TY - JOUR A1 - Lenich, Andreas A1 - Bachmeier, S. A1 - Dendorfer, Sebastian A1 - Mayr, E. A1 - Nerlich, Michael A1 - Füchtmeier, Bernd T1 - Development of a test system to analyze different hip fracture osteosyntheses under simulated walking JF - Biomedizinische Technik. Biomedical engineering N2 - The mechanical complications of osteosyntheses after hip fractures are previously investigated by mostly static or dynamic uniaxial loading test systems. However, the physiologic loading of the hip joint during a normal gait is a multiplanar, dynamic movement. Therefore, we constructed a system to test osteosyntheses for hip fractures under physiologic multiplanar loading representative of normal gait. To evaluate the testing system, 12 femora pairs were tested under 25,000 cycles with two standard osteosyntheses (Proximal Femoral Nail Antirotation/Gamma3 Nail). For angular movement, the varus collapse to cut out (∝CO) (∝CO=4.8°±2.1° for blade and ∝CO=7.8°±3.8° for screw) was the dominant failure mode, and only slight rotational angle shifts (∝Rot) (∝Rot=1.7°±0.4° for blade and ∝Rot=2.4°±0.3° for screw) of the femoral head around the implant axis were observed. Angular displacements in varus direction and rotation were higher in specimens reinforced with screws. Hence, the cut out model and the migration directions showed a distinction between helical blade and hip screw. However, there were no significant differences between the different implants. The new setup is able to create clinical failures and allows to give evidence about the anchorage stability of different implant types under dynamic gait motion pattern. KW - biomechanical implant test KW - dynamic multiplanar loading KW - hip fractureg KW - implant migration KW - patient specific loading KW - Hüftgelenk KW - Knochenbruch KW - Osteosynthese KW - Bewegungsanalyse KW - Simulation Y1 - 2012 U6 - https://doi.org/10.1515/bmt-2011-0999 VL - 57 IS - 2 SP - 113 EP - 119 ER - TY - CHAP A1 - Putzer, Michael A1 - Weber, Tim A1 - Dendorfer, Sebastian T1 - Design studies on hip prosthesis using patient specific data T2 - Simpleware Users meeting, November 9th 2011, Bristol, UK Y1 - 2011 ER - TY - JOUR A1 - Förstl, Nikolas A1 - Süß, Franz A1 - Englert, Carsten A1 - Dendorfer, Sebastian T1 - Design of a reverse shoulder implant to measure shoulder stiffness during implant component positioning JF - Medical Engineering & Physics N2 - To avoid dislocation of the shoulder joint after reverse total shoulder arthroplasty, it is important to achieve sufficient shoulder stability when placing the implant components during surgery. One parameter for assessing shoulder stability can be shoulder stiffness. The aim of this research was to develop a temporary reverse shoulder implant prototype that would allow intraoperative measurement of shoulder stiffness while varying the position of the implant components. Joint angle and torque measurement techniques were developed to determine shoulder stiffness. Hall sensors were used to measure the joint angles by converting the magnetic flux densities into angles. The accuracy of the joint angle measurements was tested using a test bench. Torques were determined by using thin-film pressure sensors. Various mechanical mechanisms for variable positioning of the implant components were integrated into the prototype. The results of the joint angle measurements showed measurement errors of less than 5° in a deflection range of ±15° adduction/abduction combined with ±45° flexion/extension. The proposed design provides a first approach for intra-operative assessment of shoulder stiffness. The findings can be used as a technological basis for further developments. Y1 - 2023 U6 - https://doi.org/10.1016/j.medengphy.2023.104059 N1 - Corresponding author: Sebstian Dendorfer VL - 121 PB - Elsevier ET - Journal Pre-proof ER - TY - GEN A1 - Förstl, Nikolas A1 - Süß, Franz A1 - Englert, Carsten A1 - Dendorfer, Sebastian T1 - Design of a reverse shoulder implant to measure shoulder stiffness during implant component positioning T2 - Book of abstracts / ESB 2023, 28th Congress of the European Society of Biomechanics, 9-12 July 2023, Maastricht, The Netherlands. N2 - Introduction Dislocation of the shoulder joint is