Ingenieurwissenschaften und zugeordnete Tätigkeiten
Filtern
Dokumenttyp
- Vortrag (8)
- Zeitschriftenartikel (4)
- Beitrag zu einem Tagungsband (1)
- Posterpräsentation (1)
Schlagworte
- IIT (4)
- ISO 14577 (4)
- Instrumentierte Eindringprüfung (4)
- Nanoindentation (4)
- Haftfestigkeit (3)
- Instrumented indentation testing (3)
- Mechanical properties (3)
- Normung (3)
- Dynamic testing (2)
- Instrumented Indentation Testing (2)
Organisationseinheit der BAM
- 6 Materialchemie (7)
- 6.7 Materialsynthese und Design (6)
- 5 Werkstofftechnik (1)
- 5.1 Mikrostruktur Design und Degradation (1)
- 6.1 Oberflächen- und Dünnschichtanalyse (1)
- 6.2 Material- und Oberflächentechnologien (1)
- 6.6 Physik und chemische Analytik der Polymere (1)
- 9 Komponentensicherheit (1)
- 9.5 Tribologie und Verschleißschutz (1)
Eingeladener Vortrag
- nein (8)
The accuracy and comparability of nanoindentation results depend significantly on the calibration of area function and instrument compliance. The area function results should not depend on the reference material used or on the calibration method (direct or indirect). This has been investigated for 18 different Berkovich tips. A novel calibration method is proposed that confirms the material independence of the area function and gives a force-dependent instrument compliance function. An agreement between direct and indirect calibration could only be achieved by considering a radial displacement correction. Further, it is shown that the transition range from a spherical cap to the correct face angle of the pyramid can extend to a depth of more than 250 nm. A better parameter for the indenter than the tip radius is the offset of the contact radius to the radius of an ideal tip at a depth where the correct face angle is reached.
The authors describe and interpret curvature-related changes to the ellipsometric readout. As model system for a concave curvature, a set of three spherical microscopic indents in silicon (100) of different sizes was prepared by instrumented indentation testing using a spherical indenter. For reference, these samples were characterized by AFM to reveal the topography of each structure. The concavelike indents were analyzed by Mueller-Matrix imaging ellipsometry to extract lateral intensity images of 12 elements of the Mueller-Matrix. As a result of the detailed analysis of the image elements m22, m23, and m14, it was possible to correlate intensity changes and symmetry properties to depolarization and cross polarization induced by the edge threshold and the curved surface of the indent.
In der DIN EN ISO 14577 Metallische Werkstoffe – Instrumentierte Eindringprüfung zur Bestimmung der Härte und anderer Werkstoffparameter sind die Durchführung und Auswertung des Verfahrens, sowie die Kalibrierung der zu verwendenden Prüfmaschine genormt. In dieser Norm werden auch grundsätzliche Aussagen zur Bestimmung der Messunsicherheit gemacht.
An Beispielen aus der täglichen Prüfpraxis werden verschiedene Möglichkeiten, aber auch die Grenzen der Bestimmung der Messunsicherheitbei der Instrumentierten Eindringprüfung beschrieben und kritisch diskutiert.
Some critical discussion of the state of the art of uncertainty evaluationin Instrumented Indentation Testing IIT
Nowadays the Instrumented Indentation Testing (IIT), in the nano range often named as nano indention, is one of the most commonly used methods to determine the mechanical properties of materials in the micro and nano range. This method is already extensive standardized in ISO 14577 part 1-4. In the past, the application of this standard in testing praxis shows that the established values have an excellent precision.If an uncertainty is calculated, the range of values within which the true value is asserted to lie with some level of confidencewill be known. In part 1 of ISO 14577 [1] two methods for evaluation of the uncertainty in IIT are mentioned:
Method 1 for determining uncertainty considers only those uncertainties associated with the overall measurement performance of the testing machine with respect to the reference blocks.
Method 2calculates a combined uncertainty from individual contributions. These may be grouped into random and systematic uncertainties.
Both methods will be described in detail using examples from the dailyexperimental praxis. The comparabilityof both methods will be critically discussed. Finally, it will be showedhow the calculated uncertaintiescan be used for performancetests and product specifications.
