@article{WiesentSchultheissLullaetal., author = {Wiesent, Lisa and Schultheiß, Ulrich and Lulla, Philipp and Noster, Ulf and Schratzenstaller, Thomas and Schmid, Christof and Nonn, Aida and Spear, Ashley}, title = {Computational analysis of the effects of geometric irregularities and post-processing steps on the mechanical behavior of additively manufactured 316L stainless steel stents}, series = {PLoS ONE}, volume = {15}, journal = {PLoS ONE}, number = {12}, publisher = {PLOS}, doi = {10.1371/journal.pone.0244463}, pages = {1 -- 30}, abstract = {Advances in additive manufacturing enable the production of tailored lattice structures and thus, in principle, coronary stents. This study investigates the effects of process-related irregularities, heat and surface treatment on the morphology, mechanical response, and expansion behavior of 316L stainless steel stents produced by laser powder bed fusion and provides a methodological approach for their numerical evaluation. A combined experimental and computational framework is used, based on both actual and computationally reconstructed laser powder bed fused stents. Process-related morphological deviations between the as-designed and actual laser powder bed fused stents were observed, resulting in a diameter increase by a factor of 2-2.6 for the stents without surface treatment and 1.3-2 for the electropolished stent compared to the as-designed stent. Thus, due to the increased geometrically induced stiffness, the laser powder bed fused stents in the as-built (7.11 ± 0.63 N) or the heat treated condition (5.87 ± 0.49 N) showed increased radial forces when compressed between two plates. After electropolishing, the heat treated stents exhibited radial forces (2.38 ± 0.23 N) comparable to conventional metallic stents. The laser powder bed fused stents were further affected by the size effect, resulting in a reduced yield strength by 41\% in the as-built and by 59\% in the heat treated condition compared to the bulk material obtained from tensile tests. The presented numerical approach was successful in predicting the macroscopic mechanical response of the stents under compression. During deformation, increased stiffness and local stress concentration were observed within the laser powder bed fused stents. Subsequent numerical expansion analysis of the derived stent models within a previously verified numerical model of stent expansion showed that electropolished and heat treated laser powder bed fused stents can exhibit comparable expansion behavior to conventional stents. The findings from this work motivate future experimental/numerical studies to quantify threshold values of critical geometric irregularities, which could be used to establish design guidelines for laser powder bed fused stents/lattice structures.}, subject = {Koronarendoprothese}, language = {en} } @article{WiesentSchultheissSchmidetal., author = {Wiesent, Lisa and Schultheiss, Ulrich and Schmid, Christof and Schratzenstaller, Thomas and Nonn, Aida}, title = {Experimentally validated simulation of coronary stents considering different dogboning ratios and asymmetric stent positioning}, series = {PlOS One}, volume = {14}, journal = {PlOS One}, number = {10}, publisher = {PLOS}, doi = {10.1371/journal.pone.0224026}, pages = {1 -- 25}, abstract = {In-stent restenosis remains a major problem of arteriosclerosis treatment by stenting. Expansion-optimized stents could reduce this problem. With numerical simulations, stent designs/ expansion behaviours can be effectively analyzed. For reasons of efficiency, simplified models of balloon-expandable stents are often used, but their accuracy must be challenged due to insufficient experimental validation. In this work, a realistic stent life-cycle simulation has been performed including balloon folding, stent crimping and free expansion of the balloon-stent-system. The successful simulation and validation of two stent designs with homogenous and heterogeneous stent stiffness and an asymmetrically positioned stent on the balloon catheter confirm the universal applicability of the simulation approach. Dogboning ratio, as well as the final dimensions of the folded balloon, the crimped and expanded stent, correspond well to the experimental dimensions with only slight deviations. In contrast to the detailed stent life-cycle simulation, a displacement-controlled simulation can not predict the transient stent expansion, but is suitable to reproduce the final expanded stent shape and the associated stress states. The detailed stent life-cycle simulation is thus essential for stent expansion analysis/optimization, whereas for reasons of computational efficiency, the displacement-controlled approach can be considered in the context of pure stress analysis.