one of the more common complications after reverse total shoulder arthroplasty [1], which is often associated with malposition of the prosthetic components [2]. Therefore, achieving sufficient shoulder stability should not be neglected when positioning the implant components. One parameter for assessing shoulder stability can be shoulder stiffness. The aim of this work is to develop a reverse shoulder implant prototype that allows intraoperative measurement of shoulder stiffness while varying the position of the implant components. The measured stiffness could provide a quantitative statement regarding the optimal positioning of the implant components, which can be adjusted accordingly in the final reverse shoulder prosthesis. Methods To measure the stiffness of the shoulder joint, it is necessary to record the joint angles and the torques generated during movement. The changes in the rotation angles were measured using 3D hall sensors and magnets. The magnets were placed under the humerosocket, and the hall sensors were integrated into the glenosphere. The strength of the magnetic field was used to determine the position of the humerosocket in relation to the glenosphere. The accuracies of the angle measurements were tested using a test bench. Three thin film pressure sensors were used to record forces at different points under the humerosocket. To obtain a force value from the sensor signal, the sensors were calibrated using a load cell. The variation of the implant components positions was integrated into the prototype implant through different constructive mechanisms to adjust the stiffness of the shoulder joint. Results In the range of ±45° flexion/extension combined with ±15° adduction/abduction, the joint position could be determined with sufficient accuracy (error e ≤ 5°). The areas near the combined maximum deflections of ±45° flexion/extension and ±45° adduction/abduction indicate the greatest deviation from the target angle. The force values of the thin film sensors enable the calculation of moments around two axes. As variable component position parameters, the tray offset, the neck-shaft angle and the humerus version were integrated into the implant prototype. Discussion Ideally, the accuracy of the angle measurements should only depend on the amount of deflection and not on the direction of deflection. The asymmetric behavior indicates a deviation from the correct positioning of the hall sensors. The application of a calibration matrix could compensate for the measurement errors and could demonstrate the potential of the new method for joint angle measurements. The accuracy of the torque measurements and the functionality of the mechanical arresting mechanisms must be investigated in further studies. Overall, the developed measurement method can help to avoid malpositioning of the implant components in reverse total shoulder arthroplasty. References 1. Clark et al, J Shoulder and elbow surgery, 21:36-41 2012. 2. Randelli et al, J Musculoskeletal surgery, 98:15-18, 2014. Y1 - 2023 UR - https://esbiomech.org/conference/archive/2023maastricht/332.pdf ER - TY - CHAP A1 - Hammer, Joachim A1 - Dendorfer, Sebastian T1 - Cyclic loading and microstructure of cancellous bone T2 - Journal of Biomechanics Y1 - 2008 U6 - https://doi.org/10.1016/S0021-9290(08)70409-3 VL - 41 IS - Suppl. 1, July 2008 SP - S410 ER - TY - JOUR A1 - Reinker, Lukas A1 - Bläsing, Dominic A1 - Bierl, Rudolf A1 - Ulbricht, Sabina A1 - Dendorfer, Sebastian T1 - Correlation of Acceleration Curves in Gravitational Direction for Different Body Segments during High-Impact Jumping Exercises JF - sensors N2 - Osteoporosis is a common disease of old age. However, in many cases, it can be very well prevented and counteracted with physical activity, especially high-impact exercises. Wearables have the potential to provide data that can help with continuous monitoring of patients during therapy phases or preventive exercise programs in everyday life. This study aimed to determine the accuracy and reliability of measured acceleration data at different body positions compared to accelerations at the pelvis during different jumping exercises. Accelerations at the hips have been investigated in previous studies