Analytische Chemie
Filtern
Erscheinungsjahr
- 2022 (28) (entfernen)
Dokumenttyp
- Vortrag (14)
- Zeitschriftenartikel (12)
- Buchkapitel (1)
- Posterpräsentation (1)
Schlagworte
- Laser ablation (8)
- Surface coating (7)
- Plasma chemistry (6)
- Emission spectroscopy (5)
- Laser induced plasma deposition (5)
- Hydrodynamic model (4)
- Laser induced plasma (4)
- Plasma modeling (3)
- Raman microspectroscopy (3)
- Titanium dioxide (3)
Organisationseinheit der BAM
- 1.3 Instrumentelle Analytik (28) (entfernen)
Probe diagnostics is used to determine the electron temperature and electron number density in a low pressure inductively coupled plasma (ICP) ignited in the mixture of SiF4, Ar and H2. Emission spectra of mixtures with different stoichiometry of components are investigated and the electron density distribution function (EDDF) is estimated. The optimal conditions for high conversion of SiF4 into Si are found by studying the dependence of the yield of silicon upon the ratio of reagents. The maximum achieved yield of silicon is 85% under the optimal conditions. Based on the analysis of IR and MS spectra of exhaust gases, 5% of initial SiF4 converts into volatile fluorosilanes. A rate of production of Si is 0.9 g/h at the energy consumption 0.56 kWh /g.
Laser-induced plasmas are widely used in many areas of science and technology; examples include spectrochemical analysis, thin film deposition, and material processing. Several topics will be addressed. First, general phenomenology of laser-induced plasmas will be discussed. Then, a chemical model will be presented based on a coupled solution of Navier-Stokes, state, radiative transfer, material transport, and chemical equations. Results of computer simulations for several chemical systems will be shown and compared to experimental observations obtained by optical imaging, spectroscopy, and tomography. The latter diagnostic tools will also be briefly discussed. Finally, a prospective application of laser-induced plasma and plasma modeling will be illustrated on the example of chemical vapor deposition of molybdenum borides and micro processing and coating of titanium dental implants.
Laser metal deposition is a rapidly evolving method for additive manufacturing that combines high performance and simplified production routine. Quality of production depends on an instrumental design and operational parameters, which require constant control during the process. In this work, feasibility of using optical spectroscopy as a control method is studied via modeling and experimentally. A simplified thermal model is developed based on the time-dependent diffusion-conduction heat equation and geometrical light collection into detection optics. Intense light emitted by a laser-heated spot moving across a sample surface is collected and processed to yield the temperature and other temperature-related parameters. In the presence of surface defects, the temperature field is distorted in a specific manner that depends on shape and size of the defect. Optical signals produced by such the distorted temperature fields are simulated and verified experimentally using a 3D metal printer and a sample with artificially carved defects. Three quantities are tested as possible metrics for monitoring the process: temperature, integral intensity, and correlation coefficient. The shapes of the simulated signals qualitatively agree with the experimental signals; this allows for a cautious inference that optical spectroscopy can detect surface defects and, possibly, predict their characters, e.g., inner or protruding.
Equilibrium model of titanium laser induced plasma in air with reverse deposition of titanium oxides
(2022)
A chemical-hydrodynamic model of laser induced plasma is developed to study a process of deposition of titanium oxides from titanium laser induced plasma to the titanium target surface. The model is relevant to texturing and coating of titanium bone implants that is done by scanning the ablation laser across implant surfaces. Such the procedure improves the biocompatibility and durability of the implants. The model considers plasma chemical reactions, formation of condensed species inside the plasma plume, and deposition and accumulation of these species on the ablation surface. A chemical part of the model is based on minimization of Gibbs free energy of the chemical system; it is used to calculate the chemical composition of the plasma. A hydrodynamic part uses the 2D fluid-dynamic equations that model a 3D axisymmetric plasma plume and assumes the mass and energy exchange between the plasma and the surface. The initial parameters for the model are inferred from experiment.
The model shows that condensed titanium oxides, mostly TiO2, form in a peripheral plasma zone and gradually adhere to the surface during the plasma plume evolution. The model predicts the major component and thickness of the deposit and can be applied for the optimization of experiments aimed at surface modification.
