) were altered by cryopreservation. The burst pressure was estimated as approx. 1000mm Hg within the limitations of the uniaxial model. HUVEC seeded onto denuded HUV formed patches (at 9E3 cells per cm2) or an almost confluent endothelium (at 3E4 cells per cm2) within three days. The capacity to retain seeded HUVEC of denuded HUV was not altered by cryopreservation (1.15±0.08E5 vs. 1.26±0.14E5 cells per cm2, n=6).
Conclusions: The burst pressure of HUV seems to be sufficiently high for the human arterial circulation and is not altered by cryopreservation. HUVEC can establish a confluent endothelium on denuded HUV. Therefore HUV appears to be a suitable storable scaffold for vascular tissue engineering.
Introduction:
Neointimal hyperplasia after percutaneous coronary intervention remains a major determinant of in-stent restenosis (ISR). The extent of mechanical vessel injury correlates with ISR. A new ex vivo porcine stent model was introduced and evaluated comparing different stent designs.
Methods:
Coronary arteries were prepared from pig hearts from the slaughterhouse and used for ex vivo implantations of coronary stents. One basic stent design in two configurations (dogbone, DB; non-dogbone, NDB) was used. Vascular injury was determined according to a modified injury score (IS).
Results:
Standardized experimental conditions ensured comparable vessel dimensions and overstretch data. DB stents caused more severe IS compared to NDB stents. The mean IS and the IS at the distal end of all stents were significantly reduced for NDB stents (ISMean, DB, 1.16 ±0.12; NDB, 1.02 ±0.12; p=0.018; ISDist, DB, 1.39 ±0.28; NDB, 1.13 ±0.24; p=0.03).
Discussion/Conclusion:
The introduced ex-vivo model allowed the evaluation of different stent designs exclude unfavorable stent designs.
Introduction
Shear induced multimerisation of von-Willebrand-factor (vWF) is supposed to play an important role in coagulation inside extracorporeal membrane oxygenators. However, there is no proof that links observed vWF structures to computed or measured flow conditions.
Methods
The structures of multimeric vWF fibers, observed in clinically used membrane oxygenators is examined using immunofluorescence microscopy (IFM) using Carstairs’ staining method (positive ethics committee vote). The flow around the membrane fibres inside the oxygenator is investigated in terms of shear rate, wall shear velocity and streamlines by using CFD (RANS, Carreau-Yasuda viscosity, geometry remodelled after high-resolution µCT-scans). By interpreting the histological and numerical results in this common context, indications for shear induced coagulation mechanisms can be identified.
Results
The fibre structures of multimeric vWF build regular but not exactly symmetric formations around the contact face (CF) between the crosswise stacked oxygenator fibres (OF), see fig.1B, vWF marked red. Annular around the CF arranged, cells are likely to be found, see fig.1B, nuclei marked blue.
The computed streamlines around the OF show attached flow around the circular fibres. However, the irregular arrangement of real OF produce considerable cross flow between the interconnected neighbouring channels, in contrast to previous 2D-simulations. Thus, the CF are washed around closely by blood, also from neighbouring channels. The wall shear velocity streamlines form regular, slightly asymmetric shapes around the contact faces. The occurring maximum shear rates are in the range of 1,000 1/s.
Discussion
The shapes of vWF structures found in clinically used oxygenators match the computational results in terms of wall shear velocity and streamlines well. The accumulation of cells close to the CF can also be explained by fluid mechanics, as there are small shear gradients and slow velocities. However, occurring shear rates between OFs are too low to trigger multimerisation of vWF. That raises the question where in the circuit the actual activation of vWF is started and how, at least partly chained, vWF multimeres are attracted towards the OF surface. A next step will be the investigation of the actual shear rate triggered (or mediated) multimerisation of vWF. Towards this end, microfluidic experiments with shear triggered coagulation will be performed. Also of big interest is the computation of the flow situation in the oxygenator in proximity to chaining threads, which have been ignored in computations so far. However, first a realistic representation of the effective viscosity in computations is needed, which is not available yet.
