TY - JOUR A1 - Pfeifer, Christian A1 - Müller, Michael A1 - Prantl, Lukas A1 - Berner, Arne A1 - Dendorfer, Sebastian A1 - Englert, Carsten T1 - Cartilage labelling for mechanical testing in T-peel configuration JF - International Orthopaedics N2 - PURPOSE: The purpose of this study was to find a suitable method of labelling cartilage samples for the measurement of distraction distances in biomechanical testing. METHODS: Samples of bovine cartilage were labelled using five different methods: hydroquinone and silver nitrate (AgNO3), potassium permanganate (KMnO4) with sodium thiosulphate (Na2S2O3), India ink, heat, and laser energy. After the labelling, we analysed the cartilage samples with regard to cytotoxity by histochemical staining with ethidiumbromide homodimer (EthD-1) and calcein AM. Furthermore, we tested cartilages labelled with India ink and heat in a T-peel test configuration to analyse possible changes in the mechanical behaviour between marked and unlabelled samples. RESULTS: Only the labelling methods with Indian ink or a heated needle showed acceptable results in the cytotoxity test with regard to labelling persistence, accuracy, and the influence on consistency and viability of the chondrocytes. In the biomechanical T-peel configuration, heat-labelled samples collapsed significantly earlier than unlabelled samples. CONCLUSION: Labelling bovine cartilage samples with Indian ink in biomechanical testing is a reliable, accurate, inexpensive, and easy-to-perform method. This labelling method influenced neither the biomechanical behaviour nor the viability of the tissue compared to untreated bovine cartilage. KW - Bovine cartilage KW - Cartilage samples KW - Indian ink KW - T-peel configuration KW - Method labeling KW - Knorpel KW - Rind KW - Kennzeichnung KW - Tinte KW - Biomechanik KW - Prüfung Y1 - 2012 U6 - https://doi.org/10.1007/s00264-011-1468-3 VL - 36 IS - 7 SP - 1493 EP - 1499 PB - Springer ER - TY - JOUR A1 - Englert, Carsten A1 - Angele, Peter A1 - Fierlbeck, J. A1 - Dendorfer, Sebastian A1 - Schubert, T. A1 - Müller, R. A1 - Lienhard, S. A1 - Zellner, Johannes A1 - Nerlich, Michael A1 - Neumann, Carsten T1 - Conductive bone substitute material with variable antibiotic delivery JF - Der Unfallchirurg N2 - A new bone substitute, consisting of hydroxylapatite and calcium sulphate, was prepared in two formulations and analysed for its mechanical strength and antibiotic elution.The bone substitute PerOssal has osteoconductive and degradable properties. The material has a built-in capillary structure, which results in an immediate fluid uptake. Antibiotics absorbed to the bone substitute resulted in a prolonged release rate. Mechanical strength was investigated by an unconfined compression test up to failure under both wet and dry conditions for both formulations of the bone substitute. Antibiotic release was analysed microbiologically for two antibiotics, vancomycin and gentamicin, over an elution period of 10 days using the agar diffusion method.The drug release analysis resulted in a prolonged release rate of both antibiotics over 10 days. In vitro the amount of gentamicin and vancomycin eluted at day 10. From one pellet still exceeded the minimal inhibitory concentration of most aetiologically important pathogens. Formulation two of the present bone substitute is significantly harder in both wet and dry conditions when compared to formulation one. Both formulations lose strength in the wet condition relative to their performance in the dry condition. However, formulation two is as hard under wet conditions as formulation one is when dry.PerOssal is a suitable new degradable osteoconductive bone substitute that can be loaded with antibiotic solutions, which are released in effective doses over 10 days. The mechanical strength of PerOssal is sufficient to support cancellous bone defects in non-weight-bearing areas or in combination with osteosynthesis. KW - Knochenersatz KW - Leitfähgkeit KW - Antibiotika KW - Biomechanik KW - Bone