In the present talk the basics of the Raman spectroscopy and particularly of Raman microscopy are explained. Advantages and disadvantages of the method are highlighted through selected case studies. In the second part of the lecture examples of correlative imaging with electron, X-ray, ion and optical microscopies from micro- to the nanoscale are highlighted.
Back Deposition of Titanium Oxides under Laser Ablation of Titanium: Simulation and Experiment
(2022)
Titanium is widely used in medicine for implants and prostheses, thanks to its high biocompatibility, good mechanical properties, and high corrosion resistance. Pure titanium, however, has low wear resistance and may release metallic titanium into surrounding tissues. Structuring and coating its surface with oxide layers are necessary for high wear resistance and improved biocompatibility. In this work, a combination of theoretical and experimental methods was used to study processes responsible for deposition of titanium oxides during ablation of titanium in air.
The deposition process was modeled via the Navier-Stokes equations that accounted for the material removal and accumulation of the deposit on the ablation surface. The chemical part was based on the equilibrium model embedded into the hydrodynamic code. Simulations showed that the most active zone of production of condensed titanium oxides were at plasma periphery whereas a zone of strong condensation of titanium metal was above the molten pool.
In experiment, a pulsed Yb fiber laser was scanned across a titanium surface. The temperature and composition of the plasma were inferred from plasma emission spectra. The post-ablation surface was analyzed by SEM, TEM, STEM, AFM, and XRD.
The developed model well reproduced the main features of experimental data. It was concluded that the deposition of condensed metal oxides from the plasma is a principal mechanism of formation of nanoporous oxide layer on the metal surface. The method of surface structuring and modification by nanosecond laser ablation can be developed into a useful technology that may find applications in medicine, photonics, and other areas.
The deposition of titanium oxides during titanium laser ablation in air has been experimentally and numerically investigated. A titanium sample was irradiated by nanosecond pulses from an Yb-fber laser with a beam scanned across the sample surface for its texturing. As a result, the hierarchical structure was observed consisting of a microrelief formed by the laser ablation and a nanoporous coating formed by the reverse deposition from the laser induced plasma plume. The chemical and phase composition of the nanoporous coating, as well as the morphology and structure of the surface, were studied using scanning electron microscopy, atomic force microscopy, and X-ray microanalysis. It was found that the deposit consists mostly of porous TiO2 with 26% porosity and inclusions of TiO, Ti2O3, and Ti2O3N. Optical emission spectroscopy was used to control the plasma composition and estimate the effective temperature of plasma plume. The chemical-hydrodynamic model of laser induced plasma was developed to get a deeper insight into the deposition process. The model predicts that condensed titanium oxides, formed in peripheral plasma
zones, gradually accumulate on the surface during the plasma plume evolution. A satisfactory agreement between the experimental and calculated chemical composition of the plasma plume as well as between the experimental and calculated composition and thickness of the deposited film was demonstrated. This allows a cautious conclusion that the formation of condensed oxides in the plasma and their consequent deposition onto the ablation surface are among the key mechanisms of formation of porous surface films.
Living organisms constantly interact with their environment, including through the exchange of gases. Respiration and the release and uptake of volatile organic compounds (VOCs) create dynamic microenvironments in biological systems. Studying the kinetics of volatiles in biological systems requires expensive instruments, and data analysis is challenging. Therefore, we aimed to design a minimal analytical device for measuring the composition of gaseous mixtures in realtime.
We built the ‘Modular Biological Mass Spectrometer’ (MoBiMS) from 3D-printed parts and custom sensors to fit a wide array of experimental set-ups. We tested the chemical detection range and temporal resolution of the MoBiMS employing pure compounds and complex biological samples.
Compounds with a higher than 0.4 mmHg vapor pressure and a molecular weight up to 154 g/mol were reliably sensed within seconds. The generated electron impact (EI) spectra were directly comparable with standard databases like the NIST EI library. Under a direct analysis approach, the MoBiMS identified the characteristic odor of banana (Musa sp.), that is, isoamyl acetate; tracked the dynamics of CO2 release while the Alka-Seltzer® reaction occurred showed the kinetics of the transient production and consumption of carbon dioxide during photosynthesis. MoBiMS also discriminated between volatile compounds ions coming from tobacco (Nicotiana benthamiana) leaves and the surrounding air through untargeted analysis. The small footprint of the MoBiMS and its relatively low energy consumption facilitate in situ analyses, as an additional gas supply is not necessary with EI ionization. The MoBiMS is easy to assemble, and its construction and operation are very cost-efficient compared to commercial devices. The analytical performance of the MoBiMS is suitable for real-time studies of biological systems, environmental monitoring, and medical diagnostics.
