Background: The 12-lead ECG is spatially limited in diagnosing cardiac abnormalities. Additional leads (right sided and posterior leads) are inconvenient in a clinical setting, however, they can be derived. In this paper we report on the development of coefficients to allow derivation of right sided and posterior leads.
Method: Analysis was performed using body surface potential maps (BSPM) recorded from 910 patients in two centres. Recordings were made up of healthy controls (n=314), peak balloon inflation during elective percutaneous coronary angioplasty (n=88), myocardial infarction (n=271) and left-ventricular hypertrophy (n=237). All recordings were expanded to the 352-node Dalhousie torso. Coefficients to allow derivation of right sided and posterior leads were generated by linear regression. Further coefficients from a previously reported study were used for performance comparisons.
Results: Correlation coefficients between recorded and derived leads were significantly improved using the new coefficients (p<0.05) in leads V7-V12.
Conclusion: We have developed coefficients that allow the derivation of 10 additional leads from the 12-lead ECG.
Linear ECG-lead transformations estimate or derive unrecorded target leads by applying a number of recorded basis leads to a so-called linear ECG-lead transformation matrix. The inverse transform of such a linear ECG-lead transformation performs a transformation in the opposite direction (from the target leads to the basis leads). The pseudo-inverse of a given transformation matrix can be used to perform such an inverse transformation. Linear regression based inverse transformation matrices are, provided that sufficient
training data for their development is available, an alternative to pseudo-inverse matrices. The aim of this research was to compare the estimation performance of pseudo-inverse and linear regression based inverse transformations. This comparison was performed for two example inverse transformations. The performance of the different transformations was assessed using root-meansquared-error (RMSE) values between the QRS-T complexes of recorded and derived leads. Typical mean RMSE values associated with the regression based
approach were found to be approximately two thirds to half of the mean RMSE values achieved by the approach based upon the pseudo-inverse. Provided that sufficient data are available, linear regression should be used for the development of inverse ECG-lead transformation matrices.
This article investigates the selection of optimal ECG leads for the detection of ST changes more likely to appear in patch systems with closely spaced leads. Method: We analysed body surface potential maps (BSPMs) from 44 subjects undergoing PTCA. BSPMs were recorded at 120 sites and these were expanded to 352 nodes (Dalhousie torso) using Laplacian interpolation. A total of 88 BSPMs were investigated. This included the 44 subjects at baseline and the 44 subjects at peak balloon inflation (PBI). At PBI the subjects had various coronary arteries occluded (14 LAD, 15 LCX, 15 RCA). All possible bipolar leads were calculated for each subject. Leads were ranked based on the maximum ST-segment change between baseline and PBI for each subject. Leads with electrode spacing of more than 100 mm were excluded. The highest ranked lead was chosen as the short spaced lead (SSL) on the anterior torso. Result: The median ST-segment change for the chosen SSL for each vessel was LAD = 134 µV, LCX = 65 µV, RCA = 166 µV. The maximum ST segment change observed for the same lead was LAD = 277 µV, LCX = 166 µV, RCA = 257 µV . For comparison, the highest median observed on the 12-lead ECG for each vessel was LAD = 137 µV (V3), LCX = 130 µV (III), RCA = 196 µV (III).
The magnitude of the spatial ventricular gradient (MSVG) is an attractive parameter in electrocardiogram (ECG)monitoring applications. The MSVG is most commonlyobtained from150 Hz low-pass filtered resting ECGs. However, monitoring applications typically utilize 40 Hz low-pass filtered ECG data. The extend to which the value of the MSVG is affected by the utilization of 40 Hz low-pass monitoring ECG filters over the commonly used 150 Hz low-pass resting ECG filters has not previously been reported. The aim of this research was to quantify the differences between MSVG values computed using 40 Hz low-pass filtered ECG data (MSVG40) and 150 Hz low-pass filtered ECG data (MSVG150). The differences between the MSVG40 and the MSVG150 were quantified as systematic error (mean difference) and random error (span of Bland-Altman 95% limits of agreement) using a study population of 726 subjects. The systematic error was found to be 0.013 mV ms [95% confidence interval: 0.008 mV ms to 0.018 mV ms]. The random error was quantified as 0.282 mV ms [95% confidence interval: 0.266 mV ms to 0.298 mV ms]. Our findings suggest that it is possible to record accurate MSVG values using 40 Hz low-pass filtered ECG data.