with regard to osteoporosis prevention. Data were collected using an IMU-based motion capture system (Xsens) consisting of 17 sensors. Forty-nine subjects were included in this study. The analysis shows the correlation between impacts and the corresponding drop height, which are dependent on the respective exercise. Very high correlations (0.83–0.94) were found between accelerations at the pelvis and the other measured segments at the upper body. The foot sensors provided very weak correlations (0.20–0.27). Accelerations measured at the pelvis during jumping exercises can be tracked very well on the upper body and upper extremities, including locations where smart devices are typically worn, which gives possibilities for remote and continuous monitoring of programs. KW - osteoporosis KW - inertial measurement units KW - wearable motion-tracking sensors KW - high-impact exercisess Y1 - 2023 U6 - http://nbn-resolving.de/urn/resolver.pl?urn:nbn:de:bvb:898-opus4-58217 N1 - Corresponding author: Lukas Reinker N1 - This work was supported by the Bavarian Academic Forum [BayWISS-Verbundkolleg Gesundheit] and the DHZK fond for behavioral cardiovascular diseases [Behaviorale kardiovaskuläre Erkrankungen; grand number: 81Z0400105]. VL - 23 IS - 4 PB - MDPI CY - Basel ER - TY - JOUR A1 - Roldán, J.C. A1 - Moralis, A. A1 - Dendorfer, Sebastian A1 - Witte, J. A1 - Reicheneder, C. T1 - Controlled central advancement of the midface after Le Fort III osteotomy by a 3-point skeletal anchorage JF - The Journal of craniofacial surgery N2 - A 3-point skeletal anchorage with taping screws for distraction osteogenesis after a Le Fort III osteotomy was applied for the first time in a severely mentally impaired patient where intraoral devices had to be avoided. All 3-force application points included the center of resistance, which allowed an optimal control on the resulting moment. A novel device for skeletal long-term retention into the nasal dorsum prevented a relapse, whereas adjustment of the midface position was observed. Fusioned three-dimensional computed tomography analysis revealed real movements not accessible by a conventional cephalometry. KW - Mund-Kiefer-Gesichts-Chirurgie KW - Le Fort III KW - distraction KW - vector KW - anchorage KW - three-dimensional computed tomography analysis KW - Distraktion KW - Knochenbildung KW - Osteotomie KW - Dreidimensionale Bildverarbeitung KW - Computertomografie Y1 - 2011 U6 - https://doi.org/10.1097/SCS.0b013e318231fc8d VL - 22 IS - 6 SP - 2384 EP - 2386 ER - TY - JOUR A1 - De Pieri, Enrico A1 - Atzori, Federica A1 - Ferguson, Stephen J. A1 - Dendorfer, Sebastian A1 - Leunig, Michael A1 - Aepli, Martin T1 - Contact force path in total hip arthroplasty: effect of cup medialisation in a whole-body simulation JF - HIP International N2 - Background: Cup medialisation down to the true acetabular floor in total hip arthroplasty with a compensatory femoral offset increase seems to be mechanically advantageous for the abductor muscles due to the relocation of the lever arms (body weight lever arm decreased, abductor lever arm increased). However, limited information is currently available about the effects of this reconstruction type at the head cup interface, compared to an anatomical reconstruction that maintains the natural lever arms. Through a whole-body simulation analysis, we compared medialised versus anatomical reconstruction in THA to analyse the effects on: (1) contact force magnitude at the head cup interface; (2) contact force path in the cup; and (3) abductor activity. Methods: Musculoskeletal simulations were performed to calculate the above-mentioned parameters using inverse dynamics analysis. The differences between the virtually implanted THAs were calculated to compare the medialised versus anatomical reconstruction. Results: Cup medialisation with compensatory femoral offset increase led to: (1) a reduction in contact force magnitude at the head cup interface up to 6.6%; (2) a similar contact force path in the cup in terms of sliding distance and aspect ratio; and (3) a reduction in abductor activity up to 17.2% (gluteus medius). Conclusions: In our opinion, these potential biomechanical gains do not generally justify a fully medialised