Purpose
MR-based attenuation correction (AC) will become an integral part of combined PET/MR systems. Here, we propose a toolbox to validate MR-AC of clinical PET/MRI data sets.
Methods
Torso scans of ten patients were acquired on a combined PET/CT and on a 1.5-T MRI system. MR-based attenuation data were derived from the CT following MR–CT image co-registration and subsequent histogram matching. PET images were reconstructed after CT- (PET/CT) and MR-based AC (PET/MRI). Lesion-to-background (L/B) ratios were estimated on PET/CT and PET/MRI.
Results
MR–CT histogram matching leads to a mean voxel intensity difference in the CT- and MR-based attenuation images of 12% (max). Mean differences between PET/MRI and PET/CT were 19% (max). L/B ratios were similar except for the lung where local misregistration and intensity transformation leads to a biased PET/MRI.
Conclusion
Our toolbox can be used to study pitfalls in MR-AC. We found that co-registration accuracy and pixel value transformation determine the accuracy of PET/MRI.
Object
The consistency of parametric registration in multi-temporal magnetic resonance (MR) imaging studies was evaluated.
Materials and methods
Serial MRI scans of adult patients with a brain tumor (glioma) were aligned by parametric registration. The performance of low-order spatial alignment (6/9/12 degrees of freedom) of different 3D serial MR-weighted images is evaluated. A registration protocol for the alignment of all images to one reference coordinate system at baseline is presented. Registration results were evaluated for both, multimodal intra-timepoint and mono-modal multi-temporal registration. The latter case might present a challenge to automatic intensity-based registration algorithms due to ill-defined correspondences. The performance of our algorithm was assessed by testing the inverse registration consistency. Four different similarity measures were evaluated to assess consistency.
Results
Careful visual inspection suggests that images are well aligned, but their consistency may be imperfect. Sub-voxel inconsistency within the brain was found for allsimilarity measures used for parametric multi-temporal registration. T1-weighted images were most reliable for establishing spatial correspondence between different timepoints.
Conclusions
The parametric registration algorithm is feasible for use in this application. The sub-voxel resolution mean displacement error of registration transformations demonstrates that the algorithm converges to an almost identical solution for forward and reverse registration.