Clot formation within membrane oxygenators (MOs) remains a critical problem during extracorporeal membrane oxygenation (ECMO). The composition of the clots—in particular, the presence of von Willebrand factor (vWF)—may be an indicator for prevalent nonphysiological flow conditions, foreign body reactions, or coagulation abnormalities in critically ill patients. Mats of interwoven gas exchange fibers from randomly collected MOs (PLS, Maquet, Rastatt, Germany) of 21 patients were stained with antibodies (anti‐vWF and anti‐P‐selectin) and counterstained with 4′,6‐diamidino‐2‐phenylindole. The extent of vWF‐loading was correlated with patient and technical data. While 12 MOs showed low vWF‐loadings, 9 MOs showed high vWF‐loading with highest accumulations close to crossing points of adjacent gas fibers. The presence and the extent of vWF‐fibers/“cobwebs,” leukocytes, platelet–leukocyte aggregates (PLAs), and P‐selectin‐positive platelet aggregates were independent of the extent of vWF‐loading. However, the highly loaded MOs were obtained from patients with a significantly elevated SOFA score, severe thrombocytopenia, and persistent liver dysfunction. The coagulation abnormalities of these critically ill patients may cause an accumulation of the highly thrombogenic and elongated high‐molecular‐weight vWF multimers in the plasma which will be trapped in the MOs during the ECMO therapy.
Coagulative disorders, especially clotting during extracorporeal membrane oxygenation, are frequent complications. Direct visualization and analysis of deposits in membrane oxygenators using computed tomography (CT) may provide an insight into the underlying mechanisms causing thrombotic events. However, the already established multidetector CT1 (MDCT) method shows major limitations. Here, we demonstrate the feasibility of applying industrial micro-CT (μCT) to circumvent these restrictions. Three clinically used membrane oxygenators were investigated applying both MDCT and μCT.
The scans were analyzed in terms of clot volume and local clot distribution. As validation, the clot volume was also determined from the fluid volume, which could be filled into the respective used oxygenator compared to a new device. In addition, cross-sectional CT images were compared with crosscut oxygenators. Based on the μCT findings, a morphological measure (sphericity) for assessing clot structures in membrane oxygenators is introduced. Furthermore, by comparing MDCT and μCT results, an augmentation of the MDCT method is proposed, which allows for improved clot volume determination in a clinical setting.
Feasibility of detecting thrombotic deposits in membrane oxygenators using micro computed tomography
(2019)
BACKGROUND: Thrombosis remains a critical complication during venovenous extracorporeal membrane oxygenation (VV ECMO). The involvement of neutrophil extracellular traps (NETs) in thrombogenesis has to be discussed. The aim was to verify NETs in the form of cell-free DNA (cfDNA) in the plasma of patients during ECMO.
METHODS: A fluorescent DNA-binding dye (QuantifFluor®, Promega) was used to detect cell-free DNA in plasma samples. cfDNA concentrations from volunteers (n = 21) and patients (n = 9) were compared and correlated with clinical/technical data before/during support, ECMO end and time of a system exchange.
RESULTS: Before ECMO, patients with a median (IQR) age of 59 (51/63) years, SOFA score of 11 (10/15), and ECMO run time of 9.0 (7.0/19.5) days presented significantly higher levels of cfDNA compared to volunteers (6.4 (5.8/7.9) ng/μL vs. 5.9 (5.4/6.3) ng/μL; p = 0.044). Within 2 days after ECMO start, cfDNA, inflammatory, and hemolysis parameters remained unchanged, while platelets decreased (p = 0.005). After ECMO removal at the end of therapy, cfDNA, inflammation, and coagulation data (except antithrombin III) remained unchanged. The renewal of a system resulted in known alterations in fibrinogen, d-dimers, and platelets, while cfDNA remained unchanged.
CONCLUSION: Detection of cfDNA in plasma of ECMO patients was not an indicator of acute and circuit-induced thrombogenesis.
