Polycyclic aromatic hydrocarbons (PAHs) are a large group of priority organic pollutants, which contaminate environmental compartments, food, and consumer products as well. Due to their frequent occurrence associated with elevated Levels of PAHs, plastic and rubber parts of consumer products and toys are particular sources of exposure. Although European maximum levels exist for eight carcinogenic PAHs in consumer products and toys according to REACH Regulation (EC) No. 1907/2006, certified reference materials (CRM) are still not available. To overcome this lack, the first CRM for the determination of PAHs in rubber toys (BAM-B001) was developed according to the requirements of ISO Guide 35. The whole process of CRM development including preparation, homogeneity and stability studies, and value assignment is presented.
The assignment of the certified mass fractions was based upon in-house study at BAM using stable isotope Dilution analysis (SIDA) gas chromatography mass spectrometry (GC–MS). The obtained values were confirmed by the results of two interlaboratory comparison (ILC) studies with more than 50 expert laboratories from Germany and China. The mass fractions of 14 PAHs including all REACH and GS mark regulated compounds were certified ranging between 0.2 and 15.4 mg/
kg accompanied by expanded uncertainties (coverage factor k = 2). In addition, informative values were determined for 4 PAHs, mainly due to higher uncertainties and/or lack of ILC data for confirmation. BAM-B001 is intended for analytical quality control particularly based on the AfPS GS 2019:01 PAK method and contributes to improve the chemical safety of consumer products including toys.
Stir bar sorptive extraction (SBSE) was compared with standardized pump sampling regarding the prospects to assess airborne levels of polycyclic aromatic hydrocarbons (PAHs) in indoor environments. A historic railway water tower, which will be preserved as a technical monument for museum purposes, was sampled with both approaches because the built-in insulationmaterial was suspected to release PAHs to the indoor air. The 16 PAHs on the US EPA list were quantified using gas chromatography with mass spectrometric detection in filters from pump sampling after solvent extraction and on SBSE devices after thermal desorption. SBSEwas seen to sample detectable PAHmasseswith excellent repeatability and a congener pattern largely similar to that observed with pump sampling. Congener patterns were however significantly different from that in the PAH source because release from the insulation material is largely triggered by the respective congener vapor pressures. Absolute masses in the ng range sampled by SBSE corresponded to airborne concentrations in the ng L−1 range determined by pump sampling. Principle differences between SBSE and pump sampling as well as prospects of SBSE as cost-effective and versatile complement of pump sampling are discussed.