Background: The 12-lead ECG is spatially limited in diagnosing cardiac abnormalities. Additional leads (right sided and posterior leads) are inconvenient in a clinical setting, however, they can be derived. In this paper we report on the development of coefficients to allow derivation of right sided and posterior leads.
Method: Analysis was performed using body surface potential maps (BSPM) recorded from 910 patients in two centres. Recordings were made up of healthy controls (n=314), peak balloon inflation during elective percutaneous coronary angioplasty (n=88), myocardial infarction (n=271) and left-ventricular hypertrophy (n=237). All recordings were expanded to the 352-node Dalhousie torso. Coefficients to allow derivation of right sided and posterior leads were generated by linear regression. Further coefficients from a previously reported study were used for performance comparisons.
Results: Correlation coefficients between recorded and derived leads were significantly improved using the new coefficients (p<0.05) in leads V7-V12.
Conclusion: We have developed coefficients that allow the derivation of 10 additional leads from the 12-lead ECG.
Linear ECG-lead transformations estimate or derive unrecorded target leads by applying a number of recorded basis leads to a so-called linear ECG-lead transformation matrix. The inverse transform of such a linear ECG-lead transformation performs a transformation in the opposite direction (from the target leads to the basis leads). The pseudo-inverse of a given transformation matrix can be used to perform such an inverse transformation. Linear regression based inverse transformation matrices are, provided that sufficient
training data for their development is available, an alternative to pseudo-inverse matrices. The aim of this research was to compare the estimation performance of pseudo-inverse and linear regression based inverse transformations. This comparison was performed for two example inverse transformations. The performance of the different transformations was assessed using root-meansquared-error (RMSE) values between the QRS-T complexes of recorded and derived leads. Typical mean RMSE values associated with the regression based
approach were found to be approximately two thirds to half of the mean RMSE values achieved by the approach based upon the pseudo-inverse. Provided that sufficient data are available, linear regression should be used for the development of inverse ECG-lead transformation matrices.
This article investigates the selection of optimal ECG leads for the detection of ST changes more likely to appear in patch systems with closely spaced leads. Method: We analysed body surface potential maps (BSPMs) from 44 subjects undergoing PTCA. BSPMs were recorded at 120 sites and these were expanded to 352 nodes (Dalhousie torso) using Laplacian interpolation. A total of 88 BSPMs were investigated. This included the 44 subjects at baseline and the 44 subjects at peak balloon inflation (PBI). At PBI the subjects had various coronary arteries occluded (14 LAD, 15 LCX, 15 RCA). All possible bipolar leads were calculated for each subject. Leads were ranked based on the maximum ST-segment change between baseline and PBI for each subject. Leads with electrode spacing of more than 100 mm were excluded. The highest ranked lead was chosen as the short spaced lead (SSL) on the anterior torso. Result: The median ST-segment change for the chosen SSL for each vessel was LAD = 134 µV, LCX = 65 µV, RCA = 166 µV. The maximum ST segment change observed for the same lead was LAD = 277 µV, LCX = 166 µV, RCA = 257 µV . For comparison, the highest median observed on the 12-lead ECG for each vessel was LAD = 137 µV (V3), LCX = 130 µV (III), RCA = 196 µV (III).
The magnitude of the spatial ventricular gradient (MSVG) is an attractive parameter in electrocardiogram (ECG)monitoring applications. The MSVG is most commonlyobtained from150 Hz low-pass filtered resting ECGs. However, monitoring applications typically utilize 40 Hz low-pass filtered ECG data. The extend to which the value of the MSVG is affected by the utilization of 40 Hz low-pass monitoring ECG filters over the commonly used 150 Hz low-pass resting ECG filters has not previously been reported. The aim of this research was to quantify the differences between MSVG values computed using 40 Hz low-pass filtered ECG data (MSVG40) and 150 Hz low-pass filtered ECG data (MSVG150). The differences between the MSVG40 and the MSVG150 were quantified as systematic error (mean difference) and random error (span of Bland-Altman 95% limits of agreement) using a study population of 726 subjects. The systematic error was found to be 0.013 mV ms [95% confidence interval: 0.008 mV ms to 0.018 mV ms]. The random error was quantified as 0.282 mV ms [95% confidence interval: 0.266 mV ms to 0.298 mV ms]. Our findings suggest that it is possible to record accurate MSVG values using 40 Hz low-pass filtered ECG data.
