CCQM key comparison K-129 for the quantitative analysis of Cu(In,Ga)Se2 (CIGS) films has been performed by the Surface Analysis Working Group (SAWG) of the Consultative Committee for Amount of Substance (CCQM). The objective of this key comparison is to compare the equivalency of the National Metrology Institutes (NMIs) and Designated Institutes (DIs) for the measurement of mole fractions of Cu, In, Ga and Se in a thin CIGS film. The measurand of this key comparison is the average mole fractions of Cu, In, Ga and Se of a test CIGS alloy film in the unit of mole fraction (mol/mol). Mole fraction with the metrological unit of mol/mol can be practically converted to atomic fraction with the unit of at%.
In this key comparison, a CIGS film with certified mole fractions was supplied as a reference specimen to determine the relative sensitivity factors (RSFs) of Cu, In, Ga and Se. The mole fractions of the reference specimen were certified by isotope dilution - inductively coupled plasma/mass spectrometry (ID-ICP/MS) and are traceable to the SI. A total number counting (TNC) method was recommended as a method to determine the signal intensities of the constituent elements acquired in the depth profiles by Secondary Ion Mass Spectrometry (SIMS), X-ray Photoelectron Spectroscopy (XPS) and Auger Electron Spectroscopy (AES). Seven NMIs and one DI participated in this key comparison. The mole fractions of the CIGS films were measured by depth profiling based-SIMS, AES and XPS. The mole fractions were also measured by non-destructive X-Ray Fluorescence (XRF) Analysis and Electron Probe Micro Analysis (EPMA) with Energy Dispersive X-ray Spectrometry (EDX).
In this key comparison, the average degrees of equivalence uncertainties for Cu, In, Ga and Se are 0.0093 mol/mol, 0.0123 mol/mol, 0.0047 mol/mol and 0.0228 mol/mol, respectively. These values are much smaller than that of Fe in a Fe-Ni alloy film in CCQM K-67 (0.0330 mol/mol). This means that the quantification of multi-element alloy films is possible by depth profiling analysis using the TNC method.
Under the Comité Consultatif pour la Quantité de Matière (CCQM), a key comparison, CCQM-K104, was coordinated by the National Institute of Metrology (NIM). The comparison was designed to demonstrate a laboratory's performance in determining the mass fraction of the main component in a complex high purity organic material. Nine NMIs or DIs participated in the comparison. Eight participants reported their results. An additional impurity was resolved from the avermectin B1a peak and was tentatively identified as an unknown impurity by NMIA (National Measurement Institute (Australia)). It was subsequently identified by NIM as a diastereoisomer of avermectin B1a at the C-26 position. Final reference value (KCRV) = 924.63 mg/g, with uncertainty (k=1) = 3.89 mg/g, and expanded uncertainty = 8.97 mg/g. The degrees of equivalence with the avermectin B1a KCRV for each participant were reported. The measurement results and degrees of equivalence should be indicative of the performance of a laboratory's measurement capability for the purity assignment of organic compounds of high structural complexity (relative molecular mass range of 500 Da -1000 Da and low polarity (-log KOW ≤ -2).
CCQM has established a framework of comparisons to demonstrate the international comparability of chemical measurements. The key point is the establishment of comparable measurements, with traceability to internationally or nationally stated references.
Originally initiated by Dr Martin P Seah, NPL, the Surface Analysis Working Group (SAWG) has been formally founded in 2003. The following terms of reference were presented to CCQM in April 2002. CCQM ratified the group as a full working group of CCQM in April 2003 with these terms of reference:
- to develop pilot studies and carry out key comparisons of national measurement standards for surface and micro/nano-analysis;
- to assist in identifying and establishing inter-laboratory work to improve the traceability of surface and micro/nano-analysis;
- to establish and update a work plan to be adopted by CCQM;
- to discuss and review the scope of the working group and to liase with other working groups related to nanotechnology.
Following the 2016 meeting, the Key Comparison K-129 "Measurement of atomic fractions in Cu(In,Ga)Se2 Films" lead by KRISS and the Key Comparison K-136 on "BET specific surface area of nanoporous Al2O3" lead by UNIIM&BAM have been finished in 2016.
The 15th meeting of SAWG will focus on
- survey on CMC claim submitted with reference to K-129 and K-136.
- the overall and SAWG specific aspects of the CCQM Strategy process,
- a joint meeting with the Inorganic Working Group addressing number concentrations of nano particles,
- discussion of future comparisons.
High purity materials can serve as a realisation of the Système International d’Unitès (SI) unit amount of substance for the specific element. Solutions prepared from such high purity materials using gravimetric preparation and the concept of molar mass are used as calibration solutions in many fields of analytical chemistry. Calibration solutions prepared this way provide the traceability to the SI and are the metrological basis in elemental analysis.
The preparation and characterization of such primary pure substances, representing the realisation of the SI unit amount of substance, is undertaken only by a small number of National Metrology Institutes (NMI) and Designated Institutes (DI).
