Quantitative analysis of a bulk sample requires that the
composition of the sample is homogeneous over the analyzed
volume. For inhomogenous samples the calculation of
the matrix effects is not correct and this can lead to wrong results in the element concentrations. For samples containing a layer structure a different quantitative evaluation has to be applied. This can be provided with the standard-based analysis in ESPRIT in combination with the STRATAGem software.
Analysis of Elemental Composition of Fe1-xNix and Si1-xGex Alloy Thin Films by EPMA and μ-XRF
(2019)
The present study reports on measurements on thin Fe-Ni films on silicon and first-time results of analysis on Si-Ge thin films deposited on a non-conductive aluminium oxide Substrate by electron probe microanalysis (EPMA). Standard-based and standardless EPMA (with EDS) results were used in combination with the thin film analysis software Stratagem for the quantification.
Further, X-ray fluorescence analysis (XRF) can be used for the determination of elemental composition and thickness of such films as well. In this case, XRF with a μ-focus X-ray source (μ-XRF) attached to a SEM was applied. For quantification, a fundamental parameter (FP) approach has been used to calculate standard-based and standardless results.
Both thin film systems have been chosen as samples of an international round robin test (RRT) organised in the frame of standardisation technical committee ISO/TC 201 ‘Surface chemical analysis’, under the lead of KRISS. The main objective of the RRT is to compare the results of atomic fractions of Fe1-xNix and Si1-xGex alloy films obtained by different surface Analysis techniques, such as X-ray photoelectron spectroscopy (XPS), Auger electron spectroscopy (AES), and secondary ion mass spectrometry (SIMS) applied in the depth-profiling operation mode.
Five samples of different atomic fractions of each thin film system, i.e., Fe1-xNix and Si1-xGex, have been grown by ion beam sputter deposition on silicon and Al2O3 wafers, respectively. Reference FeNi and SiGe films with well-known elemental composition and thickness have been also supplied for standard-based analysis. An excellent agreement has been obtained between the atomic fractions determined by EPMA and µ-XRF with the KRISS certified values.zeige mehr
It was demonstrated in the past that the electron probe microanalysis (EPMA) can be applied to determine accurately both elemental composition and thickness of thin films by using the dedicated software package for thin film analysis Stratagem. A relatively small number of film materials such as pure metallic films of platinum and nickel, binary alloys of Fe-Ni, and Pt-Ni-Co ternary alloy films has been reported in literature as working successfully. Further, the software can be applied ‘inversely’, i.e., by feeding it with the thickness of the film and using the determined mass coating, one can easily calculate the film density, which for porous layers leads us to the true film porosity.
The present study repeats measurements on an already tested system of Fe-Ni thin films on silicon and reports for the first-time results of analysis on Si-Ge thin films deposited on a non-conductive aluminium oxide substrate. Standard-based and standardless EPMA (with EDS) results were used in combination with Stratagem for the quantification.
Further, X-ray fluorescence analysis (XRF) can be used for the determination of elemental composition and thickness of such films as well. In this case, XRF with a μ-focus X-ray source (μ-XRF) attached to a SEM was applied. For quantification, a fundamental parameter (FP) approach has been used to calculate standard-based and standardless results. Compared to EPMA, XRF has a larger information depth and a higher elemental sensitivity because of a generally lower background.
Both thin film systems have been chosen as samples of an international round robin test (RRT) organised in the frame of standardisation technical committee ISO/TC 201 ‘Surface chemical analysis’, under the lead of KRISS. The main objective of the RRT is to compare the results of atomic fractions of Fe1-xNix and Si1-xGex alloy films obtained by different surface Analysis techniques, such as X-ray photoelectron spectroscopy (XPS), Auger electron spectroscopy (AES), and secondary ion mass spectrometry (SIMS) applied in the depth-profiling operation mode.
Five samples of different atomic fractions of each thin film system, i.e., Fe1-xNix and Si1-xGex, have been grown by ion beam sputter deposition on silicon and Al2O3 wafers, respectively. Reference FeNi and SiGe films with well-known elemental composition and thickness have been also supplied for standard-based analysis. The atomic fractions of all the samples including the references have been certified by RBS (Rutherford backscattering spectrometry) and ICP-AES (inductively coupled plasma atomic emission spectroscopy).
Measurement of elemental composition of Fe1-xNix and Si1-xGex alloy thin films by EMPA and µ-XRF
(2019)
It was demonstrated in the past that the electron probe microanalysis (EPMA) can be applied to determine accurately both elemental composition and thickness of thin films by using the dedicated software package for thin film analysis Stratagem. A relatively small number of film materials such as pure metallic films of platinum and nickel, binary alloys of Fe-Ni, and Pt-Ni-Co ternary alloy films has been reported in literature as working successfully. Further, the software can be applied ‘inversely’, i.e., by feeding it with the thickness of the film and using the determined mass coating, one can easily calculate the film density, which for porous layers leads us to the true film porosity.
The present study repeats measurements on an already tested system of Fe-Ni thin films on silicon and reports for the first-time results of analysis on Si-Ge thin films deposited on a non-conductive aluminium oxide substrate. Standard-based and standardless EPMA (with EDS) results were used in combination with Stratagem for the quantification.
Further, X-ray fluorescence analysis (XRF) can be used for the determination of elemental composition and thickness of such films as well. In this case, XRF with a μ-focus X-ray source (μ-XRF) attached to a SEM was applied. For quantification, a fundamental parameter (FP) approach has been used to calculate standard-based and standardless results. Compared to EPMA, XRF has a larger information depth and a higher elemental sensitivity because of a generally lower background.
Both thin film systems have been chosen as samples of an international round robin test (RRT) organised in the frame of standardisation technical committee ISO/TC 201 ‘Surface chemical analysis’, under the lead of KRISS. The main objective of the RRT is to compare the results of atomic fractions of Fe1-xNix and Si1-xGex alloy films obtained by different surface Analysis techniques, such as X-ray photoelectron spectroscopy (XPS), Auger electron spectroscopy (AES), and secondary ion mass spectrometry (SIMS) applied in the depth-profiling operation mode.
Five samples of different atomic fractions of each thin film system, i.e., Fe1-xNix and Si1-xGex, have been grown by ion beam sputter deposition on silicon and Al2O3 wafers, respectively. Reference FeNi and SiGe films with well-known elemental composition and thickness have been also supplied for standard-based analysis. The atomic fractions of all the samples including the references have been certified by RBS (Rutherford backscattering spectrometry) and ICP-AES (inductively coupled plasma atomic emission spectroscopy).
A short overview of thin film analysis by SEM/EDX is presented. Relevant steps such as optimization of instrumnetal parameters, extraction of net peak areas, use of standards, calculation of k-values and quantification with the dedicated thin film analysis software (Stratagem) are explained. Examples of different layered aystems are given together with validation of the results.