reconstruction, especially in younger patients that are more likely to undergo revision surgery in their lifetime. Cup medialisation should be performed until sufficient press fit and bony coverage of a properly sized and oriented cup can be achieved. KW - Cup medialisation KW - femoral offset KW - hip contact force KW - total hip anthroplasty KW - total hip replacement KW - Hüftgelenkprothese KW - Kontaktkraft KW - Biomechanische Analyse KW - Simulation Y1 - 2020 U6 - https://doi.org/10.1177/1120700020917321 VL - 31 IS - 5 SP - 624 EP - 631 PB - Sage ER - TY - JOUR A1 - Englert, Carsten A1 - Angele, Peter A1 - Fierlbeck, J. A1 - Dendorfer, Sebastian A1 - Schubert, T. A1 - Müller, R. A1 - Lienhard, S. A1 - Zellner, Johannes A1 - Nerlich, Michael A1 - Neumann, Carsten T1 - Conductive bone substitute material with variable antibiotic delivery JF - Der Unfallchirurg N2 - A new bone substitute, consisting of hydroxylapatite and calcium sulphate, was prepared in two formulations and analysed for its mechanical strength and antibiotic elution.The bone substitute PerOssal has osteoconductive and degradable properties. The material has a built-in capillary structure, which results in an immediate fluid uptake. Antibiotics absorbed to the bone substitute resulted in a prolonged release rate. Mechanical strength was investigated by an unconfined compression test up to failure under both wet and dry conditions for both formulations of the bone substitute. Antibiotic release was analysed microbiologically for two antibiotics, vancomycin and gentamicin, over an elution period of 10 days using the agar diffusion method.The drug release analysis resulted in a prolonged release rate of both antibiotics over 10 days. In vitro the amount of gentamicin and vancomycin eluted at day 10. From one pellet still exceeded the minimal inhibitory concentration of most aetiologically important pathogens. Formulation two of the present bone substitute is significantly harder in both wet and dry conditions when compared to formulation one. Both formulations lose strength in the wet condition relative to their performance in the dry condition. However, formulation two is as hard under wet conditions as formulation one is when dry.PerOssal is a suitable new degradable osteoconductive bone substitute that can be loaded with antibiotic solutions, which are released in effective doses over 10 days. The mechanical strength of PerOssal is sufficient to support cancellous bone defects in non-weight-bearing areas or in combination with osteosynthesis. KW - Knochenersatz KW - Leitfähgkeit KW - Antibiotika KW - Biomechanik KW - Bone substitute materials KW - Osteoconductive KW - Osteomyelitis KW - Multiresistant KW - Biomechanics Y1 - 2007 U6 - https://doi.org/10.1007/s00113-007-1229-3 VL - 110 IS - 5 SP - 408 EP - 413 PB - Springer ER - TY - CHAP A1 - Kubowitsch, Simone A1 - Süß, Franz A1 - Jansen, Petra A1 - Dendorfer, Sebastian T1 - Comparison of dynamic muscular imbalances in back pain patients and healthy controls T2 - World Congress Biomechanics Dublin, 2018 Y1 - 2018 ER - TY - JOUR A1 - Morag, Sarah A1 - Kieninger, Martin A1 - Eissnert, Christoph A1 - Auer, Simon A1 - Dendorfer, Sebastian A1 - Popp, Daniel A1 - Hoffmann, Johannes A1 - Kieninger, Bärbel T1 - Comparison of different techniques for prehospital cervical spine immobilization: Biomechanical measurements with a wireless motion capture system JF - PLOS ONE Y1 - 2023 U6 - https://doi.org/10.1371/journal.pone.0292300 VL - 18 IS - 11 SP - 1 EP - 14 PB - PLOS CY - San Francisco, California ER - TY - GEN A1 - Auer, Simon A1 - Reinker, Lukas A1 - Süß, Franz A1 - Dendorfer, Sebastian T1 - Comparing calculated and measured muscle activity of thigh muscles in dynamic motion. T2 - 27th Congress of the European Society of Biomechanics, 26 - 29 June 2022, Porto, Portugal Y1 - 2022 UR - https://drive.google.com/uc?id=1RBguxyHZE-Wr2y6ktOWK06_3lQg2M9Rb&export=download&confirm=t SP - 640 ER - TY - CHAP A1 - Gross, Simon A1 - Verkerke, Gijsbertus Jacob A1 - Dendorfer, Sebastian T1 - Combined Experimental and Numerical Approach to Investigate Changes in Muscle Recruitment Pattern of the Back Muscles during Exhausting Exercise T2 - World Congress Biomechanics Dublin, 