The composition of acoustically levitated droplets was probed by a novel combination of mid-IR laser evaporation and subsequent postionization via secondary electrospray ionization. The combination of microliter samples and subnanoliter sampling provided time-resolved interrogation of droplets and allowed for a kinetic investigation of the laser-induced release of the analyte, which was found to strongly depend on the analytes. The observed substancespecific delayed release of the analytes permitted baseline-separated discrimination of the analytes, ideal for the study of complex samples.
The additionally applied postionization scheme was found to enable efficient detection of small volatile compounds as well as peptides. The detection of small molecules and peptides occurred under very different sampling geometries, pointing to two distinct underlying ionization mechanisms. Overall, our results suggest that the experimental setup presented in this study can serve as a widely applicable platform to study chemical reactions in acoustically levitated droplets as model reactors.
Formation Mechanism of a Nano-Ring of Bismuth Cations and Mono-Lacunary Keggin-Type Phosphomolybdate
(2022)
A new hetero-bimetallic polyoxometalate (POM) nano-ring was synthesized in a one-pot procedure. The structure consists of tetrameric units containing four bismuth-substituted monolacunary Keggin anions including distorted [BiO8] cubes. The nano-ring is formed via self-assembly from metal precursors in aqueous acidic medium. The compound (NH4)16[(BiPMo11O39)4] ⋅ 22 H2O; (P4Bi4Mo44) was characterized by single-crystal X-ray diffraction, extended X-ray absorption fine structure spectroscopy (EXAFS), Raman spectroscopy, matrix-assisted laser desorption/ionisation-time of flight mass spectrometry (MALDI-TOF), and thermogravimetry/differential scanning calorimetry mass spectrometry (TG-DSC-MS). The formation of the nano-ring in solution was studied by time-resolved in situ small- and wide-angle X-ray scattering (SAXS/WAXS) and in situ EXAFS measurements at the Mo−K and the Bi−L3 edge indicating a two-step process consisting of condensation of Mo-anions and formation of Bi−Mo-units followed by a rapid self-assembly to yield the final tetrameric ring structure.
A brief introduction will be given on modeling chemical reactions in laser induced plasmas using stoichiometric and non-stoichiometric approaches. Several applications will be considered, which can benefit from such modeling. Those include plasma enhanced chemical vapor deposition (PECVD), surface modification and surface coating, and molecular analysis by LIBS. Each application will be illustrated by simulations of relevant chemical systems. For PECVD, chemical systems are BCl3/H2/Ar, BF3/H2/Ar, BCl3/BF3, Mo/BF3/H2; for surface modification/coating it is Ti/air; for molecular LIBS they are CaCO3/Ar, Ca(OH)2/Ar, and CaCl2/Ar. Advantages and shortcomings of equilibrium chemical hydrodynamic models of laser induced plasmas will be discussed.
This course will provide an introduction to plasma diagnostic techniques. The major focus of the course will be on the discussions of the practical procedures as well as the underlying physical principles for the measurements of plasma fundamental characteristics (e.g., temperatures and electron number density). Particular emphasis will be placed on laser induced plasma–atomic emission spectrometry, but other analytical plasmas will also be used as examples when appropriate. Selected examples on how one can manipulate the operating conditions of the plasma source, based on the results of plasma diagnostic measurements, to improve its performance used for spectrochemical analysis will also be covered. Topics to be covered include thermal equilibrium, line profiles, temperatures, electron densities, excitation processes, temporal and spatial resolution.
Surface modification of titanium by laser ablation is investigated theoretically and experimentally. The modification consists in texturing the surface and redeposition of chemically transformed material from the ablation plasma. The redeposition is driven by the hydrodynamic flow in the plasma. Such surface modification improves the biocompatibility of titanium implants.
