TY - CONF A1 - Birkenmaier, Clemens A1 - Krenkel, Lars T1 - Towards a realistic model of blood viscosity and coagulation in membrane oxygenators T2 - 6th European Conference on Computational Mechanics (Solids, Structures and Coupled Problems) - ECCM 6; 7th European Conference on Computational Fluid Dynamics - ECFD 7 : Glasgow, Scotland, UK, June 11-15, 2018 N2 - Modelling blood flow an shear induced coagulation in membraene oxygenators (MO) is challenging. The relevant geometry of oxygenator fibers (OF) and chaining threads is complex and spans several length scales. In relevant scales and regimes blood shows several significant non-Newtonian effects. Existing models are only capable of accounting for some, but not all relevant effects. Additionally, coagulation processes are influencing fluid properties and geometry significantly. Due to the enormous size of the discretised geometries highly detailed viscosity and coagulation properties of blodd flow in MOs. First step is to find a gemoetry dependent viscosity representation on basis of parametric micro channel experiments with anti-coagulated blood. Next step is a statistic coagulation model, based on micro channel experiments with human (re-calcified citrated) whole blood an evaluation of clinically used osygenators. Since shear rate dependent (i.e. viscosity dependet) coagulation in return influences the viscosity, a combined model with suitable implementation in a RANS framework is necessary. Towards this end, micro channel experiments with new and used single OFs triggering coagulation are performed. Structures of multimeric von Willebrand fibers (vWF), as indicator for shear induced coagulation, are compared to computed and measured flow conditions, using immunofluorescence microscopy, RANS-computations and µPIV, respectively. Preliminary examinations in clinically used MOs show good agreement between occurring structures of vWF, cell depositions and computed flow patterns (geometry form µCT-Scans). However, computed shear rates might be to low to actually trigger activation of vWF. The complex geometry of MOs results in huge meshes, which makes RANS with statistical modelling of viscosity and coagulation a reasonable approach. Towards this end, experimental data on micro channel level with evaluation on real application level is crucial. Especially regarding clotting processes, micro fluidic experiments are powerful research tool. KW - Blood Viscosity KW - Shear Raed Induced Coagulation KW - Membrane Oxygenator Y1 - 2023 UR - https://opus4.kobv.de/opus4-oth-regensburg/frontdoor/index/index/docId/5788 ER -