Intercomparison of Radiation Temperature Scales

  • The intercomparison of radiation temperature scales between PTB and NMIJ was carried out in September 2001. PTB sent their radiation thermometer LP3 to Japan and NMIJ made the comparison between the LP3 and NMIJ's Topcon radiation thermometers with a comparison blackbody from 1000 C to 2500 C. Both scales agreed within+ 0.8 C in the whole range.

Export metadata

Additional Services

Search Google Scholar
Metadaten
Author:Jürgen Hartmann
Persistent identifier:https://www.researchgate.net/profile/Juergen_Hartmann3/publication/4012565_Intercomparison_of_radiation_temperature_scales_between_PTB_and_NMIJ_from_1000C_to_2500C/links/558940cd08ae9076016e9870/Intercomparison-of-radiation-temperature-scales-between-PTB-and-NMIJ-from-1000C-to-2500C.pdf