TY - CPAPER U1 - Konferenzveröffentlichung A1 - Hartmann, Jürgen T1 - A comparison of Co-C, Pd-C, Pt-C, Ru-C and Re-C eutectic fixed points independently manufactured by three different institutes T2 - Proc. Newrad (Davos, Switzerland) N2 - In order to assess the performance and capability as well as the reproducibility of cell production five different Metal-carbon (MC) eutectic cells from three institutes were measured by radiation thermometry at PTB. The reproducibility of the melting temperature approached the requirements set by CCT/CCPR but significant differences were seen between different manufactures. AB - In order to assess the performance and capability as well as the reproducibility of cell production five different Metal-carbon (MC) eutectic cells from three institutes were measured by radiation thermometry at PTB. The reproducibility of the melting temperature approached the requirements set by CCT/CCPR but significant differences were seen between different manufactures. Y1 - 2005 UR - https://www.academia.edu/download/41206514/A_comparison_of_CoC_PdC_PtC_RuC_and_ReC_20160114-3055-1lybi4p.pdf20160115-19908-1rtghdd.pdf SP - 289 EP - 290 ER -