substitute materials KW - Osteoconductive KW - Osteomyelitis KW - Multiresistant KW - Biomechanics Y1 - 2007 U6 - https://doi.org/10.1007/s00113-007-1229-3 VL - 110 IS - 5 SP - 408 EP - 413 PB - Springer ER - TY - JOUR A1 - Lenich, Andreas A1 - Bachmeier, S. A1 - Dendorfer, Sebastian A1 - Mayr, E. A1 - Nerlich, Michael A1 - Füchtmeier, Bernd T1 - Development of a test system to analyze different hip fracture osteosyntheses under simulated walking JF - Biomedizinische Technik. Biomedical engineering N2 - The mechanical complications of osteosyntheses after hip fractures are previously investigated by mostly static or dynamic uniaxial loading test systems. However, the physiologic loading of the hip joint during a normal gait is a multiplanar, dynamic movement. Therefore, we constructed a system to test osteosyntheses for hip fractures under physiologic multiplanar loading representative of normal gait. To evaluate the testing system, 12 femora pairs were tested under 25,000 cycles with two standard osteosyntheses (Proximal Femoral Nail Antirotation/Gamma3 Nail). For angular movement, the varus collapse to cut out (∝CO) (∝CO=4.8°±2.1° for blade and ∝CO=7.8°±3.8° for screw) was the dominant failure mode, and only slight rotational angle shifts (∝Rot) (∝Rot=1.7°±0.4° for blade and ∝Rot=2.4°±0.3° for screw) of the femoral head around the implant axis were observed. Angular displacements in varus direction and rotation were higher in specimens reinforced with screws. Hence, the cut out model and the migration directions showed a distinction between helical blade and hip screw. However, there were no significant differences between the different implants. The new setup is able to create clinical failures and allows to give evidence about the anchorage stability of different implant types under dynamic gait motion pattern. KW - biomechanical implant test KW - dynamic multiplanar loading KW - hip fractureg KW - implant migration KW - patient specific loading KW - Hüftgelenk KW - Knochenbruch KW - Osteosynthese KW - Bewegungsanalyse KW - Simulation Y1 - 2012 U6 - https://doi.org/10.1515/bmt-2011-0999 VL - 57 IS - 2 SP - 113 EP - 119 ER - TY - JOUR A1 - Dendorfer, Sebastian A1 - Maier, Hans Jürgen A1 - Hammer, Joachim T1 - Fatigue damage in cancellous bone: an experimental approach from continuum to micro scale JF - Journal of the Mechanical Behavior of Biomedical Materials N2 - Repeated loadings may cause fatigue fractures in bony structures. Even if these failure types are known, data for trabecular bone exposed to cyclic loading are still insufficient as the majority of fatigue analyses on bone concentrate on cortical structures. Despite its highly anisotropic and inhomogeneous structure, trabecular bone is treated with continuum approaches in fatigue analyses. The underlying deformation and damage mechanism within trabecular specimens are not yet sufficiently investigated. In the present study different types of trabecular bone were loaded in monotonic and cyclic compression. In addition to the measurement of integral specimen deformations, optical deformation analysis was employed in order to obtain strain distributions at different scale levels, from the specimens' surface to the trabeculae level. These measurements allowed for the possibility of linking the macroscopic and microscopic mechanical behaviour of cancellous bone. Deformations were found to be highly inhomogeneous across the specimen. Furthermore strains were found to already localise at very low load levels and after few load cycles. Microcracks in individual trabeculae were induced in the very early stage of cyclic testing. The results provide evidence of the capability of the method to supply essential data on the failure behaviour of individual trabeculae in future studies. KW - Knochen KW - Spongiosa KW - Ermüdungsbruch KW - Fatigue KW - Cancellous bone KW - Deformation behaviour KW - Damage KW - Measurement and testing KW - Mechanical behaviour Y1 - 2009 U6 - https://doi.org/10.1016/j.jmbbm.2008.03.003 