Refine
Document Type
- Article (9)
- ZIB-Report (4)
- In Proceedings (2)
- Poster (2)
Is part of the Bibliography
- no (17)
Keywords
- image registration (2)
- 3D reconstruction (1)
- Digitally Reconstructed Radiograph (DRR), Anatomy Reconstruction, Statistical Shape and Intensity Model (SSIM), GPU acceleration (1)
- GPU acceleration (1)
- acetabular orientation (1)
- articulated shape and intensity models (1)
- bone anchored hearing implant (1)
- digitally reconstructed radiographs (1)
- mesh deformation (1)
- segmentation (1)
Information about the temporal bone size and variations of anatomical structures are crucial for a safe positioning of the Vibrant Bonebridge B-FMT. A radiological based preoperative planning of the surgical procedure decreases the surgical time and minimizes the risk of complications.
We developed a software tool, which allows a catch up of foreign DICOM data based CT temporal bone scans. The individual CT scan is transmitted into a 3D reconstructed pattern of the temporal bone. In this 3D reconstruction the individually favored position of the B- FMT should be found.
The software allows a determination of a safe B-FMT position by identifying the individual relation of middle fossa, jugular bulb and external auditory canal. Skull thickness and screw length are contained parameters for the surgical planning.
An easy to handle software tool allows a radiologically data based safe and fast surgical positioning of the B-FMT.
The aim of this paper is to provide a comprehensive overview of the MICCAI 2020 AutoImplant Challenge. The approaches and publications submitted and accepted within the challenge will be summarized and reported, highlighting common algorithmic trends and algorithmic diversity. Furthermore, the evaluation results will be presented, compared and discussed in regard to the challenge aim: seeking for low cost, fast and fully automated solutions for cranial implant design. Based on feedback from collaborating neurosurgeons, this paper concludes by stating open issues and post-challenge requirements for intra-operative use.
We present an automated method for extrapolating missing
regions in label data of the skull in an anatomically plausible manner. The ultimate goal is to design patient-specic cranial implants for correcting large, arbitrarily shaped defects of the skull that can, for example, result from trauma of the head. Our approach utilizes a 3D statistical shape model (SSM) of the skull and a 2D generative adversarial network (GAN) that is trained in an unsupervised fashion from samples of healthy patients alone. By tting the SSM to given input labels containing the skull defect, a First approximation of the healthy state of the patient is obtained. The GAN is then applied to further correct and smooth the output of the SSM in an anatomically plausible manner. Finally, the defect region is extracted using morphological operations and subtraction between the extrapolated healthy state of the patient and the defective input labels. The method is trained and evaluated based on data from the MICCAI 2020 AutoImplant challenge. It produces state-of-the art results on regularly
shaped cut-outs that were present in the training and testing data of the challenge. Furthermore, due to unsupervised nature of the approach, the method generalizes well to previously unseen defects of varying shapes that were only present in the hidden test dataset.
This study’s objective was the generation of a standardized geometry of the healthy nasal cavity.
An average geometry of the healthy nasal cavity was generated using a statistical shape model based on 25 symptom-free subjects. Airflow within the average geometry and these geometries was calculated using fluid simulations. Integral measures of the nasal resistance, wall shear stresses (WSS) and velocities were calculated as well as cross-sectional areas (CSA). Furthermore, individual WSS and static pressure distributions were mapped onto the average geometry.
The average geometry featured an overall more regular shape that resulted in less resistance, reduced wall shear stresses and velocities compared to the median of the 25 geometries. Spatial distributions of WSS and pressure of average geometry agreed well compared to the average distributions of all individual geometries. The minimal CSA of the average geometry was larger than the median of all individual geometries (83.4 vs. 74.7 mm²).
The airflow observed within the average geometry of the healthy nasal cavity did not equal the average airflow of the individual geometries. While differences observed for integral measures were notable, the calculated values for the average geometry lay within the distributions of the individual parameters. Spatially resolved parameters differed less prominently.