Laser induced dielectric breakdown (LIDB) on a surface of solid Mo in H2/BF3 atmosphere at 30-760 Torr and in a gaseous mixture MoF6/H2/BF3 + at 760 Torr pressure is tested for synthesis and deposition of superhard molybdenum borides that are needed in many areas of industry and technology. The emission spectra of the plasma and the dynamics of the gas discharge near the substrate are investigated. A comparative analysis of the gas mixture before and after exposure to LIDB plasma is carried out using IR spectroscopy. The conditions for the formation of molybdenum borides are determined. A thermodynamic analysis of the MoF6/H2/BF3 and Mo/H2/BF3 systems is carried out to determine the temperature range for the formation of molybdenum borides and establish the main chemical reactions responsible for their formation. Deposits containing MoB and MoB2 phases are obtained. For the mixture MoF6/H2/BF3, the deposit exhibits an amorphous layered structure, which contains 19.15 wt% F, 30.45% O, and 0.8% Si. For the Mo/H2/BF3 system at the pressures 30 and 160 Torr, nanopowder of molybdenum boride is produced with a characteristic grain size of 100 nm. At pressures above 160 Torr, Mo nanopowder with a grain size <30 nm is obtained.
Raman microspectroscopy enables imaging of the distributions of mineral phases as well as physical properties of materials, such as crystal orientations and crystallinities, with down to sub-micrometre resolution. In a combination with other spectroscopic and microscopic techniques, this approach was applied to the analysis and elucidation of ancient production technologies of stucco fragments made of high-fired gypsum mortar and Egyptian blue pigment discovered on a monochrome wall painting fragment originating from the Early Medieval (5th/6th century AD) construction phase of the church St. Peter above Gratsch in South Tyrol (Northern Italy).
The discussed comparative analyses of Roman Imperial pigment balls and fragmentary murals unearthed in the ancient cities of Aventicum and Augusta Raurica (Switzerland) by means of Raman microspectroscopy pertain to a predecessor study on trace compounds in Early Medieval Egyptian blue (St. Peter, Gratsch, South Tyrol, Northern Italy). The plethora of newly detected associated minerals of the raw materials surviving the synthesis procedure validate the use of quartz sand matching the composition of sediments transported by the Volturno river into the Gulf of Gaeta (Campania, Southern Italy) with a roasted sulphidic copper ore and a mixed-alkaline plant ash as fluxing agent. Thus, the results corroborate a monopolised pigment production site located in the northern Phlegrean Fields persisting over the first centuries A.D., this in line with statements of the antique Roman writers Vitruvius and Pliny the Elder and recent archaeological evidences. Beyond that, Raman spectra reveal through gradual peak shifts and changes of band width locally divergent process conditions and compositional inhomogeneities provoking crystal lattice disorder in the chromophoric cuprorivaite as well as the formation of a copper-bearing green glass phase, the latter probably in dependency of the concentration of alkali flux, notwithstanding that otherwise solid-state reactions predominate the synthesis.
In this presentation, I will give a brief overview of my personal experience with laser induced plasma (LIP). I will start from my and colleagues’ early works, where we used LIP as an atomic reservoir for laser induced fluorescence (LIP). We applied LIP-LIF for a sensitive detection of trace elements in various materials and demonstrated that under certain conditions the technique can even be used for isotope analysis. Next, I will discuss the application of LIP spectroscopy, i.e., LIBS, to material identification that nowadays constitutes one of the best applications of this technique. In those early days, we used correlation analysis for spectra processing; it is now replaced by more powerful chemometric methods. Further, I will stop on our efforts in modeling LIP that we first intended for the improved quality of spectroscopic analysis and later extended to non-spectroscopic fields such as chemical vapor deposition and surface structuring. We developed a version of calibration-free LIBS, in which we iterated model-generated spectra until a close match was achieved between experimental and synthetic spectra to determine concentrations. Next, I will briefly overview our recent developments in plasma modeling that include plasma chemistry. This was important in view of widening application of LIBS as a molecular technique. I will also address several plasma diagnostics, e.g., Radon transform tomography that we developed to get more insight about LIP that was helpful for both analytic spectroscopy and modeling. Finally, I will mention several exotic applications of LIP such as LIP-based lasers and chemical reactors to illustrate a real multifaceted character of laser induced plasma and usefulness of its study for many science fields.