VL - 2 IS - 1 SP - 113 EP - 119 ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Maier, Hans Jürgen A1 - Hammer, Joachim T1 - How do age and anisotropy affect the fatigue behaviour of cancellous bone? T2 - Medicine Meets Engineering N2 - The fatigue behaviour of materials is of particular interest for the failure prediction of materials and structures exposed to cyclic loading. For trabecular bone structures only a few sets of lifetime data have been reported in the literature and structural measures are commonly not considered. The influence of load contributions not aligned with the main physiological axis remains unclear. Furthermore age effects on the fatigue behaviour are not well described. In the present study, different groups of human vertebral cancellous bone were exposed to cyclic compression. The inital modulus and therefore lifetimes were found to be highly dependent on age. The decrease in both with increasing age was much more pronounced in specimens which were not aligned with the main physiological axis. This implies that old bone is much more sensitive to (cyclic) failure loads in general but particularly to loads which are not coincident with the physiological main axis. KW - Knochenbruch KW - Spongiosa KW - Alter KW - Anisotropie Y1 - 2008 SP - 68 EP - 74 PB - IOS Press ER - TY - JOUR A1 - Hoenicka, Markus A1 - Lehle, Karla A1 - Jacobs, V. R. A1 - Dendorfer, Sebastian A1 - Kostorz, A. A1 - Schmid, F. X. A1 - Birnbaum, D. E. T1 - Mechanical and seeding properties of human umbilical vein – a potential scaffold for a tissue-engineered vessel graft JF - The Thoracic and Cardiovascular Surgeon N2 - Objectives: The mechanical properties and seeding with endothelial cells were investigated in fresh and cryopreserved human umbilical vein. Methods: Human umbilical veins (HUV) were frozen in Euro-Collins/1M DMSO at –1°C/min and stored in liquid nitrogen. Stress-strain relationships of fresh and thawed veins were determined in an uniaxial tension-testing rig. HUV endothelial cells (HUVEC) were seeded onto denuded HUV under static conditions and grown for 3d. Luminal surfaces were analyzed by scanning electron microscopy. Calcein-stained cells were seeded hyperconfluently to determine the cell retention capacity of fresh and cryopreserved veins. Results: The stress-strain relationships of HUV followed a biphasic pattern typical for natural vessels. Neither the failure stress (2.71±0.36 vs. 3.25±0.97 N, n=3) nor the displacement required to achieve failure (9.73±0.9 vs. 7.43±2.07mm, n=3) were altered by cryopreservation. The burst pressure was estimated as approx. 1000mm Hg within the limitations of the uniaxial model. HUVEC seeded onto denuded HUV formed patches (at 9E3 cells per cm2) or an almost confluent endothelium (at 3E4 cells per cm2) within three days. The capacity to retain seeded HUVEC of denuded HUV was not altered by cryopreservation (1.15±0.08E5 vs. 1.26±0.14E5 cells per cm2, n=6). Conclusions: The burst pressure of HUV seems to be sufficiently high for the human arterial circulation and is not altered by cryopreservation. HUVEC can establish a confluent endothelium on denuded HUV. Therefore HUV appears to be a suitable storable scaffold for vascular tissue engineering. KW - Nabelvene KW - Tissue Engineering Y1 - 2007 U6 - https://doi.org/10.1055/s-2007-967592 VL - 55 IS - S 1 SP - P_37 PB - Thieme ER - TY - JOUR A1 - Wong, Christian A1 - Rasmussen, John A1 - Simonsen, Erik B. A1 - Hansen, Lone A1 - de Zee, Mark A1 - Dendorfer, Sebastian T1 - The Influence of Muscle Forces on the Stress Distribution in the Lumbar Spine JF - The Open Spine Journal N2 - Introduction: Previous studies of bone stresses in the human lumbar spine have relied on simplified models when modeling the spinal musculature, even though muscle forces are likely major contributors to the stresses in the vertebral bones. Detailed musculoskeletal spine models have recently become available and show good correlation with experimental findings. A