2018 N2 - In recent years, musculoskeletal computation has become a widely used tool to investigate joint and muscle forces within the human body. However, the issue of muscle fatigue is not considered adequately in most models and is a challenging task. One aspect that needs to be examined is the interaction of muscles during an exhausting task. Therefore, an experimental study was designed to analyze the changes of back muscle recruitment pattern during such exercises. In this study 38 subjects (27 male, 11 female, height = 177±8.5 cm, weight = 74.0±13.6 kg) participated. Each subject had to perform three static and three dynamic exhausting exercises where the back muscles were loaded with subject specific forces using a dynamometer adapter especially designed for the trunk muscles. To collect the muscle activity, twelve surface electromyography sensors were applied on the back, and four on the abdominal muscles. Muscle activity and fatigue were analyzed by calculating the maximum voluntary contraction normalized signal and the median frequency. At first the fatigue of m. erector spinae and m. multifidi was analyzed, since these muscles carry the main load during the exercises. Subsequently the activity of the m. trapezius, m. rectus abdominis and m. obliquus externus were investigated to determine recruitment patterns. To gain more detailed information of these patterns a numerical model was built using the AnyBody Modeling System™. Analyzing the measurements, we can observe an increasing muscle activity during isokinetic exercises while the force is constant. Since the activity in the simulation is defined as the current force output divided by the strength of the muscle, the strength parameter was scaled down based on the measured data, assuming a linear force – activity correlation, and using a numerical algorithm considering the influence of cross talk. The results show, that changes in recruitment pattern can be divided into three major subgroups. Prior to total exhaustion, some of the subjects show additional activation of muscles in the trapezius region, while other subjects show an additional activation of abdominal muscles, increasing the intra-abdominal pressure which supports the spine. In the third group an activation in both regions can be observed. The numerical simulations show an increasing activity of abdominal muscles as well as muscles in the upper back. Especially the m. latissimus dorsi shows a significantly higher activity. The results lead to the conclusion that prior to total exhaustion, additional muscles are recruited to support the main muscles. It was shown that abdominal muscles are activated to support back muscles by pressurizing the trunk cavity to delay total exhaustion as long as possible. In conclusion, the results show that changes in muscle recruitment pattern need to be considered when introducing muscle fatigue to musculoskeletal models. Y1 - 2018 ER - TY - GEN A1 - Auer, Simon A1 - Niebler, Michael A1 - Eiglsperger, Josef A1 - Kubowitsch, Simone A1 - Renkawitz, Tobias A1 - Achenbach, Leonard A1 - Krutsch, Werner A1 - Dendorfer, Sebastian T1 - Cognitive stress increases muscle forces in dynamic football specific movements T2 - European Society of Biomechanics meeting 2019, Vienna, Austria Y1 - 2019 UR - https://owncloud.tuwien.ac.at/index.php/s/dovqqcj02VeZHze ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Feldotto, Benedikt A1 - Walch, Blasius A1 - Koch, Patrick A1 - Knoll, Alois T1 - Co-Development of an Infant Prototype in Hardware and Simulation based on CT Imaging Data T2 - IEEE International Conference on Cyborg and Bionic Systems (CBS), 2019, Munich N2 - The development of biomimetic robots has gained research interest in the last years as it may both help under-standing processes of motion execution in biological systems as well as developping a novel generation of intelligent and energy efficient robots. However, exact model generation that builds up on observations and robot design is very time intensive. In this paper we present a novel pipeline for co-development of biomimetic hardware and simulation models based on biological Computer Tomography (CT) data. For this purpose we exploit State of the Art rapid