@inproceedings{Hartmann2004, author = {Hartmann, J{\"u}rgen}, title = {Long term stability of the spectral responsivity of filter radiometers at the PTB}, booktitle = {Proc. of TEMPMEKO}, editor = {Zvizdic, D.}, pages = {977 -- 982}, year = {2004}, abstract = {The Physikalisch-Technische Bundesanstalt (PTB) uses calibrated filter radiometers to measure the thermodynamic temperature of blackbodies applied for the realisation and dissemination of the spectral radiance and spectral irradiance. This paper presents the PTB experience with the stability of the spectral responsivities of two different sets of filter radiometers, a set of four broad-band filter radiometers with centroid wavelengths in the visible spectral range (437 nm, 510 nm, 558 nm and 728 nm), using a combination of coloured glass filters as wavelength selective elements and a set of five narrow-band interference filter radiometers with centroid wavelengths in the visible and in the near infrared wavelength range (676 nm, 802 nm, 903 nm and 1003 nm). The time period of their continuous investigation with respect to the relevant stability parameters, the integrated spectral responsivity and the centroid wavelength, is about seven years for the broad-band filter radiometers and up to ten years for the narrow-band interference filter radiometers. Both types of filter radiometers show sufficient stability for their application as high-accuracy temperature measurement instruments. Relative changes in the integrated spectral responsivity up to some parts in 103 and for the interference filter radiometers a linear increase of the centroid wavelength with time in the order of several ten picometers per year have been observed. A comparison between the two types of filter radiometers is given, with respect to the contribution of the stability of these parameters to the uncertainty budget of thermodynamic temperature determination. Reasons for the observed …}, language = {en} }