BACKGROUND:
Shear-induced conformational changes of von Willebrand factor (vWF) may be responsible for coagulation disorder and clot formation inside membrane oxygenators (MOs) during extracorporeal membrane oxygenation (ECMO) therapy.
OBJECTIVE:
The aim was to identify vWF structures inside clinically used MOs and employ computational fluid dynamics to verify the corresponding flow conditions.
METHODS:
Samples from gas exchange membranes (GEM) from MOs were analysed for accumulations of vWF and P-selectin-positive platelets using immunofluorescence techniques. Streamlines and shear rates of the flow around GEMs were computed using a laminar steady Reynolds-Averaged-Navier-Stokes approach.
RESULTS:
Most samples were colonized with equally distributed leukocytes, integrated in thin cobweb-like vWF-structures. Only 25 % of the samples showed extended accumulations of vWF. Computed streamlines showed considerable cross flow between interconnected neighbouring channels. Stagnation points were non-symmetric and contact faces were washed around closely. The occurring maximum shear rates ranged from 2,500 to 3,000 1/s.
CONCLUSIONS:
If pronounced vWF structures are present, shape and extent match the flow computations well. Computed shear rates bear a critical degree of uncertainty due to the improper viscosity model. If flow conditions inside the MO were sufficient to affect vWF, a more consistent distribution of vWF across the samples should be present.
Clot formation within membrane oxygenators (MOs) remains a critical problem during extracorporeal membrane oxygenation (ECMO). The composition of the clots-in particular, the presence of von Willebrand factor (vWF)-may be an indicator for prevalent nonphysiological flow conditions, foreign body reactions, or coagulation abnormalities in critically ill patients. Mats of interwoven gas exchange fibers from randomly collected MOs (PLS, Maquet, Rastatt, Germany) of 21 patients were stained with antibodies (anti-vWF and anti-P-selectin) and counterstained with 4 ',6-diamidino-2-phenylindole. The extent of vWF-loading was correlated with patient and technical data. While 12 MOs showed low vWF-loadings, 9 MOs showed high vWF-loading with highest accumulations close to crossing points of adjacent gas fibers. The presence and the extent of vWF-fibers/"cobwebs," leukocytes, platelet-leukocyte aggregates (PLAs), and P-selectin-positive platelet aggregates were independent of the extent of vWF-loading. However, the highly loaded MOs were obtained from patients with a significantly elevated SOFA score, severe thrombocytopenia, and persistent liver dysfunction. The coagulation abnormalities of these critically ill patients may cause an accumulation of the highly thrombogenic and elongated high-molecular-weight vWF multimers in the plasma which will be trapped in the MOs during the ECMO therapy.
Over the past decade, veno-venous extracorporeal membrane oxygenation (vvECMO) has been increasingly utilized in respiratory failure in patients. This study presents our institution´s experience focusing on the life span of ECMO systems reflecting the performance of a particular system. A retrospective review of our ECMO database identified 461 adult patients undergoing vvECMO (2010-2017). Patients that required more than one system and survived the first exchange >24 hours (n = 139) were included. Life span until the first exchange and exchange criteria were analyzed for all systems (PLS, Cardiohelp HLS-set, both Maquet Cardiopulmonary, Rastatt, Germany; Deltastream/Hilite7000LT, iLA-activve, Xenios/NovaLung, Heilbronn, Germany; ECC.O5, LivaNova, Mirandola, Italy). At our ECMO center, the frequency of a system exchange was 30%. The median (IQR) life span was 9 (6-12) days. There was no difference regarding the different systems (p = 0.145 and p = 0.108, respectively). However, the Deltastream systems were exchanged more frequently due to elective technical complications (e. g. worsened gas transfer, development of coagulation disorder, increased bleedings complications) compared to the other exchanged systems (p = 0.013). In summary, the used ECMO systems are safe and effective for acute respiratory failure. There is no evidence for the usage of a specific system. Only the increased predictability of an imminent exchange preferred the usage of a Deltastream system. However, the decision to use a particular system should not depend solely on the possible criteria for an exchange.