combined inverse dynamics and finite element analysis study was conducted in the lumbar spine to investigate the effects of muscle forces on a detailed musculoskeletal finite element model of the 4th lumbar vertebral body. Materials and Methodology: The muscle forces were computed with a detailed and validated inverse dynamics musculoskeletal spine model in a lifting situation, and were then applied to an orthotropic finite element model of the 4th lumbar vertebra. The results were compared with those from a simplified load case without muscles. Results: In general the von Mises stress was larger by 30%, and even higher when looking at the von Mises stress distribution in the superio-anterior and central part of the vertebral body and in the pedicles. Conclusion: The application of spine muscles to a finite element model showed markedly larger von Mises stress responses in the central and anterior part of the vertebral body, which can be tolerated in the young and healthy spine, but it would increase the risk of compression fractures in the elderly, osteoporotic spine. KW - Lumbar spine KW - Muscle influence KW - Inverse dynamics KW - Finite element analysis KW - Lendenwirbelsäule KW - Muskelkraft KW - Belastung KW - Finite-Elemente-Methode Y1 - 2011 U6 - https://doi.org/10.2174/1876532701103010021 VL - 3 IS - 1 SP - 21 EP - 26 ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Hammer, Joachim A1 - Lenich, Andreas ED - Lee, T. Clive ED - Niederer, Peter F. T1 - Characterisation and testing of biomaterials T2 - Basic Engineering for Medics and Biologists: An ESEM Primer KW - Biomaterial KW - Werkstoff KW - Test Y1 - 2010 SN - 978-1607505266 IS - Chapter IV.4. PB - Non Basic Stock Line ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Feldotto, Benedikt A1 - Walch, Blasius A1 - Koch, Patrick A1 - Knoll, Alois T1 - Co-Development of an Infant Prototype in Hardware and Simulation based on CT Imaging Data T2 - IEEE International Conference on Cyborg and Bionic Systems (CBS), 2019, Munich N2 - The development of biomimetic robots has gained research interest in the last years as it may both help under-standing processes of motion execution in biological systems as well as developping a novel generation of intelligent and energy efficient robots. However, exact model generation that builds up on observations and robot design is very time intensive. In this paper we present a novel pipeline for co-development of biomimetic hardware and simulation models based on biological Computer Tomography (CT) data. For this purpose we exploit State of the Art rapid prototyping technologies such as 3D Printing and the Neurorobotics Platform for musculoskeletal simulations in virtual environments. The co-development integrates both advantages of virtual and physical experimental models and is expected to increase development speed of controllers that can be tested on the simulated counterpart before application to a printed robot model. We demonstrate the pipeline by generating a one year old infant model as a musculoskeletal simulation model and a print-in-place 3D printed skeleton as a single movable part. Even though we hereonly introduce the initial body generation and only a first testsetup for a modular sensory and control framework, we can clearly spot advantages in terms of rapid model generation and highly biological related models. Engineering costs are reducedand models can be provided to a wide research community for controller testing in an early development phase. KW - Biomechanische Analyse KW - Robotik KW - Computertomographie KW - 3D-Druck Y1 - 2019 UR - http://mediatum.ub.tum.de/doc/1520042/985669104579.pdf ER - TY - CHAP A1 - Dendorfer, Sebastian A1 - Kubowitsch, Simone A1 - Süß, Franz T1 - How to determine the effect of working conditions on the human body T2 - 11th International Scientific Conference on Production Engineering DEVELOPMENT AND MODERNIZATION OF PRODUCTION (RIM 2017), Sarajevo, Bosnia and Herzegovina N2 - Work places and conditions strains