}, subject = {Stent}, language = {en} } @inproceedings{WiesentGeithWagner, author = {Wiesent, Lisa and Geith, Markus A. and Wagner, Marcus}, title = {Simulation of Fluid-Structure Interaction between injection medium and balloon catheter using ICFD}, series = {11th European LS-DYNA Conference 2017, 9 - 11 May, Salzburg, Austria}, booktitle = {11th European LS-DYNA Conference 2017, 9 - 11 May, Salzburg, Austria}, isbn = {978-3981621549}, abstract = {Arteriosclerosis is a major health issue worldwide. While it is commonly treated by the implantation of an balloon-expandable stent, micro injuries may occur during stent deployment, and induce in-stent restenosis, whose consequence can be fatal. Studying this undesirable phenomenon is usually limited as experimental data is hard to obtain on ethical ground. Numerical simulation are performed to better understand this problem. To construct a more realistic simulation of a balloon-expandable stent, a partitioned strongly-coupled FSI simulation of the balloon deployment was set up using the ICFD solver of LS-DYNA, - a quite innovative approach. The complex balloon configuration as well as the interaction of the injection medium and the balloon structure was considered. The balloon structure consisting of shell elements was obtained from preliminary balloon folding and pleating simulations. The balloon consists of a flexible thin walled polyamide. The injection fluid is implemented using volume elements. Balloon deployment was initiated by a pressure boundary condition inducing a volume flow into the balloon. The initial feasibility analysis showed promising result including a continuous balloon deployment and a reasonable development of the fluid pressure and velocity field. However, applying this FSI approach to a more complex balloon structure led to a non convergent solution. The non-convergence could be mainly reduced to mechanical factors including the low wall thickness of the balloon (< 0.05 mm) and the flexibility of the polyamide. Further, the ICFD solver shows less accuracy concerning the FSI conditions when dealing with thin flexible structures as well as enclosed volumes. A shell thickness of 0.06 mm is believed to result in a convergent solution.}, subject = {Koronare Herzkrankheit}, language = {en} } @misc{WiesentHupkeBalketal., author = {Wiesent, Lisa and Hupke, Constantin and Balk, Christian and Schultheiss, Ulrich and Schratzenstaller, Thomas}, title = {Optimization of the cardiovascular stent design towards improved expansion behaviour and radial stiffness properties}, series = {Biomedizinische Technik}, volume = {63}, journal = {Biomedizinische Technik}, number = {s1}, doi = {10.1515/bmt-2018-6031}, abstract = {- Development of a FEA Tool for a realistic stent simulation - investigation on minor modification on the stent design on the expansion behaviour - analysis of three stent designs: classical stent design with pronounced dogbone effect, two modified stent design (non-dogbone-design)}, subject = {Kardiovaskul{\"a}res System}, language = {en} } @article{WiesentSpearNonn, author = {Wiesent, Lisa and Spear, Ashley and Nonn, Aida}, title = {Computational analysis of the effects of geometric irregularities on the interaction of an additively manufactured 316L stainless steel stent and a coronary artery}, series = {Journal of the Mechanical Behavior of Biomedical Materials}, volume = {125}, journal = {Journal of the Mechanical Behavior of Biomedical Materials}, publisher = {Elsevier}, issn = {1751-6161}, doi = {10.1016/j.jmbbm.2021.104878}, abstract = {Customized additively manufactured (laser powder bed fused (L-PBF)) stents could improve the treatment of complex lesions by enhancing stent-artery conformity. However, geometric irregularities inherent for L-PBF stents are expected to influence not only their mechanical behavior but also their interaction with the artery. In this study, the influence of geometrical irregularities on stent-artery interaction is evaluated within a numerical framework. Thus, computed arterial stresses induced by a reconstructed L-PBF stent model are compared to those induced by the intended stent model (also representing a stent geometry obtained from conventional manufacturing processes) and a modified CAD stent model that accounts for the increased strut thickness inherent for L-PBF stents. It was found that, similar to conventionally manufactured stents, arterial stresses are initially related to the basic stent design/topology, with the highest stresses occurring at the indentations of the stent struts. Compared to the stent CAD model, the L-PBF stent induces distinctly higher and more maximum volume stresses within the plaque and the arterial wall. In return, the