TY - JOUR A1 - Weigl, Stefan A1 - Feldmeier, Florian A1 - Bierl, Rudolf A1 - Matysik, Frank-Michael T1 - Photoacoustic detection of acetone in N2 and synthetic air using a high power UV LED JF - Sensors Actuators B Chemical N2 - The performance of a photoacoustic trace gas sensor for the detection of acetone in N2 and synthetic air is reported. The sensor system utilises an amplitude modulated UV LED. The light source has an emission maximum at 278 nm and a maximum CW output power of 300 mW according to the datasheet. Three different collimating and focusing approaches have been investigated to guide the highly divergent LED light into the acoustic resonator of the photoacoustic measurement cell. A 3D printed aluminium cell was designed to optimize light coupling by simultaneously minimizing the photoacoustic background signal generation. Hence, the diameter of the resonator was set to a comparable large diameter of 10 mm and the inner walls of the resonator were mirror polished. The additive manufacturing procedure allowed for integration of a spirally formed gas channel, enabling gas heating prior to detection. The sensor performance was investigated by measuring acetone in N2 and synthetic air at different concentrations. The UV LED current was set to 86 % of the maximum value according to the datasheet of the light source in order to increase the lifetime and thermal stability. An Allan-Werle deviation analysis validates a stable sensor performance. The limit of detection (LoD) was determined at a 3σ noise level with a 10 s lock-in amplifier time constant by sampling data points over 20 s with a data acquisition rate of 5 Hz. LoDs of 80.8 ppbV and 19.6 ppbV were obtained for acetone in N2 and synthetic air, respectively. KW - Acetone detection KW - Photoacoustic spectroscopy KW - High power UV LED KW - UV LED collimation system KW - Acetone breath analysis Y1 - 2020 U6 - https://doi.org/10.1016/j.snb.2020.128109 N1 - Corresponding author at: Sensorik-Applikationszentrum (SappZ) der Ostbayerischen Technischen Hochschule Regensburg VL - 316 IS - August SP - 1 EP - 11 PB - Elsevier ER - TY - CHAP A1 - Hausler, Peter A1 - Jobst, Simon A1 - Fischer, Johannes A1 - Roth, Carina A1 - Bierl, Rudolf T1 - Homogeneous Light Source for Surface Plasmon Resonance Imaging T2 - PHOTOPTICS 2020: proceedings of the 8th International Conference on Photonics, Optics and Laser Technology, Valletta, Malta, February 27-29, 2020 N2 - We describe how to build a homogeneous light source for Surface Plasmon Resonance Imaging (SPRi) which mainly finds its applications in pharmaceutical screening and biotechnology so far. SPR spectroscopy is a label-free, non-destructive and highly sensitive measurement principle for detecting changes in the refractive index in close vicinity of a gold surface. A transfer of this technology to a miniaturized sensor will broaden the range of possible applications. Commercial SPR assays are mainly working with a small number of sensing spots. In contrast, the SPR imaging system shown here will allow the use of an array of many sensing spots. In combination with chemical receptors designed as an artificial nose or an electronic tongue, the simultaneous detection of many analytes is envisioned. So far, lasers or other inhomogeneous light sources were used to illuminate the sensing surface, which is decreasing the systems sensitivity. We show a compact ( 300mm2) homogeneous top hat profile. The combination of a high bit-resolution camera with our new light source enables a reflectivity based surface plasmon resonance imaging system with a high refractive index unit (RIU) resolution. KW - Surface Plasmon Resonance Spectroscopy KW - Sensor KW - SPR-imaging KW - Miniaturization KW - Micro-opto-electro-mechanical Systems Y1 - 2020 U6 - https://doi.org/10.5220/0009168701630167 SP - 163 EP - 167 PB - SCITEPRESS - Science and Technology Publications ER - TY - PAT A1 - Olbrich, Florian A1 - Pongratz, Christian A1 - Bierl, Rudolf A1 - Ehrlich, Ingo T1 - Method and System for Evaluating a Structural Integrity of an Aerial Vehicle T2 - Europäisches Patentblatt (09.12.2020) zur Anmeldung: OTH Regensburg, Anmeldenummer: 19179054.2, Patentnummer 1001/3748327, Veröffentlichungsnummer: 3 748 327 Y1 - 2020 SP - 395 EP - 396 ER -