Development and characterization of a laboratory setup for photoacoustic NO2 determination based on the excitation of electronic 2B2 and 2B1 states using a low-cost semiconductor laser

  • This work gives a detailed characterization of a laboratory setup for photoacoustic NO2 trace gas detection at the ppb level. The signal generation is based on the excitation of electronic 2B2 and 2B1 states using a low-cost semiconductor laser emitting at 450 nm. An acoustic resonator was used for signal amplification and the modulation frequency of the laser was determined to 3395 Hz in order toThis work gives a detailed characterization of a laboratory setup for photoacoustic NO2 trace gas detection at the ppb level. The signal generation is based on the excitation of electronic 2B2 and 2B1 states using a low-cost semiconductor laser emitting at 450 nm. An acoustic resonator was used for signal amplification and the modulation frequency of the laser was determined to 3395 Hz in order to gain maximum signal amplification. The quality of resonant amplification was determined to 7.9. The gas samples were NO2 calibration gases diluted with pure nitrogen. The signal-to-noise ratio (SNR) dependency on the flow rate Q and the lock-in time constant τLIA was investigated, respectively, and the optimum values were specified to Q = 500 mL/min and τLIA = 2 s. In case of ambient noise, increasing τLIA to 10 s was evaluated as sufficient for SNR preservation. The noise level was measured in the absence of NO2 and it was determined to be composed of 51% electronic noise and 49% gas flow noise. With the analyte concentration ranging from 300 ppbV to 100 ppmV, the linear dependency of the photoacoustic amplitude on the NO2 concentration was specified, the sensitivity was determined to 110 μV/ppmV and the maximum measurement error was calculated to ±0.8%. The detection limit was determined to 2.0 ppbV. Furthermore, the stability of the signal was investigated and a maximum drift of ±1% was observed within a measuring period of 30 min. The response time τ90 was specified to 58 s. All results considered, this photoacoustic measuring system, which is based on low-cost signal generation and detection units, provides an excellent basis in view of developing a portable device for photoacoustic trace gas detection.show moreshow less

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author:Thomas RückOTHORCiD, Rudolf BierlOTH, Frank-Michael Matysik
DOI:https://doi.org/10.1016/j.sna.2017.03.024
Parent Title (English):Sensors and Actuators A: Physical
Publisher:Elsevier
Place of publication:Amsterdam, Niederlande
Document Type:Article
Language:English
Year of first Publication:2017
Release Date:2021/03/15
Volume:258
First Page:193
Last Page:200
Note:
Within the scope of the German government’s 6th Energy Research Program, the Federal Ministry of Economics and Technology (BMWi) has provided essential financial support for this work. The funding codes of the grant projects are 03ET1137A and 03ET1317A
Institutes:Fakultät Angewandte Natur- und Kulturwissenschaften
Fakultät Angewandte Natur- und Kulturwissenschaften / Sensorik-Applikationszentrum (SappZ)
Begutachtungsstatus:peer-reviewed
research focus:Digitale Transformation
Frontdoor-URL:https://opus4.kobv.de/opus4-oth-regensburg/1432
Einverstanden ✔