Lead (Pb) isotope amount ratios are commonly used in applications ranging from archaeology and forensic sciences to terrestrial and extra-terrestrial geochemistry. Despite their utility and frequency of use, only three certified isotope amount ratio reference materials are currently available for Pb: NIST SRMs 981, 982 and 983. Because SRM 981 has a natural Pb isotopic composition, it is mainly used for correcting instrumental mass discrimination or fractionation. This means that, at present, there are no other certified isotope reference materials with natural Pb isotopic composition that could be used for validating or verifying an analytical procedure involving the measurement of Pb isotope amount ratios.
To fill this gap, two new reference materials, both certified for their Pb isotopic composition, have been produced together with a complete uncertainty assessment. These new reference materials offer SI traceability and an independent means of validating or verifying analytical procedures used to produce Pb isotope amount ratio measurements.
ERM-EB400 is a bronze material containing a nominal Pb mass fraction of 45 mg/kg. ERM-AE142 is a high purity solution of Pb with a nominal mass fraction of 100 mg/kg. Both materials have been specifically produced to assist analysts in verifying or validating their analytical procedures. Note that while one of these reference materials requires the chemical separation of Pb from its matrix (ERM-EB400), the other does not (ERM-AE142). Details on the certification of these isotope reference materials are provided in this report.
Certification report for the reference materials ERM-AE102a, -AE104a, -AE120, -AE121 and -AE122
(2010)
Isotope reference materials are essential to enable reliable and comparable isotope data. Besides the correction of mass fractionation or mass discrimination isotope reference materials are indispensible for validation and quality control of analytical procedures. This article describes the production and certification of a set of five isotope reference materials ERM-AE102a, 104a, AE120, 121 and 122, for boron isotope analysis. The isotopic composition of all materials has been adjusted by mixing boron mother solutions enriched in 10B or 11B with a boron mother solution having natural-like isotopic composition under full gravimetric control. All mother solutions have been analysed for their boron mass fraction as well as their boron isotopic composition by TIMS using IDMS as calibration technique. For all five reference materials the isotopic composition obtained on the basis of the gravimetric data agrees very well with the isotopic composition obtained from different TIMS techniques. Performed stability and homogeneity studies show no significant influence on the isotopic composition as well as on the related uncertainties.
The certified isotope abundances for 10B are 0.29995 (27) for ERM-AE102a and 0.31488 (28) for ERM-AE104a. The certified δ11B values are -20.2 (6) ‰ for ERM-AE120, 19.9(6) ‰ for ERM-AE121 and 39.7 (6) ‰ for ERM-AE122.
Together with the formerly certified ERM-AE101 and -AE103 a unique set of seven certified reference materials (CRM) for boron isotope analysis is now available from BAM.
This Report describes the certification of the reference material antimony implanted in Si/SiO2 intended to be used for calibration of surface and near surface analytical methods. It describes the preparation, homogeneity measurements and the analytical work performed for the certification of both Areal density of antimony Atoms (retained dose) and the isotope amount Ratio as well as giving considerations on the stability of the material.