Institute
Regional hyperthermia, a clinical cancer therapy, is the main topic of the Sonderforschungsbereich Hyperthermia: Scientific Methods and Clinical Applications'' at Berlin. In recent years, technological improvements towards a better concentration of heat to the desired target region have been achieved. These include a rather sophisticated integrated software environment for therapy planning and a new hyperthermia applicator. In a next step, a detailed closed loop monitoring of the actual treatment is to be developed. For this purpose the hyperthermia applicator is combined with an MRI system, which will allow to check the positioning of the patients and to measure individual blood perfusion as well as the 3D temperature distribution. The measurements will then be employed for an on-line control of the whole treatment. In this intended setting, new fast feedback control algorithms will come into play.
After a short summary on therapy planning and the underlying technologies we discuss quantitative medicine by giving a short overview on medical image data, summarizing some applications of computer based treatment planning, and outlining requirements on medical planning systems. Then we continue with a description of our medical planning system {\sf HyperPlan}. It supports typical working steps in therapy planning, like data aquisition, segmentation, grid generation, numerical simulation and optimization, accompanying these with powerful visualization and interaction techniques.
HyperPlan is a software system for performing 3D-simulations and treatment planning in regional hyperthermia. It allows the user to understand the complex effects of electromagnetic wave propagation and heat transport inside a patient's body. Optimized power amplitudes and phase settings can be calculated for the BSD radiowave applicators Sigma 60 and Sigma 2000 (eye-applicator). HyperPlan is built on top of the modular, object-oriented visualization system Amira. This system already contains powerful algorithms for image processing, geometric modelling and 3D graphics display. HyperPlan provides a number of hyperthermia-specific modules, allowing the user to create 3D tetrahedral patient models suitable for treatment planning. In addition, all numerical simulation modules required for hyperthermia simulation are part of HyperPlan. This guide provides a step-by-step introduction to hyperthermia planning using HyperPlan. It also describes the usage of the underlying visualization system Amira.
Hyperthermia Treatment Planning in Clinical Cancer Therapy:Modelling, Simulation, and Visualization
(1997)
Hyperthermia Treatment Planning in Clinical Cancer Therapy: Modelling, Simulation and Visualization
(1997)
\noindent The speaker and his co-workers in Scientific Computing and Visualization have established a close cooperation with medical doctors at the Rudolf--Virchow--Klinikum of the Humboldt University in Berlin on the topic of regional hyperthermia. In order to permit a patient--specific treatment planning, a special software system ({\sf\small HyperPlan}) has been developed. \noindent A mathematical model of the clinical system ({\it radio frequency applicator with 8 antennas, water bolus, individual patient body}) involves Maxwell's equations in inhomogeneous media and a so--called bio--heat transfer PDE describing the temperature distribution in the human body. The electromagnetic field and the thermal phenomena need to be computed at a speed suitable for the clinical environment. An individual geometric patient model is generated as a quite complicated tetrahedral ``coarse'' grid (several thousands of nodes). Both Maxwell's equations and the bio--heat transfer equation are solved on that 3D--grid by means of {\em adaptive} multilevel finite element methods, which automatically refine the grid where necessary in view of the required accuracy. Finally optimal antenna parameters for the applicator are determined . \noindent All steps of the planning process are supported by powerful visualization methods. Medical images, contours, grids, simulated electromagnetic fields and temperature distributions can be displayed in combination. A number of new algorithms and techniques had to be developed and implemented. Special emphasis has been put on advanced 3D interaction methods and user interface issues.