Background: We have previously reported on the potential of patch-based ECG leads to observe changes typical during ischaemia. In this study we aim to assess the utility of patch-based leads in the detection of these changes.
Method: Body surface potential maps (BSPM) from subjects (n=45) undergoing elective percutaneous coronary angioplasty (PTCA) were used. The short spaced lead (SSL), that was previously identified as having the greatest ST-segment change between baseline and peak balloon inflation (PBI), was selected as the basis for a patch based lead system. A feature set of J-point amplitudes for all bipolar leads available within the same 100 mm region were included (n=6). Current 12-lead ECG criteria were applied to 12-lead ECGs for the same subjects to benchmark performance.
Results: The previously identified single SSL achieved sensitivity and specificity of 87% and 71% respectively using a Naive Bayes classifier. Adding other combinations of leads to this did not improve performance significantly. The 12-lead ECG performance was 62/93% (sensitivity/specificity).
Conclusion: This study suggests that short spaced leads can be sensitive to ischaemic ECG changes. However, due to the short distance between leads, they lack the specificity of the 12-lead ECG.
This study assessed the performance of a deep neural network (PulseAI, Belfast, United Kingdom) used in conjunction with a dry-electrode ECG sensor device (RhythmPad, D&FT, United Kingdom) to detect AF automatically. Simultaneous pairs of 12-lead ECGs and single-lead dry-electrode ECGs were collected from 622 patients. The 12-lead ECGs were manually overread and used as reference diagnoses. Twenty-two patients were confirmed with AF and had an interpretable 12-lead and single-lead dry-electrode ECG recording. The deep neural network analysed the dry-electrode ECGs, and performance was compared to the 12-lead interpretation. Overall, the deep neural network algorithm yielded a sensitivity of 96% (95% CI, 87%-100%), specificity of 99% (95% CI, 98%-100%) and positive predictive value of 81% (95% CI, 66%-96%) for detection of AF episodes. When coupled with dry-electrode ECG sensors, the PulseAI neural network allows for large-scale and low-cost screening for AF. Widespread implementation of this technology may allow for earlier detection, treatment, and management of patients with AF.
Linear ECG-lead transformations (LELTs) are used to estimate unrecorded leads by applying a number of recorded leads to a LELT matrix. Such LELT matrices are commonly developed using a training dataset and linear regression analysis. An important performance metric of LELTs is the subject-to-subject variability (SSV) of their estimation performance. In this research, we assess the relationship between an increasing training set size (from n=10 to n=370 subjects) and the SSV of LELTs. A total of 200 LELT matrices were developed for each training sets size. The developed LELT matrices and 12-lead ECG data of a testing dataset (n=123 subjects) were used for the estimation of Frank VCGs. Root-mean-squared-error (RMSE) values between recorded and estimated Frank VCG leads were used for the quantification of the estimation performance. The SSV associated with each LELT matrix was quantified as the standard deviation of the corresponding RMSE values. This was followed by an analysis of the relationship between the training set size and the associated SSV values. Increasing the training set size from 10 to 180, to 160 and to 200 subjects, for Frank VCG leads X,Y and Z respectively, was associated with a reduction of the observed SSV. Further increases in training set size were found to only have a marginal effect on the observed SSV.
Background
The application of artificial intelligence to interpret the electrocardiogram (ECG) has predominantly included the use of knowledge engineered rule-based algorithms which have become widely used today in clinical practice. However, over recent decades, there has been a steady increase in the number of research studies that are using machine learning (ML) to read or interrogate ECG data.
Objective
The aim of this study is to review the use of ML with ECG data using a time series approach.
Methods
Papers that address the subject of ML and the ECG were identified by systematically searching databases that archive papers from January 1995 to October 2019. Time series analysis was used to study the changing popularity of the different types of ML algorithms that have been used with ECG data over the past two decades. Finally, a meta-analysis of how various ML techniques performed for various diagnostic classifications was also undertaken.