Acknowledgement
This work was performed under the support of the EMPIR project 17NRM05Advancing measurement uncertainty̶ comprehensiveexamples for key international standards
References
[1] ISO 14577 part 1 (2017)
Nowadays the Instrumented Indentation Testing (IIT) is one of the most commonly used methods to determine the mechanical properties of materials in the nano range. This method is already extensive standardized in EN ISO 14577 part 1-4. Because of the great interest of researchers and industries in investigations of time depending material behavior mostly all suppliers of IIT equipment are offering the possibility of dynamic testing. Realizing this development ISO/TC 164/SC3 Hardness Testing has proposed to start the new standardization project “Linear elastic dynamic instrumented indentation testing DIIT”. The development of this standard is accompanied by the first international intercomparing exercise comparing results of dynamic instrumented indentation testing from testing machines using different hardware solutions and different models for data evaluation.
The draft of part 5 of ISO 14577 “Linear elastic dynamic instrumented indentation testing DIIT” specifies verification and calibration of testing machines for carrying out the measurement of the dynamic material response when an oscillatory force or displacement, with amplitudes small in comparison to the prescribed target values, is imparted to the indenter while the indenter is continuously loaded to a prescribed target load or target depth or while the load or displacement is held constant at a prescribed target value. In case of a material showing plastic-elastic behavior, the measured dynamic response is used for continuous evaluation of the dynamic stiffness of the contact as a function of depth and frequency. Using the dynamic stiffness of the contact a reduced dynamic modulus will be calculated. In case of a material showing visco-elastic behavior from the measured dynamic response also the dynamic contact damping coefficient as function of depth and frequency is evaluated continuously. Using dynamic contact stiffness and dynamic contact damping coefficient reduced lost and storage modulus for visco-elastic materials will be calculated.
The main normative requirements of the draft will be presented and discussed in the light of the first results of the intercomparing excise.
Commercial grade-1 titanium samples (Ti, 99.6%) were treated using three alternative methods, (i) femtosecond laser processing, (ii) thermal heat treatment, and (iii) electrochemical anodization, respectively, resulting in the formation of differently conditioned superficial titanium oxide layers. The laser processing (i) was carried out by a Ti:sapphire laser (pulse duration 30 fs, central wavelength 790 nm, pulse repetition rate 1 kHz) in a regime of generating laser-induced periodic surface structures (LIPSS). The experimental conditions (laser fluence, spatial spot overlap) were optimized in a sample-scanning setup for the processing of several square-millimeters large surface areas covered homogeneously by these nanostructures. The differently oxidized titanium surfaces were characterized by optical microscopy, micro Raman spectroscopy, variable angle spectroscopic ellipsometry, and instrumented indentation testing. The tribological performance was characterized in the regime of mixed friction by reciprocating sliding tests against a sphere of hardened steel in fully formulated engine oil as lubricant. The specific tribological performance of the differently treated surfaces is discussed with respect to possible physical and chemical mechanisms.
The draft of part 5 of ISO 14577 specifies verification and calibration of testing machines for carrying out the measurement of the dynamic material response when an oscillatory force or displacement, with amplitudes small in comparison to the prescribed target values, is imparted to the indenter while the indenter is continuously loaded to a prescribed target load or target depth or while the load or displacement is held constant at a prescribed target value. In case of a material showing plastic-elastic behavior, the measured dynamic response is used for continuous evaluation of the dynamic stiffness of the contact as a function of depth and frequency. Using the dynamic stiffness of the contact a reduced dynamic modulus will be calculated. In case of a material showing visco-elastic behavior from the measured dynamic response also the dynamic contact damping coefficient as function of depth and frequency is evaluated continuously. Using dynamic contact stiffness and dynamic contact damping coefficient reduced lost and storage modulus for visco-elastic materials will be calculated. The main normative requirements of the draft are discussed and their applications are illustrated by examples from daily experimental praxis.
Nowadays the Instrumented Indentation Testing (IIT), in the nano range often named as nano indention, is one of the most commonly used methods to determine the mechanical properties of materials in the micro and nano range. This method is already extensive standardized in DIN EN ISO 14577 part 1-4. In the past, the application of this standard in testing praxis shows that the established values have an excellent precision. On the other side, the trueness as comparability of obtained results with reference values is not so good. To improve accuracy of IIT the use of a variable ε and the consideration of the lateral displacement during indentation became normative requirements during the last revision of ISO 14577.
Starting with the beginning of instrumented indentation testing in 1898 the historical background and the main assumption of the Oliver and Pharr method will be explained and critical discussed. It will be shown how to apply the use of a variable ε and the consideration of the lateral displacement during indentation in the daily testing work to improve accuracy of IIT.
Finally, ongoing standardization projects in the field of IIT will be briefly presented.
