NO2 trace gas monitoring in air using off-beam quartz enhanced photoacoustic spectroscopy (QEPAS) and interference studies towards CO2, H2O and acoustic noise

  • We present the development and characterization as well as comprehensive interference studies of a photoacoustic NO2 trace gas detection system. The system is based on an off-beam quartz enhanced photoacoustic scheme (off-beam QEPAS) and signal generation was initiated by amplitude modulating a low-cost diode laser emitting at 450 nm. The QEPAS sensor element features double-resonantWe present the development and characterization as well as comprehensive interference studies of a photoacoustic NO2 trace gas detection system. The system is based on an off-beam quartz enhanced photoacoustic scheme (off-beam QEPAS) and signal generation was initiated by amplitude modulating a low-cost diode laser emitting at 450 nm. The QEPAS sensor element features double-resonant amplification, still it is only ∼ 5 × 5 × 2.5 mm in size. The individual and combined resonance characteristics were investigated and specified to 52 dB amplification, adding up 15 dB acoustic- and 37 dB mechanical-resonance amplification. The linearity of the photoacoustic signal dependency on the analyte concentration was verified from 200 ppbV to 100 ppmV NO2 in synthetic air. The detection limit (3σ) was determined to 1.8 ppbV using a lock-in time constant of 10 s and an averaging time of 20 s. The normalized noise equivalent absorption coefficient was specified to 2.5·10−8 W cm−1 Hz−0.5. The stability of the signal was investigated over time and a slight drift by 1‰ was observed after 30 min without temperature stabilizing the photoacoustic cell (PAC). Noise analysis was performed by means of Allan deviation and the inverse dependency of response time and precision of the system on the lock-in time constant was outlined. We performed interference analyses towards N2, O2, CO2, H2O and acoustic noise, respectively. Although neither spectral interferences nor losses due to slow NO2 VT-relaxation were observed, O2 was identified to cause a 15% signal drop due to VVNO2-O2-relaxation. Changing H2O concentrations were found to cause acoustic detuning, which cannot be compensated by adjusting the frequency of modulation, because of the double-resonant feature of the PAC. However, alternative approaches of compensation were discussed. Finally, we carried out heavy traffic noise simulations and determined the QEPAS setup to be 46 times less susceptible towards ambient noise compared to standard microphone-based photoacoustic setups.show moreshow less

Export metadata

Additional Services

Share in Twitter Search Google Scholar Statistics
Metadaten
Author:Thomas RückOTHORCiD, Rudolf BierlOTH, Frank-Michael Matysik
DOI:https://doi.org/10.1016/j.snb.2017.09.039
Parent Title (English):Sensors and Actuators B: Chemical
Publisher:Elsevier
Document Type:Article
Language:English
Year of first Publication:2018
Release Date:2021/03/12
Volume:255
Issue:Part 3, February
First Page:2462
Last Page:2471
Note:
Within the scope of the German government's 6th Energy Research Program, the Federal Ministry of Economic Affairs and Energy (BMWi) has provided essential financial support for this work. The funding codes of the grant projects are 03ET1137A and 03ET1317A.
Institutes:Fakultät Angewandte Natur- und Kulturwissenschaften
Fakultät Angewandte Natur- und Kulturwissenschaften / Sensorik-Applikationszentrum (SappZ)
Begutachtungsstatus:peer-reviewed
research focus:Digitale Transformation
Frontdoor-URL:https://opus4.kobv.de/opus4-oth-regensburg/1431
Einverstanden ✔
Diese Webseite verwendet technisch erforderliche Session-Cookies. Durch die weitere Nutzung der Webseite stimmen Sie diesem zu. Unsere Datenschutzerklärung finden Sie hier.