Modeling is an important tool for understanding a physical phenomenon. It helps to interpret results of experiments and optimize experimental parameters for obtaining a desirable result. Modeling laser induced plasma is beneficial for many scientific and industrial fields, e.g., analytical chemistry, pulsed laser deposition, plasma enhanced chemical vapor deposition, laser welding, additive manufacturing etc. In this presentation, a personal experience in development of a physical model of laser induced plasma will be given in a chronological sequence starting from early 2000th and until now.
Over the time, the model evolved from its simple analytical form that described plasma emission spectra to its current numerical form that describes plasma dynamics, chemistry, and interaction with a substrate surface. Several examples will be given for the application of the model to practical problems such as spectroscopic chemical analysis, plasma enhanced chemical vapor deposition, and surface modification by laser ablation.
Concrete structures often show severe damage during their lifetime. One such damage is pitting corrosion of the steel reinforcement caused by chloride ingress into the porous concrete structure. Laser-induced breakdown spectroscopy (LIBS) is a promising method in civil engineering, which is used for detection of chlorine in concrete structures in addition to conventional methods of wet chemistry. To assess LIBS as a trustful analytical technique, its accuracy and robustness is carefully tested. The presentation will outline the results of the interlaboratory comparison of chlorine quantification in cement paste samples, which was carried out by 12 laboratories in 10 countries. Two sets of samples with chloride content ranging from 0.06-1.95 wt.% in the training set and 0.23-1.51 wt.% in the test sample set (“unknowns”), with additional variations in the type of cement and chlorine source (salt type) were sent to the laboratories. The overall result demonstrates that LIBS is suitable for the quantification of the investigated sample compositions: average relative bias was mostly below 15 %. Considering that the laboratories did not receive instructions on how to perform the analysis or how to process the data, the results can be evaluated as a true status quo of the LIBS technique for this type of analysis.
The performance of the Monte Carlo (MC) algorithm for calibration-free LIBS was studied on the example of a simulated spectrum that mimics a metallurgical slag sample. The underlying model is that of a uniform, isothermal, and stationary plasma in local thermodynamical equilibrium.
Based on the model, the algorithm generates from hundreds of thousands to several millions of simultaneous configurations of plasma parameters and the corresponding number of spectra. The parameters are temperature, plasma size, and concentrations of species. They are iterated until a cost function, which indicates a difference between synthetic and simulated slag spectra, reaches its minimum. After finding the minimum, the concentrations of species are read from the model and compared to the certified values. The algorithm is parallelized on a graphical processing unit (GPU) to reduce computational time. The minimization of the cost function takes several minutes on the GPU NVIDIA Tesla K40 card and depends on the number of elements to be iterated. The intrinsic accuracy of the MC calibration-free method is found to be around 1% for the eight elements tested. For a real experimental spectrum, however, the efficiency may turn out to be worse due to the idealistic nature of the model, as well as incorrectly chosen experimental conditions. Factors influencing the performance of the method are discussed.
The Boltzmann plot is one of the most widely used methods for determining the temperature in different types of laboratory plasmas. It operates on the logarithm as a function of the dimensional argument, which assumes that the correct physical units are used. In many works using the Boltzmann method, there is no analysis of the dimension of this argument, which may be the cause of a potential error. This technical note offers a brief description of the method and shows how to correctly use physical units when using transcendental functions like the logarithm.
Calibration-free methods in laser-induced breakdown spectroscopy, CF LIBS, serve as an alternative to calibration-based LIBS techniques. Their major advantage is the ability for fast chemical analysis in situations where matrix-matched standards are not readily available (as, e.g., in the analysis of biological materials and remote analysis) or amount of samples are limited. Their main applications are in the industry, geology, biology, archeology, and even space exploration. This chapter overviews the principle of operation and performance of CF LIBS techniques.
