Intercomparison of Radiation Temperature Scales between PTB and NMIJ from 1000-degree Celsius to 2500-degree Celsius
- The intercomparison of radiation temperature scales between PTB and NMIJ was carried out in September 2001. PTB sent their radiation thermometer LP3 to Japan and NMIJ made the comparison between the LP3 and NMIJ’s Topcon radiation thermometers with a comparison blackbody from 1000 C to 2500 C. Both scales agreed within±0.8 C in the whole range.
Author: | Jürgen Hartmann, F. Sakuma, L. Ma |
---|---|
Persistent identifier: | https://doi.org/10.11499/sicep.2002.0.13.0 |
Parent Title (English): | SICE Annual Conference Program and Abstracts SICE Annual Conference 2002 |
Publisher: | The Society of Instrument and Control Engineers |
Document Type: | Conference Proceeding |
Language: | English |
Year of publication: | 2002 |
Publishing Institution: | Hochschule für Angewandte Wissenschaften Würzburg-Schweinfurt |
Release Date: | 2020/12/04 |
First Page: | 13 |
Last Page: | 13 |
Dewey Decimal Classification: | 5 Naturwissenschaften und Mathematik / 53 Physik / 536 Wärme |