the human body, both psychologically and biomechanically. In order to analyse working conditions and in the following to improve them, detailed knowledge about the effect of the different stressors on the body is needed. This manuscript discusses methods on how to evaluate biomechanical and mental loading and its effect on the musculoskeletal system. A possible workflow for the analysis is presented. KW - Arbeitsbedingungen KW - Körper KW - Mensch Y1 - 2017 UR - https://tfb.ba/repozitorij/2/RIM/RIM2017/01A-Sebastian%20Dendorfer,%20Simone%20Kubowitsch,%20Franz%20S%C3%BC%C3%9F-HOW%20TO%20DETERMINE%20THE%20EFFECT%20OF%20WORKING%20CONDITIONS%20ON%20THE%20HUMAN%20BODY.pdf ER - TY - JOUR A1 - Benditz, Achim A1 - Auer, Simon A1 - Spörrer, J.F. A1 - Wolkerstorfer, S. A1 - Grifka, Joachim A1 - Süß, Franz A1 - Dendorfer, Sebastian T1 - Regarding loads after spinal fusion, every level should be seen separately: a musculoskeletal analysis JF - European Spine Journal N2 - The number of spinal fusion surgeries is steadily increasing and biomechanical consequences are still in debate. The aim of this study is to provide biomechanical insights into the sagittal balance of the spine and to compare spinal load before and after spinal fusion. METHOD: The joint reaction forces of 52 patients were analyzed in proximo-distal and antero-posterior direction from the levels T12-L1 to L5-S1 using musculoskeletal simulations. RESULTS: In 104 simulations, pre-surgical forces were equal to post-surgical. The levels L4-L5 and T12-L1, however, showed increased spinal forces compression forces with higher sagittal displacement. Improved restauration of sagittal balance was accompanied by lower spinal load. AP shear stress, interestingly decreased with sagittal imbalance. CONCLUSION: Imbalanced spines have a risk of increased compression forces at Th12-L1. L4-L5 always has increased spinal loads. These slides can be retrieved under Electronic Supplementary Material. KW - AnyBody Modeling System KW - Musculoskeletal analysis KW - Sagittal balance KW - Spinal fusion KW - Spine biomechanics KW - Biomechanische Analyse KW - Wirbelsäulenversteifung KW - Vergleichende Anatomie Y1 - 2018 U6 - https://doi.org/10.1007/s00586-018-5476-5 VL - 27 IS - 8 SP - 1905 EP - 1910 PB - Springer-Verlag ER - TY - JOUR A1 - Völlner, Florian A1 - Weber, Tim A. A1 - Weber, Markus A1 - Renkawitz, Tobias A1 - Dendorfer, Sebastian A1 - Grifka, Joachim A1 - Craiovan, Benjamin T1 - A simple method for determining ligament stiffness during total knee arthroplasty in vivo JF - Scientific Reports N2 - A key requirement in both native knee joints and total knee arthroplasty is a stable capsular ligament complex. However, knee stability is highly individual and ranges from clinically loose to tight. So far, hardly any in vivo data on the intrinsic mechanical of the knee are available. This study investigated if stiffness of the native ligament complex may be determined in vivo using a standard knee balancer. Measurements were obtained with a commercially available knee balancer, which was initially calibrated in vitro. 5 patients underwent reconstruction of the force-displacement curves of the ligament complex. Stiffness of the medial and lateral compartments were calculated to measure the stability of the capsular ligament complex. All force-displacement curves consisted of a non-linear section at the beginning and of a linear section from about 80 N onwards. The medial compartment showed values of 28.4 ± 1.2 N/mm for minimum stiffness and of 39.9 ± 1.1 N/mm for maximum stiffness; the respective values for the lateral compartment were 19.9 ± 0.9 N/mm and 46.6 ± 0.8 N/mm. A commercially available knee balancer may be calibrated for measuring stiffness of knee ligament complex in vivo, which may contribute to a better understanding of the intrinsic mechanical behaviour of knee joints. KW - Biomechanische Analyse KW - Kniegelenkband KW - Steifigkeit Y1 - 2019 U6 - https://doi.org/10.1038/s41598-019-41732-x VL - 9 SP - 