The study, dedicated to Beautiful Pietàs conserved in South Tyrol (Northern Italy), aims to establish, for the first time, a connection between Austroalpine raw materials and the high-fired gypsum mortars constituting the Gothic figure groups in question. The origin and chronology of this stylistically and qualitatively differing ensemble have been subject of art historical debate for nearly a century. The discourse is dominated by three main hypotheses: itinerary of an Austrian artist versus itinerary of the work of art created in an artist’s workshop in Austria versus itinerary of the stylistic vocabulary via graphical or three-dimensional models. The comparison of the δ34S values and the 87Sr/86Sr ratios of the gypsum mortars and Austroalpine sulphate deposits (in a compilation of own reference samples and literature data) points to the exploitation of sediments in the Salzkammergut and possibly also in the evaporite district of the Eastern Calcareous Alps, thus evidencing the import of the sculptures and not the activities of local South Tyrolean or itinerant artists. Two geochronological units are distinguishable: The Pietà in the Church St. Martin in Göflan can be assigned to Upper Permian raw material, whereas the metrologically consistent sculptures in the Church of Our Lady of the Benedictine Abbey Marienberg and in the Chapel St. Ann in Mölten correlate with deposits of the Early Triassic (or the Lower-Middle Triassic transition). The medieval gypsum mortars also differ in their mineralogical characteristics, i.e. in their geologically related minor components, as in the first case, characterised by a significant proportion of primary anhydrite, natural carbonate impurities mainly consist of calcite (partly converted to lime-lump-like aggregates), whereas in the second group dolomite (or rather its hydration products after pyrometamorphic decomposition) predominates, accompanied by celestine, quartz and potassium feldspar. The Pietà in the Cathedral Maria Himmelfahrt in Bozen turned out to be made of Breitenbrunn calcareous sandstone (Leitha Mountains, Burgenland, Austria), which is why the sample is not considered in the geochemical analysis.
The amount of data generated worldwide is constantly increasing. These data come from a wide variety of sources and systems, are processed differently, have a multitude of formats, and are stored in an untraceable and unstructured manner, predominantly in natural language in data silos. This problem can be equally applied to the heterogeneous research data from materials science and engineering. In this domain, ways and solutions are increasingly being generated to smartly link material data together with their contextual information in a uniform and well-structured manner on platforms, thus making them discoverable, retrievable, and reusable for research and industry. Ontologies play a key role in this context. They enable the sustainable representation of expert knowledge and the semantically structured filling of databases with computer-processable data triples.
In this perspective article, we present the project initiative Materials-open-Laboratory (Mat-o-Lab) that aims to provide a collaborative environment for domain experts to digitize their research results and processes and make them fit for data-driven materials research and development. The overarching challenge is to generate connection points to further link data from other domains to harness the promised potential of big materials data and harvest new knowledge.
An alternative method for lithium isotope analysis by using high-resolution atomic absorption spectrometry (HR-CS-AAS) is proposed herein. This method is based on monitoring the isotope shift of approximately 15 pm for the electronic transition 22P←22S at around the wavelength of 670.8 nm, which can be measured by state-of-the-art HR-CS-AAS. Isotope analysis can be used for (i) the traceable determination of Li concentration and (ii) isotope amount ratio analysis based on a combination of HR-CS-AAS and spectral data analysis by machine learning (ML).
In the first case, the Li spectra are described as the linear superposition of the contributions of the respective isotopes, each consisting of a spin-orbit doublet, which can be expressed as Gaussian components with constant spectral position and width and different relative intensity, reflecting the isotope ratio in the sample. Precision was further improved by using lanthanum as internal spectral standard. The procedure has been validated using human serum-certified reference materials. The results are metrologically comparable and compatible with the certified values.
In the second case, for isotope amount ratio analysis, a scalable tree boosting ML algorithm (XGBoost) was employed and calibrated using a set of samples with 6Li isotope amount fractions ranging from 0.06 to 0.99 mol mol−1. The training ML model was validated with certified reference materials. The procedure was applied to the isotope amount ratio determination of a set of stock chemicals and a BAM candidate reference material NMC111 (LiNi1/3Mn1/3Co1/3O2), a Li-battery cathode material. These determinations were compared with those obtained by MC-ICP-MS and found to be metrologically comparable and compatible. The residual bias was −1.8‰, and the precision obtained ranged from 1.9‰ to 6.2‰. This precision was sufficient to resolve naturally occurring variations. The NMC111 cathode candidate reference material was analyzed using high-resolution continuum source atomic absorption spectrometry with and without matrix purification to assess its suitability for technical applications. The results obtained were metrologically compatible with each other.
Die BAM setzt in ihren analytischen Labors zunehmend Labor-Robotik ein, um gefähliche oder zeitraubende Routineaufgaben zu automatisieren. Durch Automation kann auch die Reproduzierbarkeit solcher Anwendungen erhöht werden.