prototyping technologies such as 3D Printing and the Neurorobotics Platform for musculoskeletal simulations in virtual environments. The co-development integrates both advantages of virtual and physical experimental models and is expected to increase development speed of controllers that can be tested on the simulated counterpart before application to a printed robot model. We demonstrate the pipeline by generating a one year old infant model as a musculoskeletal simulation model and a print-in-place 3D printed skeleton as a single movable part. Even though we hereonly introduce the initial body generation and only a first testsetup for a modular sensory and control framework, we can clearly spot advantages in terms of rapid model generation and highly biological related models. Engineering costs are reducedand models can be provided to a wide research community for controller testing in an early development phase. KW - Biomechanische Analyse KW - Robotik KW - Computertomographie KW - 3D-Druck Y1 - 2019 UR - http://mediatum.ub.tum.de/doc/1520042/985669104579.pdf ER - TY - CHAP A1 - Weber, Tim A1 - Dendorfer, Sebastian A1 - Renkawitz, Tobias A1 - Dullien, Silvia A1 - Grifka, Joachim T1 - Clinical gait analysis combined with musculoskeletal modelling – coding a new generation of evaluation instruments T2 - Deutsche Gesellschaft für Biomechanik, Murnau, 2011 Y1 - 2011 ER - TY - JOUR A1 - Dendorfer, Sebastian A1 - Hammer, Joachim A1 - Lenich, Andreas T1 - Characterisation and testing of biomaterials JF - Technology and Health Care KW - Testing biomaterials KW - Characterisation testing KW - Biomaterial KW - Test Y1 - 2011 U6 - https://doi.org/10.3233/THC-2011-0644 VL - 19 IS - 5 SP - 357 EP - 371 ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Hammer, Joachim A1 - Lenich, Andreas ED - Lee, T. Clive ED - Niederer, Peter F. T1 - Characterisation and testing of biomaterials T2 - Basic Engineering for Medics and Biologists: An ESEM Primer KW - Biomaterial KW - Werkstoff KW - Test Y1 - 2010 SN - 978-1607505266 IS - Chapter IV.4. PB - Non Basic Stock Line ER - TY - JOUR A1 - Pfeifer, Christian A1 - Müller, Michael A1 - Prantl, Lukas A1 - Berner, Arne A1 - Dendorfer, Sebastian A1 - Englert, Carsten T1 - Cartilage labelling for mechanical testing in T-peel configuration JF - International Orthopaedics N2 - PURPOSE: The purpose of this study was to find a suitable method of labelling cartilage samples for the measurement of distraction distances in biomechanical testing. METHODS: Samples of bovine cartilage were labelled using five different methods: hydroquinone and silver nitrate (AgNO3), potassium permanganate (KMnO4) with sodium thiosulphate (Na2S2O3), India ink, heat, and laser energy. After the labelling, we analysed the cartilage samples with regard to cytotoxity by histochemical staining with ethidiumbromide homodimer (EthD-1) and calcein AM. Furthermore, we tested cartilages labelled with India ink and heat in a T-peel test configuration to analyse possible changes in the mechanical behaviour between marked and unlabelled samples. RESULTS: Only the labelling methods with Indian ink or a heated needle showed acceptable results in the cytotoxity test with regard to labelling persistence, accuracy, and the influence on consistency and viability of the chondrocytes. In the biomechanical T-peel configuration, heat-labelled samples collapsed significantly earlier than unlabelled samples. CONCLUSION: Labelling bovine cartilage samples with Indian ink in biomechanical testing is a reliable, accurate, inexpensive, and easy-to-perform method. This labelling method influenced neither the biomechanical behaviour nor the viability of the tissue compared to untreated bovine cartilage. KW - Bovine cartilage KW - Cartilage samples KW - Indian ink KW - T-peel configuration KW - Method labeling KW - Knorpel KW - Rind KW - Kennzeichnung KW - Tinte KW - Biomechanik KW - Prüfung Y1 - 2012 U6 - https://doi.org/10.1007/s00264-011-1468-3 VL - 36 IS - 7 SP - 1493 EP - 1499 PB - Springer ER - TY - GEN A1 - Dendorfer, Sebastian T1 - Biomechanische Aspekte bei Sportverletzungen T2 - OT World Leipzig, Internationale Fachmesse und Weltkongress für Orthopädie und Rehatechnik, 10.-13. Mai 2022 Y1 - 2022 ER -