Results
A total of 757 papers was identified. Based on results, the use of ML with ECG data started to increase sharply (p < 0.001) from 2012. Healthcare applications, especially in heart abnormality classification, were the most common application of ML when using ECG data (p < 0.001). However, many new emerging applications include using ML and the ECG for biometrics and driver drowsiness. The support vector machine was the technique of choice for a decade. However, since 2018, deep learning has been trending upwards and is likely to be the leading technique in the coming few years. Despite the accuracy paradox, accuracy was the most frequently used metric in the studies reviewed, followed by sensitivity, specificity, F1 score and then AUC.
Conclusion
Applying ML using ECG data has shown promise. Data scientists and physicians should collaborate to ensure that clinical knowledge is being applied appropriately and is informing the design of ML algorithms. Data scientists also need to consider knowledge guided feature engineering and the explicability of the ML algorithm as well as being transparent in the algorithm's performance to appropriately calibrate human-AI trust. Future work is required to enhance ML performance in ECG classification.
Deep Convolutional Neural Networks (DCNNs) have been shown to provide improved performance over traditional heuristic algorithms for the detection of arrhythmias from ambulatory ECG recordings. However, these DCNNs have primarily been trained and tested on device-specific databases with standardized electrode positions and uniform sampling frequencies. This work explores the possibility of training a DCNN for Atrial Fibrillation (AF) detection on a database of single‑lead ECG rhythm strips extracted from resting 12‑lead ECGs. We then test the performance of the DCNN on recordings from ambulatory ECG devices with different recording leads and sampling frequencies.
We developed an extensive proprietary resting 12‑lead ECG dataset of 549,211 patients. This dataset was randomly split into a training set of 494,289 patients and a testing set of the remaining 54,922 patients. We trained a 34-layer convolutional DCNN to detect AF and other arrhythmias on this dataset. The DCNN was then validated on two Physionet databases commonly used to benchmark automated ECG algorithms (1) MIT-BIH Arrhythmia Database and (2) MIT-BIH Atrial Fibrillation Database. Validation was performed following the EC57 guidelines, with performance assessed by gross episode and duration sensitivity and positive predictive value (PPV). Finally, validation was also performed on a selection of rhythm strips from an ambulatory ECG patch that a committee of board-certified cardiologists annotated.
On MIT-BIH, The DCNN achieved a sensitivity of 100% and 84% PPV in detecting episodes of AF. and 100% sensitivity and 94% PPV in quantifying AF episode duration. On AFDB, The DCNN achieved a sensitivity of 94% and PPV of 98% in detecting episodes of AF, and 98% sensitivity and 100% PPV in quantifying AF episode duration. On the patch database, the DCNN demonstrated performance that was closely comparable to that of a cardiologist.
The results indicate that DCNN models can learn features that generalize between resting 12‑lead and ambulatory ECG recordings, allowing DCNNs to be device agnostic for detecting arrhythmias from single‑lead ECG recordings and enabling a range of clinical applications.
Linear electrocardiographic lead transformations (LELTs) are used to estimate unrecorded ECG leads by applying a number of recorded leads to a LELT matrix. Such matrices are commonly developed using a training dataset. The size of the training dataset has an influence on the estimation performance of a LELT matrix. However, an estimate of the minimal size required for the development of LELTs has previously not been reported.
The aim of this research was to determine such an estimate. We generated LELT matrices from differently sized (from n = 10 to n = 540 subjects in steps of 10 subjects) training datasets. The LELT matrices and the 12-lead ECG data of a testing dataset (n = 186 subjects) were used for the estimation of Frank VCGs. Root-mean-squared-error values between recorded and estimated Frank leads of the testing dataset were used for the quantification of the estimation performance associated with a given size of the training dataset.
The performance of the LELTs was, after an initial phase of improvement, found to only marginally improve with additional increases in the size of the training dataset. Our findings suggest that the training dataset should have
a minimal size of 170 subjects when developing LELTs that utilise the 12-lead ECG for the estimation of unrecorded
ECG leads.