modified CAD model overestimates the arterial stresses induced by the L-PBF stent due to its homogeneously increased strut thickness and thus its homogeneously increased geometric stiffness compared with the L-PBF stent. Therefore, the L-PBF-induced geometric irregularities must be explicitly considered when evaluating the L-PBF stent-induced stresses because the intended stent CAD model underestimates the arterial stresses, whereas the modified CAD model overestimates them. The arterial stresses induced by the L-PBF stent were still within the range of values reported for conventional stents in literature, suggesting that the use of L-PBF stents is conceivable in principle. However, because geometric irregularities, such as protruding features from the stent surface, could potentially damage the artery or lead to premature stent failure, further improvement of L-PBF stents is essential.}, language = {en} } @article{WiesentSchultheissLullaetal., author = {Wiesent, Lisa and Schultheiß, Ulrich and Lulla, Philipp and Nonn, Aida and Noster, Ulf}, title = {Mechanical properties of small structures built by selective laser melting 316 L stainless steel - a phenomenological approach to improve component design}, series = {Materials Science \& Engineering Technology}, volume = {51}, journal = {Materials Science \& Engineering Technology}, number = {12}, publisher = {Wiley}, doi = {10.1002/mawe.202000038}, url = {http://nbn-resolving.de/urn:nbn:de:bvb:898-opus4-14718}, pages = {1615 -- 1629}, abstract = {Experimental investigations are conducted to quantify the influence of specimen thickness and orientation on the mechanical properties of selective laser melted stainless steel 316 L. The results indicate that the mechanical strength and ductility increase with increasing specimen thickness until a saturation value is reached from a specimen thickness of about 2 mm. Specimen orientation dependency is pronounced for thin specimens (<1.5 mm), whereas only small deviations in strength are observed for thicker specimens with orientations of 30°, 45° and 90° to build direction. The mechanical properties of the specimen orientation of 0° to build direction shows great deviation to the other orientations and the smallest overall strength. A reliable design of selective laser melted components should account for specimen thickness and orientation, e. g. by a correction factor. Furthermore, it is recommended to avoid loads vertical (90°) and parallel (0°) to build direction to guarantee higher ductility and strength.}, language = {en} } @article{WiesentStockerNonn, author = {Wiesent, Lisa and Stocker, Felix and Nonn, Aida}, title = {Investigating the influence of geometric parameters on the deformation of laser powder bed fused stents using low-fidelity thermo-mechanical analysis}, series = {Materialia}, volume = {28}, journal = {Materialia}, publisher = {Elsevier}, doi = {10.1016/j.mtla.2023.101774}, abstract = {Maintaining dimensional accuracy is a major challenge of laser powder bed fusion (L-PBF) preventing its application for more complex and filigree L-PBF structures in industrial practice. Previous studies have shown that residual stresses and distortion of benchmark L-PBF components may be predicted by sequential thermo-mechanical analyses. However, the reliability of these analyses for more complex structures must be critically questioned, as comprehensive validation and sensitivity analyses are scarce. In this paper, we present a calibrated and validated low-fidelity sequential thermo-mechanical finite element analysis (FEA) of a tubular L-PBF lattice structure, i.e., an aortic stent, where pronounced local deformation is expected. As a first step, the finite element model was extensively calibrated using experimental data to ensure reproducibility of the simulation results. Thereupon, geometric features critical to the distortion of L-PBF lattice structures and measures to compensate for the distortion, such as inversion of the distorted L-PBF structure, were investigated. It was found that the distortion of the L-PBF lattice structures can be reduced, but not completely prevented, by increasing the strut angles, increasing the strut thickness, and decreasing the transition radius in the area of merging struts. FEA-based inversion of the numerically predicted deformed structure minimized distortion, resulting in the L-PBF aortic stent approximating the intended CAD geometry even with a small strut thickness. This work shows that low-fidelity sequential thermo-mechanical FEA can be used not only for the analysis and deformation compensation of reference structures, but also for the analysis of more complex filigree structures with pronounced local deformation.}, language = {en} }