1 EP - 8 PB - Nature ER - TY - JOUR A1 - Putzer, Michael A1 - Ehrlich, Ingo A1 - Rasmussen, John A1 - Gebbeken, Norbert A1 - Dendorfer, Sebastian T1 - Sensitivity of lumbar spine loading to anatomical parameters JF - Journal of Biomechanics N2 - Musculoskeletal simulations of lumbar spine loading rely on a geometrical representation of the anatomy. However, this data has an inherent inaccuracy. This study evaluates the influence of defined geometrical parameters on lumbar spine loading utilising five parametrised musculoskeletal lumbar spine models for four different postures. The influence of the dimensions of vertebral body, disc, posterior parts of the vertebrae as well as the curvature of the lumbar spine was studied. Additionally, simulations with combinations of selected parameters were conducted. Changes in L4/L5 resultant joint force were used as outcome variable. Variations of the vertebral body height, disc height, transverse process width and the curvature of the lumbar spine were the most influential. These parameters can be easily acquired from X-rays and should be used to morph a musculoskeletal lumbar spine model for subject-specific approaches with respect to bone geometry. Furthermore, the model was very sensitive to uncommon configurations and therefore, it is advised that stiffness properties of discs and ligaments should be individualised. KW - Musculoskeletal simulation KW - Lumbar spine KW - Parameter study KW - Vertebra KW - Wirbelsäule KW - Belastung KW - Simulation Y1 - 2015 U6 - https://doi.org/10.1016/j.jbiomech.2015.11.003 VL - 49 IS - 6 SP - 953 EP - 958 PB - Elsevier Science ER - TY - JOUR A1 - Dendorfer, Sebastian A1 - Hammer, Joachim A1 - Lenich, Andreas T1 - Characterisation and testing of biomaterials JF - Technology and Health Care KW - Testing biomaterials KW - Characterisation testing KW - Biomaterial KW - Test Y1 - 2011 U6 - https://doi.org/10.3233/THC-2011-0644 VL - 19 IS - 5 SP - 357 EP - 371 ER - TY - JOUR A1 - Hölscher, Thomas A1 - Weber, Tim A. A1 - Lazarev, Igor A. A1 - Englert, Carsten A1 - Dendorfer, Sebastian T1 - The influence of rotator cuff tears on glenohumeral stability during abduction tasks JF - Journal of Orthopaedic Research N2 - One of the main goals in reconstructing rotator cuff tears is the restoration of glenohumeral joint stability, which is subsequently of utmost importance in order to prevent degenerative damage such as superior labral anterior posterior (SLAP) lesion, arthrosis, and malfunction. The goal of the current study was to facilitate musculoskeletal models in order to estimate glenohumeral instability introduced by muscle weakness due to cuff lesions. Inverse dynamics simulations were used to compute joint reaction forces for several static abduction tasks with different muscle weakness. Results were compared with the existing literature in order to ensure the model validity. Further arm positions taken from activities of daily living, requiring the rotator cuff muscles were modeled and their contribution to joint kinetics computed. Weakness of the superior rotator cuff muscles (supraspinatus; infraspinatus) leads to a deviation of the joint reaction force to the cranial dorsal rim of the glenoid. Massive rotator cuff defects showed higher potential for glenohumeral instability in contrast to single muscle ruptures. The teres minor muscle seems to substitute lost joint torque during several simulated muscle tears to maintain joint stability. Joint instability increases with cuff tear size. Weakness of the upper part of the rotator cuff leads to a joint reaction force closer to the upper glenoid rim. This indicates the comorbidity of cuff tears with SLAP lesions. The teres minor is crucial for maintaining joint stability in case of massive cuff defects and should be uprated in clinical decision-making. KW - Abduction tasks KW - Glenohumeral stability KW - Musculoskeletal Modeling KW - Muscle weakness KW - Rotator cuff tears KW - Rotatorenmanschettenriss KW - Schultergelenk KW - Stabilität Y1 - 2016 U6 - https://doi.org/10.1002/jor.23161 VL - 34 IS - 9 SP - 1628 EP - 1635 ER - TY - JOUR A1 - Putzer, Michael A1 - Auer, Stefan A1 - Malpica, William A1 - Süß, Franz A1 - Dendorfer, Sebastian T1 - A numerical study to determine the effect of ligament stiffness on kinematics of the lumbar spine during flexion JF - BMC Musculoskeletal Disorders N2 - Background There is a wide range of mechanical properties of spinal ligaments documented in literature. Due to the fact that ligaments contribute in stabilizing the spine by limiting excessive intersegmental motion, those properties are of particular interest for the implementation in musculoskeletal models. The aim of this study was to investigate the effect of varying ligament stiffness on the kinematic behaviour of the lumbar spine. Methods A musculoskeletal model with a detailed lumbar spine was modified according to fluoroscopic recordings and corresponding data files of three different subjects. For flexion, inverse dynamics analysis with a variation of the ligament stiffness matrix were conducted. The influence of several degrees of ligament stiffness on the lumbar spine model were investigated by tracking ligament forces, disc forces and resulting moments generated by the ligaments. Additionally, the kinematics of the motion segments were evaluated. Results An increase of ligament stiffness resulted in an increase of ligament and disc forces, whereas the relative change of disc force increased at a higher rate at the L4/L5 level (19 %) than at the L3/L4 (10 %) level in a fully flexed posture. The same behaviour applied to measured moments with 67 % and 45 %. As a consequence, the motion deflected to the lower levels of the lumbar spine and the lower discs had to resist an increase in loading. Conclusions Higher values of ligament stiffness over all lumbar levels could lead to a shift of the loading and the motion between segments to the lower lumbar levels. This could lead to an increased risk for the lower lumbar parts. KW - Lumbar spine KW - Ligament stiffness KW - Musculoskeletal Modeling KW - Biomechanics Y1 - 2016 U6 - https://doi.org/10.1186/s12891-016-0942-x VL - 17 IS - 95 ER - TY - JOUR A1 - Schmitz, Paul A1 - Neumann, Christoph Cornelius A1 - Neumann, Carsten A1 - Nerlich, Michael A1 - Dendorfer, Sebastian T1 - Biomechanical analysis of iliac crest loading following cortico-cancellous bone harvesting JF - Journal of Orthopaedic Surgery and Research N2 - Background Iliac crest bone harvesting is a frequently performed surgical procedure widely used to treat bone defects. The objective of this study is to assess the biomechanical quantities related to risk for pelvic fracture after harvesting an autologous bone graft at the anterior iliac crest. Methods Finite element models with a simulated harvest site (sized 15 × 20 mm, 15 × 35 mm, 30 × 20 mm and 30 × 35 mm) in the iliac wing are created. The relevant loading case is when the ipsilateral leg is lifted off the ground. Musculoskeletal analysis is utilized to compute the muscle and joint forces involved in this motion. These forces are used as boundary conditions for the finite element analyses. Bone tissue stress is analyzed. Results Critical stress peaks are located between the anterior superior iliac spine (ASIS) and the anterior edge of the harvest site. Irrespective of the graft size, the iliac wing does not show any significant stress peaks with the harvest site being 20 to 25 mm posterior to the ASIS. The harvest area itself inhibits the distribution of the forces applied on the ASIS to extend to the posterior iliac wing. This leads to a lack of stress posterior to the harvest site. A balanced stress distribution with no stress peaks appears when the bone graft is taken below the iliac crest. Conclusion A harvest site located at least 20 to 25 mm posterior to the ASIS should be preferred to minimize the risk of iliac fatigue fracture. KW - Bone harvesting KW - Autologous bone graft KW - Iliac crest KW - Fatigue fracture