Polycrystalline Ni materials with grain sizes less than 100 nm (nano crystalline NC) and with grain sizes in the micrometer range (micro crystalline MC) in form of thin films have become important in many technologies due to their improved physical, chemical and mechanical properties. Usual the mechanical properties of such coatings are described by a Hardness value and a Young´s modulus measured by Instrumented Indentation Testing (IIT).The behavior of such coatings during indentation test is influenced by different size effects having their representative length scale – grain size, coating thickness, length that characterizes the depth dependence of the hardness (Indentation Size Effect ISE). To estimate realistic values for the intrinsic coating Hardness and Young´s modulus all size effects have to be considered.
For this work thin nano crystalline Ni – coatings (average grain size 30 nm) with thickness from 1 μm to 5 μm were electrodeposited on brass substrates. Indentation tests in the continuous stiffness measurement (CSM) mode were provided on as prepared Ni – coatings using a G200 Nanoindentation system (Fa. Keysight). For estimation of the intrinsic hardness of the coatings from composite hardness values calculated from the measured force –displacement curve using the Oliver & Pharr method, the model described by Z.S. Ma [1] was used.
It was found that the experimental data can be well described by the model. The fitted values for the intrinsic hardness and the fitting parameters will be given. The different values of intrinsic hardness and of length characterizing depth dependence of the hardness for different coating thickness are discussed as results of changes in the coating structure because of changes in deposition parameters.
[1] Z.S. Ma, Y.C. Zhou, S.G. Long, C. Lu: On the intrinsic hardness of a metallic film/substrate system: Indentation size and substrate effects. International Journal of Plasticity 34 (2012) 1-11.
This work is supported by European Metrology Program for Innovation and Research (EMPIR)
(JRP 14IND03 Strength – ABLE)
Durch faseroptische Sensoren können Bauteile im Einsatz kontinuierlich überwacht und frühzeitig Informationen über Materialveränderungen gewonnen werden. Mittels einer magnetostriktiven Aktorschicht, welche den Sensor umhüllt, kann nun auch die korrekte Sensorfunktion jederzeit kontrolliert werden. Als Aktorschicht eignet sich galvanisch abgeschiedenes reines Nickel oder Nickel-Eisen in der Zusammensetzung 50:50. Um diese haftfest mit dem Sensor zu verbinden wurde ein ECD/PVD-Kombinationsschichtsystem entwickelt. Die mechanischen Eigenschaften der ECD-Schicht können sowohl mit einer auf die Fasergeometrie angepassten instrumentierten Eindringprüfung als auch einem 2-Punkt-Biegeversuch bestimmt werden.
Mechanische Eigenschaften von Materialien im Mikro- und Nanobereich werden heute mit der Instrumentierten Eindringprüfung bestimmt. Das Verfahren ist bereits umfangreich in der DIN EN ISO 14577 Teil 1-4 genormt. Im Ergebnis des EU EMRP Projekts Dynamic mechanical properties and long-term deformation behavior of viscous materials (MEPROVISC)“ wurden die zwei neuen Normprojekte für die Instrumentierte Eindringprüfung „Linear elastische dynamische Eindringprüfung“ und „Eindringkriechen und zeitabhängige Eigenschaften“ vorgeschlagen und mit deren Bearbeitung im ISO/TC 164/SC3 Hardness Testing begonnen.
Ausgehend von der Darstellung der wesentlichen Ergebnisse des Projekts MEPROVISC werden die Grundzüge der neuen Normen in Bezug auf Kalibrierung, Methodik, Auswertung und Bestimmung der Messunsicherheit erläutert. Weiterhin werden die Möglichkeiten der aktiven Mitarbeit an diesen Normprojekten aufgezeigt.
Der Vortrag widmet sich dem „RITZ-TEST: EVALUIERUNG UND STAND DER NORMUNG“ und beschreibt die Komponenten, Ergebnis und Messgrößen, nicht registrierende Prüfgeräte, Evaluierung Ritz-Test (Makrobereich), den Atlas der Versagensformen, Geräte für den instrumentierten Ritz-Test und die Normung des Ritz-Tests (ASTM) zur Bewertung der Haftfestigkeit“ im Einzelnen.
Der Vortrag widmet sich der Kratzbeständigkeit-Prüfmethoden und deren Bewertung und beschreibt Begriffe/Definitionen, Wirtschaftliche Bedeutung der Kratzbeständigkeit, Mess- und Prüfverfahren – allgemein (Anwendbarkeit, Messunsicherheit, Validierung), Mess- und Prüfverfahren – speziell die Erzeugung, Prüfverfahren von Einzel- und Vielfach-Kratzern, die Charakterisierung von Verkratzungen und eine zusammenfassende Bewertung im Einzelnen