Background: We have previously reported on the potential of patch-based ECG leads to observe changes typical during ischaemia. In this study we aim to assess the utility of patch-based leads in the detection of these changes.
Method: Body surface potential maps (BSPM) from subjects (n=45) undergoing elective percutaneous coronary angioplasty (PTCA) were used. The short spaced lead (SSL), that was previously identified as having the greatest ST-segment change between baseline and peak balloon inflation (PBI), was selected as the basis for a patch based lead system. A feature set of J-point amplitudes for all bipolar leads available within the same 100 mm region were included (n=6). Current 12-lead ECG criteria were applied to 12-lead ECGs for the same subjects to benchmark performance.
Results: The previously identified single SSL achieved sensitivity and specificity of 87% and 71% respectively using a Naive Bayes classifier. Adding other combinations of leads to this did not improve performance significantly. The 12-lead ECG performance was 62/93% (sensitivity/specificity).
Conclusion: This study suggests that short spaced leads can be sensitive to ischaemic ECG changes. However, due to the short distance between leads, they lack the specificity of the 12-lead ECG.
There are limited datasets available to facilitate the evaluation of patch-based lead systems, so the leads must be derived from existing data, mainly the 12-lead ECG. We have previously introduced a short spaced lead (SSL) system consisting of two leads with the largest ST segment changes during ischaemic-type episodes. In this study, we aim to evaluate the derivation of this patch-based lead system from the 12-lead ECG.
Method: Thoracic body surface potential maps (BSPM) were recorded from n=734 patients. Using Laplacian interpolation, each recording was expanded to the 352-node Dalhousie torso. The eight independent channels of the 12-lead ECG were extracted (I, II, V1-V6) with the two leads of the SSL patch Coefficients were derived using linear regression from the 12-lead ECG to the SSL patch.
Results: The median Pearson correlation coefficients (CC) and root mean square error (RMSE) for each lead were calculated as follows (CC/RMSE): 0.986/74.3 µV (ST monitoring lead); 0.976/65.3 µV (spatially orthogonal lead).
Conclusion: We have developed coefficients that allow the derivation of a patch-based lead system from the 12-lead ECG. Given the high correlation, it is possible to generate short spaced lead systems from existing diagnostic lead systems, however, amplitude errors are introduced in the process.
Linear electrocardiographic lead transformations (LELTs) are used to estimate unrecorded ECG leads by applying a number of recorded leads to a LELT matrix. Such matrices are commonly developed using a training dataset. The size of the training dataset has an influence on the estimation performance of a LELT matrix. However, an estimate of the minimal size required for the development of LELTs has previously not been reported.
The aim of this research was to determine such an estimate. We generated LELT matrices from differently sized (from n = 10 to n = 540 subjects in steps of 10 subjects) training datasets. The LELT matrices and the 12-lead ECG data of a testing dataset (n = 186 subjects) were used for the estimation of Frank VCGs. Root-mean-squared-error values between recorded and estimated Frank leads of the testing dataset were used for the quantification of the estimation performance associated with a given size of the training dataset.
The performance of the LELTs was, after an initial phase of improvement, found to only marginally improve with additional increases in the size of the training dataset. Our findings suggest that the training dataset should have
a minimal size of 170 subjects when developing LELTs that utilise the 12-lead ECG for the estimation of unrecorded
ECG leads.
Automated interpretation of the 12-lead ECG has remained an underpinning interest in decades of research that has seen a diversity of computing applications in cardiology.
The application of computers in cardiology began in the 1960s with early research focusing on the conversion of analogue ECG signals (voltages) to digital samples. Alongside this, software techniques that automated the extraction of wave measurements and provided basic diagnostic statements, began to emerge. In the years since then there have been many significant milestones which include the widespread commercialisation of 12-lead ECG interpretation software, associated clinical utility and the development of the related regulatory frameworks to promote standardised development.
In the past few years, the research community has seen a significant rejuvenation in the development of ECG interpretation programs. This is evident in the research literature where a large number of studies have emerged tackling a variety of automated ECG interpretation problems. This is largely due to two factors. Specifically, the technical advances, both software and hardware, that have facilitated the broad adoption of modern artificial intelligence (AI) techniques, and, the increasing availability of large datasets that support modern AI approaches.
In this article we provide a very high-level overview of the operation of and approach to the development of early 12-lead ECG interpretation programs and we contrast this to the approaches that are now seen in emerging AI approaches. Our overview is mainly focused on highlighting differences in how input data are handled prior to generation of the diagnostic statement.