Fakultät Maschinenbau
Refine
Year of publication
- 2018 (2)
Document Type
Language
- English (2)
Has Fulltext
- no (2)
Is part of the Bibliography
- no (2)
Keywords
- Membrane Oxygenator (2) (remove)
Institute
Begutachtungsstatus
- begutachtet (1)
- peer-reviewed (1)
Modelling blood flow an shear induced coagulation in membraene oxygenators (MO) is challenging. The relevant geometry of oxygenator fibers (OF) and chaining threads is complex and spans several length scales. In relevant scales and regimes blood shows several significant non-Newtonian effects. Existing models are only capable of accounting for some, but not all relevant effects. Additionally, coagulation processes are influencing fluid properties and geometry significantly. Due to the enormous size of the discretised geometries highly detailed viscosity and coagulation properties of blodd flow in MOs. First step is to find a gemoetry dependent viscosity representation on basis of parametric micro channel experiments with anti-coagulated blood. Next step is a statistic coagulation model, based on micro channel experiments with human (re-calcified citrated) whole blood an evaluation of clinically used osygenators. Since shear rate dependent (i.e. viscosity dependet) coagulation in return influences the viscosity, a combined model with suitable implementation in a RANS framework is necessary. Towards this end, micro channel experiments with new and used single OFs triggering coagulation are performed. Structures of multimeric von Willebrand fibers (vWF), as indicator for shear induced coagulation, are compared to computed and measured flow conditions, using immunofluorescence microscopy, RANS-computations and µPIV, respectively. Preliminary examinations in clinically used MOs show good agreement between occurring structures of vWF, cell depositions and computed flow patterns (geometry form µCT-Scans). However, computed shear rates might be to low to actually trigger activation of vWF. The complex geometry of MOs results in huge meshes, which makes RANS with statistical modelling of viscosity and coagulation a reasonable approach. Towards this end, experimental data on micro channel level with evaluation on real application level is crucial. Especially regarding clotting processes, micro fluidic experiments are powerful research tool.
BACKGROUND:
Shear-induced conformational changes of von Willebrand factor (vWF) may be responsible for coagulation disorder and clot formation inside membrane oxygenators (MOs) during extracorporeal membrane oxygenation (ECMO) therapy.
OBJECTIVE:
The aim was to identify vWF structures inside clinically used MOs and employ computational fluid dynamics to verify the corresponding flow conditions.
METHODS:
Samples from gas exchange membranes (GEM) from MOs were analysed for accumulations of vWF and P-selectin-positive platelets using immunofluorescence techniques. Streamlines and shear rates of the flow around GEMs were computed using a laminar steady Reynolds-Averaged-Navier-Stokes approach.
RESULTS:
Most samples were colonized with equally distributed leukocytes, integrated in thin cobweb-like vWF-structures. Only 25 % of the samples showed extended accumulations of vWF. Computed streamlines showed considerable cross flow between interconnected neighbouring channels. Stagnation points were non-symmetric and contact faces were washed around closely. The occurring maximum shear rates ranged from 2,500 to 3,000 1/s.
CONCLUSIONS:
If pronounced vWF structures are present, shape and extent match the flow computations well. Computed shear rates bear a critical degree of uncertainty due to the improper viscosity model. If flow conditions inside the MO were sufficient to affect vWF, a more consistent distribution of vWF across the samples should be present.