Many other NMIs and DIs, however, prepare elemental calibration solutions as calibrants for their measurement services, such as the certification of matrix Reference Materials or the provision of reference values for Proficiency Testing schemes. The elemental calibration solutions used for this purpose are not a direct service to customers, such as preparing secondary calibration solutions, but provide the source of traceability for the other services. Hence, it is necessary for the NMI or DI to obtain data on the purity of the pure metals or other materials used to prepare the solutions with measurement uncertainties meeting the needs of the above described services. This is commonly undertaken as a “fit for purpose” assessment, appropriate for the uncertainty requirement of the service provided to customers.
As a consequence, total purity measurements are a long-term strategy of CCQM-IAWG. Several studies were conducted (CCQM-P107, CCQM-K72 and CCQM-P149) on the measurement of the purity of zinc. From these studies, several conclusions can be drawn for the purity assessment of a pure (metallic) element. These conclusions will be put together in this document in order to assist all NMIs/DIs in performing a purity assessment, whenever needed.
Within the Working Group on Inorganic Analysis (IAWG) of the Consultative Committee for Amount of Substance: Metrology in Chemistry and Biology (CCQM) international key comparisons and pilot studies related to inorganic analysis are carried to ensure consistency in this field at the highest level. Some of these comparisons deal directly with the preparation and characterization of monoelemental solutions or with topics, closely related. The importance of monoelemental solutions lies in the fact that almost every measurement in inorganic analysis relies on the comparison with either a reference material, or references in form of solutions, usually (mono)elemental solutions. All quantitative measurement approaches, e.g. isotope dilution or standard addition, need an accurate reference solution made from a well characterized reference material, prepared under full gravimetric control. These primary (monoelemental) solutions do not only serve as arbitrary references/calibration solutions, but they also link up measurement results to the International System of units (SI), this way establishing the so-called metrological traceability to a measurement unit of the SI. Without such solutions on the highest possible level of accuracy and with the smallest possible associated uncertainties (for e.g. element content and/or impurities), an analysis itself can never be as good as it could be with appropriate reference solutions. This article highlights select key comparisons and pilot studies dealing with monoelemental solution related topics within the IAWG from the foundation of CCQM – 25 years ago – up to latest achievements in the field of inorganic analysis.
Shortly after founding the CCQM in 1993, the first key comparison (CCQM-K2) about the determination of Cd and Pb content in river water was hosted by IRMM. It has been clear from the beginning that for the determination of an element mass fraction in a matrix, accurate reference solutions would be the key point for a metrologically sound analysis. Triggered by the not entirely satisfactory results it was decided to investigate the basis of all measurements: the reference solutions. Thus in 1999, CCQM-K8, hosted by EMPA/LNE, started the tedious but indispensable work on elemental solutions standards.
Additionally, PTB, BAM and Merck KGaA initiated a project to establish a traceability system for inorganic analysis based on accurate monoelemental solutions. Within this unique approach ten high purity reference materials were fully characterized, and solutions were prepared gravimetrically thereof. These materials (forming the primary standards of elemental analysis) are available for NMIs/DIs from BAM.
CCQM-P46 (Cu, Mg, Rh) demonstrated the actuality of the topic and the challenges to prepare elemental solutions for every element. The accurate preparation and use of monoelemental solutions is consequently reflected in the recurrent conduct of (key) comparisons: CCQM-P149 (purity of zinc), CCQM-K87 (Co, Cr, Pb), and the current CCQM-K143/P181 (Cu solutions), to name only a few in a long series.
Is it necessary to put so much effort into the preparation of reference solutions? It clearly is! An example outside the world of academia: Recently, EDQM*, PTB, BAM and JRC have developed monoelemental reference standards traceable to the SI for toxic elements in support of the chapters of the European Pharmacopoeia, related to the analysis of elemental impurities in medicinal products. A new ICP OES method has been established allowing the comparison of two 1 g/kg mercury solutions with an outstandingly small uncertainty of Urel(w(Hg)) = 0.16 %. However, this technical progress would not have been possible without the ongoing effort of the metrological community within 25 years of research in the field of high accuracy monoelemental solutions.
* EDQM: European Directorate for the Quality of Medicines and HealthCare
ICP-MS has played a key role in inorganic chemical metrology for 25 years, from the 1993 CIPM feasibility study which led to establishment of the CCQM. Since that time, the Inorganic Analysis Working Group of the CCQM has organised 56 international comparisons involving measurements by ICP-MS and, in a recent comparison, 16 different national institutes submitted their results using the technique. Metrological applications of ICP-MS currently address an enormous range of measurements using a wide variety of instrumentation, calibration strategies and methodologies. This review provides an overview of the ICP-MS field with an emphasis on developments which are of particular relevance to chemical metrology.
Examples from CCQM comparisons and the services available from the participants are used to illustrate how the capability and scope of ICP-MS methods have expanded far beyond the expectations of 1993. This is due in part to the research and development Programmes of the national institutes which participate in the CCQM. They have played a key role in advancing new instrumentation and applications for elemental analysis, isotope dilution mass spectrometry, determination of isotopic ratio or composition, and speciation of organometallic compounds. These developments are continuing today, as demonstrated by work in new fields such as heteroatom quantitation of proteins, characterisation and counting of nanoparticles using spICP-MS, and LA-ICP-MS analysis of solid materials.