We present an efficient GPU-based method to generate virtual X-ray images from tetrahedral meshes which are
associated with attenuation values. In addition, a novel approach is proposed that performs the model deformation
on the GPU. The tetrahedral grids are derived from volumetric statistical shape and intensity models (SSIMs) and describe anatomical structures. Our research targets at reconstructing 3D anatomical shapes by comparing virtual X-ray images generated using our novel approach with clinical data while varying the shape and density of the SSIM in an optimization process. We assume that a deformed SSIM adequately represents an anatomy of interest when the similarity between the virtual and the clinical X-ray image is maximized. The OpenGL implementation presented here generates accurate (virtual) X-ray images at interactive rates, thus qualifying it for its use in the reconstruction process.
Die Positionierung des B-FMT der Vibrant Bonebridge kann aufgrund der anatomischen Verhältnisse des Mastoids und der Größe des Aktuators ohne eine vorherige Beurteilung der individuellen Computertomographie (CT) des Felsenbeins problematisch sein. Die Entwicklung eines einfach zu bedienenden Viewers, welcher eine Positionierung des B-FMT im Felsenbeinmodell ermöglicht und hier auf individuelle potenzielle anatomische Konflikte hinweist sowie Lösungsmöglichkeiten anbietet, kann ein hilfreiches Werkzeug zur präoperativen Positionierung sein.
Ziel der Arbeit war die Definition von Anforderungen und die Anfertigung eines Prototyps eines Vibrant-Bonebridge-Viewers.
Auf der Basis einer ZIBAmira-Software-Version und der Inklusion eines B-FMT-Modells unter Erstellung eines Felsenbeinmodells, welches die intuitive Beurteilung von Konflikten ermöglicht, erfolgte die Erstellung des Prototyps eines Vibrant-Bonebridge-Viewers.Ergebnisse. Die Segmentierungszeit der individuellen DICOM-Daten („digital imaging and communications in medicine“) beträgt etwa 5 min. Eine Positionierung im individuellen 3-D-Felsenbeinmodell ermöglicht die quantitative und qualitative Beurteilung von Konflikten (Sinus sigmoideus, mittlere Schädelgrube) und das Aufsuchen einer bevorzugten Position. Das Anheben des B-FMT mittels virtueller Unterlegscheiben kann simuliert werden.
Der erstellte Vibrant-Bonebridge-Viewer ermöglicht verlässlich eine Simulation der B-FMT-Positionierung. Die klinische Anwendbarkeit muss evaluiert werden.
We propose a novel GPU-based approach to render virtual X-ray projections of deformable tetrahedral meshes. These meshes represent the shape and the internal density distribution of a particular anatomical structure and are derived from statistical shape and intensity models (SSIMs). We apply our method to improve the geometric reconstruction of 3D anatomy (e.g.\ pelvic bone) from 2D X-ray images. For that purpose, shape and density of a tetrahedral mesh are varied and virtual X-ray projections are generated within an optimization process until the similarity between the computed virtual X-ray and the respective anatomy depicted in a given clinical X-ray is maximized. The OpenGL implementation presented in this work deforms and projects tetrahedral meshes of high resolution (200.000+ tetrahedra) at interactive rates. It generates virtual X-rays that accurately depict the density distribution of an anatomy of interest. Compared to existing methods that accumulate X-ray attenuation in deformable meshes, our novel approach significantly boosts the deformation/projection performance. The proposed projection algorithm scales better with respect to mesh resolution and complexity of the density distribution, and the combined deformation and projection on the GPU scales better with respect to the number of deformation parameters. The gain in performance allows for a larger number of cycles in the optimization process. Consequently, it reduces the risk of being stuck in a local optimum. We believe that our approach contributes in orthopedic surgery, where 3D anatomy information needs to be extracted from 2D X-rays to support surgeons in better planning joint replacements.