Im Beitrag werden einige aktuelle Beispiele aus den Analytiklabors diskutiert, wie z. B. die Anwendung für die Herstellung und RFA-Analytik von Gläsern, eine Feinwäge-Robotik für 10-L-Gaszylinder im Rahmen der Herstellung von Primärnormalen oder die Automatisierung von Probenpräsentationen für optische Spektroskopie und chemical Imaging. Auch wird kurz auf den Einsatz von Speicherprogrammierbaren Steuerungen in der modularen Laborautomation und die Virtualisierung von analytischen Laborrechnern eingengangen.
Electrochemical methods make great promise to meet the demand for user-friendly on-site devices for monitoring important parameters. Food industry often runs own lab procedures, e.g., for mycotoxin analysis, but it is a major goal to simplify analysis, linking analytical methods with miniaturized technologies. Enzyme-linked immunosorbent assays, with photometric detection of the horseradish peroxidase (HRP) substrate 3,3’,5,5’-tetramethylbenzidine (TMB), form a good basis for sensitive detection. To provide a straight-forward approach for the miniaturization of the detection step, we have studied the pitfalls of the electrochemical TMB detection. By cyclic voltammetry, it could be shown that TMB electrochemistry is strongly dependent on the pH and the electrode material.
It was found that screen-printed gold electrodes and a very low pH value (pH 1) are well-suited to perform the electrochemical detection of TMB, due to the reversible character of the redox reaction under these conditions. Under these conditions, a good signal stability over several measuring cycles is achieved, providing the basis for analyzing multiple samples. In contrast to this, for carbon screen-printed electrodes, it was found that the signal response has changed after the electrochemical reaction with TMB at pH 1. At moderately acidic conditions (pH 4), neither with carbon nor with gold electrodes a reproducible electrochemical detection of TMB could be achieved.
Based on these findings, we created a smartphone-based, electrochemical, immunomagnetic assay for the detection of ochratoxin A (OTA) and ergometrine in food samples. A competitive assay is performed on magnetic beads using HRP and TMB/H2O2 to generate the signal. Enzymatically oxidized TMB is quantified after addition of H2SO4 by amperometry with screen-printed gold electrodes in a custom-made wall-jet flow cell. The results are in good correlation with the established photometric detection method, providing a solid basis for sensing of further analytes in HRP-based assays using the newly developed miniaturized smartphone-based, electrochemical, immunomagnetic assay.
Steps Forward in the Application of 3,3’,5,5’- Tetramethylbenzidine (TMB) in Amperometric Assays
(2022)
Electrochemical methods make great promise to meet the demand for user-friendly on-site devices for monitoring important parameters. Food industry often runs own lab procedures, e.g. for mycotoxin analysis, but it is a major goal to simplify analysis, linking analytical methods with miniaturized technologies. Enzyme-linked immunosorbent assays, with photometric detection of the horseradish peroxidase (HRP) substrate, 3,3’,5,5’-tetramethylbenzidine (TMB), form a good basis for sensitive detection. To provide a straight-forward approach for the miniaturization of the detection step, we have studied the pitfalls of the electrochemical TMB detection. By cyclic voltammetry it was found that the TMB electrochemistry is strongly dependent on the pH and the electrode material.
It was found that screen-printed gold electrodes and a highly acidic pH value (pH 1) are well-suited to perform the electrochemical detection of TMB, due to the reversible character of the redox reaction under these conditions. This set-up provides a good signal stability over several measuring cycles, providing the basis for analysing multiple samples. In contrast to this, for carbon screen-printed electrodes, it was found that the signal response has changed after the electrochemical reaction with TMB at pH 1. At a weakly acidic pH value (pH 4), neither with carbon nor with gold electrodes a reproducible electrochemical detection of TMB could be achieved [1].
Based on these findings we created a smartphone-based, electrochemical, immunomagnetic assay for the detection of ochratoxin A and ergometrine in real samples. Therefore, a competitive assay was performed on magnetic beads using HRP and TMB/H2O2 to generate the signal. Enzymatically oxidized TMB was quantified after addition of H2SO4 by amperometry with screen-printed gold electrodes in a custom-made wall-jet flow cell. The results are in good correlation with the established photometric detection method, providing a solid basis for sensing of further analytes in HRP-based assays using the newly developed miniaturized smartphone-based, electrochemical, immunomagnetic assay.