Refine
Language
- English (5)
Has Fulltext
- no (5)
Is part of the Bibliography
- no (5)
Keywords
- ECMO (3)
- MDCT (1)
- blood (1)
- cell- free DNA (1)
- coagulation (1)
- hemocompatibility (1)
- hemolysis (1)
- histological evaluation (1)
- inflammation (1)
- membrane lung (1)
Institute
Begutachtungsstatus
- peer-reviewed (3)
Thrombus formation within extracorporeal membrane oxygenation (ECMO) devices remained a critical complication. One reason seems to be the contact of blood with large artificial surfaces within the membrane lung (ML). The aim was to test the hemocompatibility of different naïve ECMO materials. Blood and platelets from five healthy volunteers were incubated with gas exchange (GF) and heat exchange fibers (HE) from four different commercial available new MLs representing different antithrombogenic coatings. Adherent platelets were stained with rhodamine-phalloidin. Surface coverage was quantified with ImageJ. Non-adherent platelets were stained with antibodies (CD62P, PAC-1, CD61) and fibrinogen to detect platelet activation with flow cytometry. Hemolysis of red blood cells after material contact was detected. All ECMO-materials were non-hemolytic and did not induce platelet activation. However, platetelet adhesion (median (IQR)) was significantly elevated on uncoated GFs made of polymethylpentene (GF-PMP; 12 (7-19)%) and on GFs from the Hilite-MLs (GF-Hilite; 13 ((8-19)%) compared to the other materials. In vitro testing of platelet adhesion disclosed significant differences of ECMO-materials with different antithrombogenic surface coatings. Instead, circulating platelets remained non-activated. ECMO-materials and its coatings were non-hemolytic. Finally, this study confirmed the good hemocompatibility of GFs and HEs from commerciall available MLs.
Extracorporeal membrane oxygenation (ECMO) was established as a treatment for severe cardiac or respiratory disease. Intra-device clot formation is a common risk. This is based on complex coagulation phenomena which are not yet sufficiently understood. The objective was the development and validation of a methodology to capture the key properties of clots deposed in membrane lungs (MLs), such as clot size, distribution, burden, and composition. One end-oftherapy PLS ML was examined. Clot detection was performed using multidetector computed tomography (MDCT), microcomputed tomography (μCT), and photography of fiber mats (fiber mat imaging, FMI). Histological staining was conducted for von Willebrand factor (vWF), platelets (CD42b, CD62P), fibrin, and nucleated cells (4′, 6-diamidino-2-phenylindole, DAPI). The three imaging
methods showed similar clot distribution inside the ML. Independent of the imaging method, clot loading was detected predominantly in the inlet chamber of the ML. The μCT had the highest accuracy. However, it was more expensive and time consuming than MDCT or FMI. The MDCT detected the clots with low scanning time. Due to its lower resolution, it only showed clotted areas but not
the exact shape of clot structures. FMI represented the simplest variant, requiring little effort and resources. FMI allowed clot localization and calculation of clot volume. Histological evaluation indicated omnipresent immunological deposits throughout the ML. Visually clot-free areas were covered with leukocytes and platelets forming platelet-leukocyte aggregates (PLAs). Cells were embedded in vWF cobwebs, while vWF fibers were negligible. In conclusion, the presented
methodology allowed adequate clot identification and histological classification
of possible thrombosis markers such as PLAs.
BACKGROUND: Thrombosis remains a critical complication during venovenous extracorporeal membrane oxygenation (VV ECMO). The involvement of neutrophil extracellular traps (NETs) in thrombogenesis has to be discussed. The aim was to verify NETs in the form of cell-free DNA (cfDNA) in the plasma of patients during ECMO.
METHODS: A fluorescent DNA-binding dye (QuantifFluor®, Promega) was used to detect cell-free DNA in plasma samples. cfDNA concentrations from volunteers (n = 21) and patients (n = 9) were compared and correlated with clinical/technical data before/during support, ECMO end and time of a system exchange.
RESULTS: Before ECMO, patients with a median (IQR) age of 59 (51/63) years, SOFA score of 11 (10/15), and ECMO run time of 9.0 (7.0/19.5) days presented significantly higher levels of cfDNA compared to volunteers (6.4 (5.8/7.9) ng/μL vs. 5.9 (5.4/6.3) ng/μL; p = 0.044). Within 2 days after ECMO start, cfDNA, inflammatory, and hemolysis parameters remained unchanged, while platelets decreased (p = 0.005). After ECMO removal at the end of therapy, cfDNA, inflammation, and coagulation data (except antithrombin III) remained unchanged. The renewal of a system resulted in known alterations in fibrinogen, d-dimers, and platelets, while cfDNA remained unchanged.
CONCLUSION: Detection of cfDNA in plasma of ECMO patients was not an indicator of acute and circuit-induced thrombogenesis.
Thrombosis continues to be a significant complication during venovenous extracorporeal membrane oxygenation (V-V ECMO). Platelet activation markers might serve as indicators of inflammation and thrombogenesis. The aim was to identify these markers in ECMO patients. Blood from 10 ECMO patients (before, during, after ECMO) and 11 healthy volunteers were collected to determine platelet-neutrophil-aggregates (PNAs), platelet-monocyte-aggregates (PMAs), fibrinogen-binding, and P-selectin-expression on platelets by flow cytometry. Critical illness was associated with significantly elevated levels of PNAs and PMAs, increased P-selectin expression, reduced fibrinogen-binding, and restricted activation of platelets. Although PNAs and PMAs decreased significantly within 2 hours after the initiation of ECMO and remained at those levels, ECMO did not affect basal P-selectin expression and fibrinogen-binding. These results correlated with coagulation activation. Platelet markers before ECMO were not indicators for an imminent system exchange and end of therapy. In conclusion, platelet dysfunction during ECMO was mainly attributed to the critical illness. Extracorporeal membrane oxygenation support strengthened the restricted response of platelets to exogenous agonists (P-selectin). Furthermore, a decrease in PNAs/PMAs after ECMO started identified a reduced inflammatory response. There was no correlation of analyzed platelet parameters with the incidence of thrombotic complications.
Contact of blood with artificial surfaces triggers platelet activation. The aim was to compare platelet kinetics after venovenous extracorporeal membrane oxygenation (V-V ECMO) start and after system exchange in different etiologies of acute lung failure. Platelet counts and coagulation parameters were analyzed from adult patients with long and exchange-free (≥8 days) ECMO runs (n = 330) caused by bacterial (n = 142), viral (n = 76), or coronavirus disease 2019 (COVID-19) (n = 112) pneumonia. A subpopulation requiring a system exchange and with long, exchange-free runs of the second oxygenator (≥7 days) (n = 110) was analyzed analogously. Patients with COVID-19 showed the highest platelet levels before ECMO implantation. Independent of the underlying disease and ECMO type, platelet counts decreased significantly within 24 hours and reached a steady state after 5 days. In the subpopulation, at the day of a system exchange, platelet counts were lower compared with ECMO start, but without differences between underlying diseases. Subsequently, platelets remained unchanged in the bacterial pneumonia group, but increased in the COVID-19 and viral pneumonia groups within 2–4 days, whereas D-dimers decreased and fibrinogen levels increased. Thus, overall platelet counts on V-V ECMO show disease-specific initial dynamics followed by an ongoing consumption by the ECMO device, which is not boosted by new artificial surfaces after a system exchange.