KW - Pelvis KW - ASIS KW - FEA KW - Biomechanical investigation KW - Beckenkammknochen KW - Knochenentnahme KW - Spongiosa KW - Biomechanische Analyse Y1 - 2018 U6 - https://doi.org/10.1186/s13018-018-0822-1 VL - 13 IS - 108 SP - 1 EP - 8 PB - Springer Nature ER - TY - JOUR A1 - Renkawitz, Tobias A1 - Weber, Tim A. A1 - Dullien, Silvia A1 - Woerner, Michael A1 - Dendorfer, Sebastian A1 - Grifka, Joachim A1 - Weber, Markus T1 - Leg length and offset differences above 5 mm after total hip arthroplasty are associated with altered gait kinematics JF - Gait & Posture N2 - We aimed to investigate the relationship between postoperative leg length/offset (LL/OS) reconstruction and gait performance after total hip arthroplasty (THA). In the course of a prospective randomized controlled trial, 60 patients with unilateral hip arthrosis received cementless THA through a minimally-invasive anterolateral surgical approach. One year post-operatively, LL and global OS restoration were analyzed and compared to the contralateral hip on AP pelvic radiographs. The combined postoperative limb length/OS reconstruction of the operated hip was categorized as restored (within 5 mm) or non-restored (more than 5 mm reduction or more than 5 mm increment). The acetabular component inclination, anteversion and femoral component anteversion were evaluated using CT scans of the pelvis and the femur. 3D gait analysis of the lower extremity and patient related outcome measures (HHS, HOOS, EQ-5D) were obtained pre-operatively, six months and twelve months post-operatively by an observer blinded to radiographic results. Component position of cup and stem was comparable between the restored and non-restored group. Combined LL and OS restoration within 5 mm resulted in higher Froude number (p < 0.001), normalized walking speed (p < 0.001) and hip range-of-motion (ROM) (p = 0.004) during gait twelve months postoperatively, whereas gait symmetry was comparable regardless of LL and OS reconstruction at both examinations. Clinical scores did not show any relevant association between the accuracy of LL or OS reconstruction and gait six/twelve months after THA. In summary, postoperative LL/OS discrepancies larger than 5 mm relate to unphysiological gait kinematics within the first year after THA. DRKS00000739, German Clinical Trials Register. KW - THA KW - Leg length KW - Offset KW - Gait analysis KW - Biomechanics Y1 - 2016 U6 - https://doi.org/10.1016/j.gaitpost.2016.07.011 VL - vol. 49 SP - 196 EP - 201 ER - TY - JOUR A1 - Weber, Tim A. A1 - Dendorfer, Sebastian A1 - Bulstra, Sjoerd K. A1 - Grifka, Joachim A1 - Verkerke, Gijsbertus Jacob A1 - Renkawitz, Tobias T1 - Gait six month and one-year after computer assisted Femur First THR vs. conventional THR. Results of a patient- and observer- blinded randomized controlled trial JF - Gait & Posture N2 - A prospective randomized controlled trial is presented that is used to compare gait performance between the computer assisted Femur First (CAS FF) operation method and conventional THR (CON). 60 patients underwent a 3D gait analysis of the lower extremity at pre-operative, 6 months post-operative and twelve months post-operative. Detailed verification experiments were facilitated to ensure the quality of data as well as to avoid over-interpreting of the data. The results confirm a similar data-quality as reported in the literature. Walking speed, range of motion and symmetry thereof improved over the follow-up period, without significant differences between the groups. While all parameters do significantly increase over the follow-up period for both groups, there were no significant differences between them at any given time-point. Patients undergoing CAS FF showed a trend to improved hip flexion angle indicating a possible long-term benefit. KW - Biomechanics KW - Combined anteversion KW - Computer-assisted surgery KW - Femur first KW - Gait analysis Y1 - 2016 U6 - https://doi.org/10.1016/j.gaitpost.2016.06.035 VL - vol. 49 SP - 418 EP - 425 ER -